How to Use Linshang LS252D Leeb Hardness Tester for Metal Testing Malaysia

How to Use Linshang LS252D Leeb Hardness Tester for Metal Testing Malaysia

 

Portable hardness testing is useful when large, heavy or installed metal components cannot conveniently be moved to a laboratory.

 

The Linshang LS252D Leeb Hardness Tester is a compact handheld instrument with an integrated D-type impact device, designed to help engineers, technicians and QA/QC personnel perform hardness measurements directly on suitable metallic workpieces.

 

However, obtaining a reliable Leeb hardness reading involves more than simply pressing the tester against the metal.

 

Correct surface preparation, workpiece support, material selection and impact technique are essential.

 

This guide explains the basic considerations when using the Linshang LS252D for industrial metal hardness testing in Malaysia.

 

Step 1: Identify the Material Being Tested

 

Before testing, determine the material of the workpiece.

 

The LS252D supports various metallic material groups, including:

 

Steel

 

Cast steel

 

Alloy tool steel

 

Stainless steel

 

Grey cast iron

 

Nodular cast iron

 

Cast aluminium alloy

 

Brass

 

Bronze

 

Copper alloy

 

Forged steel

 

Correct material selection is important because hardness conversion relationships differ between materials.

 

A hardness conversion that is appropriate for steel, for example, should not automatically be assumed to apply to aluminium or copper alloys.

 

Step 2: Check the Workpiece Size and Rigidity

 

Leeb hardness testing uses an impact and rebound principle.

 

Because of this, the physical characteristics of the workpiece can influence the result.

 

Before measurement, consider:

 

Workpiece mass

 

Thickness

 

Rigidity

 

Shape

 

Curvature

 

Surface condition

 

Stability during impact

 

Large and heavy workpieces are generally well suited to portable Leeb testing.

 

This is one reason the LS252D can be useful for applications such as:

 

Large machinery

 

Machine components

 

Mold cavities

 

Automotive components

 

Pressure vessel components

 

Heavy fabricated parts

 

Installed equipment

 

Step 3: Properly Support Small Workpieces

 

Small or relatively light workpieces require additional attention.

 

If the component moves or vibrates when the impact device is released, the measurement can be affected.

 

For suitable smaller components, proper coupling to a solid and rigid supporting body may be required.

 

The workpiece should not shift during the measurement.

 

For very small, thin or flexible components, another hardness testing method may be more appropriate.

 

Step 4: Prepare the Testing Surface

 

Surface condition is one of the most important factors in portable hardness testing.

 

The test location should be clean and suitable for measurement.

 

Remove contaminants such as:

 

Oil

 

Dirt

 

Rust

 

Loose coating

 

Scale

 

Heavy oxidation

 

Other surface contamination

 

Where necessary and permitted by the inspection procedure, the measurement area may need appropriate surface preparation.

 

A rough, uneven or unstable surface can contribute to inconsistent results.

 

Step 5: Select the Correct Material Setting

 

Configure the LS252D according to the material being measured.

 

This is especially important when using converted hardness values.

 

The LS252D supports multiple hardness scales, including:

 

HL – Leeb

 

HV – Vickers

 

HB – Brinell

 

HS – Shore

 

HRA – Rockwell A

 

HRB – Rockwell B

 

HRC – Rockwell C

 

For many applications, users may perform the Leeb measurement and view a converted hardness value corresponding to the required engineering specification.

 

Step 6: Select the Required Hardness Scale

 

Choose the hardness scale relevant to the inspection requirement.

 

For example, an engineering drawing or heat-treatment specification may require hardness to be reported in HRC.

 

Other applications may use HB or HV.

 

Do not select a hardness scale simply because it is familiar. The correct scale should correspond to the material, applicable specification and testing requirement.

 

Step 7: Position the LS252D Correctly

 

Place the instrument firmly against the prepared measurement surface.

 

The support ring should sit properly on the workpiece.

 

Avoid unstable positioning.

 

Poor contact, excessive movement or incorrect positioning can cause inconsistent readings.

 

Step 8: Perform the Impact Measurement

 

Load the impact mechanism according to the operating procedure and perform the measurement while keeping the tester stable.

 

Avoid moving the instrument during impact.

 

The LS252D measures the Leeb hardness value based on the relationship between the impact and rebound velocities of the impact body.

 

The result is then displayed on the OLED screen.

 

Step 9: Take Multiple Measurements

 

For industrial inspection, relying on a single measurement is generally not ideal.

 

Take several measurements within the appropriate test area while maintaining sufficient spacing between impact points and edges.

 

Multiple readings help identify:

 

Measurement variation

 

Local material differences

 

Surface preparation problems

 

Incorrect positioning

 

Abnormal results

 

The inspection procedure or applicable standard should determine how results are evaluated and reported.

 

Step 10: Review Any Unusual Reading

 

Suppose several measurements are similar:

 

Reading 1: 612 HL

Reading 2: 608 HL

Reading 3: 615 HL

Reading 4: 610 HL

 

and another reading is significantly different.

 

Instead of immediately concluding that the material hardness has changed, inspect the measurement location and testing conditions.

 

Possible causes include:

 

Surface contamination

 

Roughness

 

Poor contact

 

Workpiece movement

 

Curved measurement surface

 

Incorrect material selection

 

Incorrect measurement technique

 

Local material variation

 

Repeat the test correctly before making an engineering conclusion.

 

Why Impact Direction Matters

 

Leeb hardness measurements can be affected by impact orientation.

 

Field measurements are not always performed vertically downward. Engineers may need to test:

 

Horizontal surfaces

 

Vertical surfaces

 

Side walls

 

Installed machinery

 

Components at difficult orientations

 

The LS252D provides automatic impact direction compensation, making it more convenient when testing components in different orientations.

 

Nevertheless, the instrument must still be positioned correctly and used according to the manufacturer's instructions.

 

Why Surface Preparation Matters

 

A Leeb hardness tester measures a dynamic rebound response.

 

This means poor surface conditions can influence the impact behavior.

 

For example, testing directly on a heavily rusted or rough surface may produce results that do not accurately represent the underlying material.

 

Good surface preparation improves measurement consistency.

 

This is especially important when comparing hardness before and after:

 

Heat treatment

 

Repair

 

Fabrication

 

Welding-related processes

 

Maintenance

 

Component replacement

 

Why Workpiece Weight Matters

 

Imagine hitting two objects with the same impact energy.

 

One is a large steel block.

 

The other is a small loose steel plate.

 

The small plate may move or vibrate during impact.

 

A similar issue can occur during Leeb hardness testing.

 

This is why workpiece mass and support are important.

 

The LS252D is particularly useful for large components because heavy workpieces generally provide a stable base for the impact measurement.

 

Common Mistakes When Using a Portable Leeb Hardness Tester

 

Testing Directly on Rust

 

Rust, scale and contamination can interfere with the measurement surface.

 

Prepare the test area properly.

 

Testing a Loose Small Component

 

A lightweight workpiece can move during impact.

 

Use an appropriate support or coupling method where applicable.

 

Using the Wrong Material Setting

 

Incorrect material selection can result in inappropriate hardness conversion.

 

Always verify the material before testing.

 

Depending on Only One Reading

 

One measurement may not adequately represent the component.

 

Take multiple readings according to the applicable inspection procedure.

 

Testing Too Close to an Edge

 

Measurement location matters.

 

Avoid unsuitable locations near edges or other areas that may affect impact behavior.

 

Assuming HRC Is Directly Measured

 

The LS252D fundamentally performs Leeb hardness measurement. Other supported hardness scales are obtained through conversion according to the applicable material relationship.

 

This distinction is important when a project requires testing according to a specific standard or laboratory method.

 

Example: Heat-Treated Steel Component

 

Consider a Malaysian machine shop receiving a batch of heat-treated steel components.

 

The specification requires a certain hardness range.

 

Instead of transporting every large component to a laboratory, the QA/QC team can use a portable hardness tester to perform suitable on-site checks.

 

A typical workflow may involve:

 

Confirming the material

 

Preparing the measurement surface

 

Selecting the correct material setting

 

Selecting the required hardness scale

 

Performing several measurements

 

Comparing the results

 

Investigating abnormal readings

 

Recording the results

 

This can make hardness inspection significantly more convenient for large components.

 

Example: Installed Machinery

 

Another useful application involves machinery that has already been installed.

 

Removing a component simply to perform a hardness check may require significant downtime.

 

For suitable components, the LS252D allows maintenance personnel to bring the instrument to the machine rather than bringing the machine component to the instrument.

 

This is one of the key advantages of portable hardness testing.

 

Is the LS252D Suitable for Your Application?

 

The Linshang LS252D is particularly worth considering when you need to test:

 

Large steel components

 

Heavy metal parts

 

Installed machinery

 

Mold and die components

 

Pressure vessel components

 

Automotive components

 

Machine tool components

 

Fabricated metal structures

 

Heat-treated components

 

For small, thin or lightweight workpieces, the application should be evaluated more carefully before selecting a Leeb hardness tester.

 

Portable Hardness Tester for QA/QC in Malaysia

 

For Malaysian manufacturers and engineering companies, portable hardness testing can help bring inspection closer to the actual production or maintenance location.

 

Instead of depending entirely on laboratory testing for every suitable large component, QA/QC teams can perform convenient on-site hardness checks using a portable instrument such as the Linshang LS252D.

 

Correct technique remains essential.

 

A good hardness tester combined with poor testing practice can still produce unreliable results.

 

Understanding the material, workpiece and testing conditions is therefore just as important as choosing the instrument itself.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd.

 

Showroom & Service Centre

 

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia

 

🌐 Website: www.mtmpre.com.my

📧 Email: mtmpre@yahoo.com

📱 WhatsApp: +6016-660 7346

 

Contact MTM Precision for Linshang LS252D Leeb Hardness Tester enquiries and portable hardness testing solutions in Malaysia.


 

12 Aug 2026