How to Use OTYWELL G40S for Confined Space Gas Testing Malaysia

How to Use OTYWELL G40S for Confined Space Gas Testing Malaysia

 

Confined spaces can contain oxygen deficiency, toxic gases, combustible gases or a combination of atmospheric hazards. A space that appears clean and safe from the outside may still contain an atmosphere capable of causing unconsciousness, fire, explosion or fatal exposure.

 

The OTYWELL G40S portable four-gas detector can be configured to monitor the four gases commonly associated with confined-space work:

 

Combustible gas – LEL

 

Oxygen – O₂

 

Carbon monoxide – CO

 

Hydrogen sulphide – H₂S

 

The G40S also provides functions such as TWA and STEL alarms, fall detection, data storage, sound-light-vibration alarms and optional wireless communication.

 

However, purchasing a four-gas detector does not by itself make a confined-space entry safe. The instrument must be correctly configured, tested, positioned and used as part of an established entry procedure.

 

What Is a Confined Space?

 

A confined space is not defined only by its physical size. It may be large enough for a person to enter but have restricted entry or exit, unfavourable ventilation or the potential for a hazardous atmosphere.

 

Examples include:

 

Manholes

 

Sewers

 

Storage tanks

 

Process vessels

 

Silos

 

Pits

 

Tunnels

 

Pipelines

 

Ducts

 

Ship cargo holds

 

Engine rooms

 

Underground chambers

 

Wastewater sumps

 

Boiler interiors

 

Atmospheric conditions inside these spaces can change due to decomposition, chemical residues, cleaning work, welding, combustion, leakage, displacement of oxygen or inadequate ventilation.

 

Malaysia Confined-Space Gas Testing Requirements

 

Malaysia’s Department of Occupational Safety and Health states that an Authorised Gas Tester is responsible for conducting atmospheric testing within a confined space, while an Entry Supervisor oversees activities before, during and after entry.

 

DOSH also states that employers must appoint or engage the appropriate competent person to conduct gas testing and supervise the entry process as prescribed in the Industry Code of Practice for Safe Working in a Confined Space 2010.

 

The G40S is a measurement instrument. It does not replace:

 

A competent Authorised Gas Tester

 

An Entry Supervisor

 

An entry permit

 

Risk assessment

 

Isolation and lockout procedures

 

Ventilation

 

Standby personnel

 

Communication

 

Rescue planning

 

Required training

 

Users should follow the current DOSH requirements, their organisation’s safe-work procedures and instructions provided by the responsible competent person.

 

Why Are Four Gases Commonly Monitored?

 

Oxygen – O₂

 

Normal outdoor air contains approximately 20.9% oxygen. Oxygen levels inside a confined space may decrease when oxygen is consumed or displaced by another gas.

 

An oxygen-enriched atmosphere can also create additional combustion risk.

 

A worker should never assume that a space contains sufficient oxygen merely because there is no visible smoke or unusual smell.

 

Combustible Gas – LEL

 

Combustible-gas readings are commonly displayed as a percentage of the Lower Explosive Limit.

 

Possible sources include:

 

Fuel vapour

 

Solvents

 

Methane

 

Hydrocarbons

 

Process leakage

 

Decomposing organic material

 

A combustible atmosphere can become dangerous when an ignition source is introduced. Gas testing should therefore be completed before hot work, electrical work or other activities capable of producing heat or sparks.

 

Carbon Monoxide – CO

 

Carbon monoxide can be produced by incomplete combustion.

 

Potential sources include:

 

Petrol or diesel engines

 

Generators

 

Burners

 

Welding and cutting activity

 

Machinery exhaust

 

Fire or smouldering material

 

CO is colourless and cannot be reliably identified without an appropriate detector.

 

Hydrogen Sulphide – H₂S

 

Hydrogen sulphide may be present in:

 

Sewers

 

Wastewater-treatment areas

 

Sludge tanks

 

Pits

 

Manholes

 

Drainage systems

 

Petroleum operations

 

Spaces containing decomposing organic matter

 

Smell must not be used as a safety indicator. The ability to detect H₂S by odour can become unreliable during exposure.

 

Three Stages of Confined-Space Gas Monitoring

 

Gas monitoring should not be treated as a single reading taken at the entrance. It may be required at several stages.

 

Stage 1: Test Before Opening or Entry

 

The atmosphere should be assessed before a worker enters the space.

 

Depending on the site and risk assessment, this may involve testing from outside the space at different locations or depths.

 

Because the G40S uses natural diffusion, it measures the atmosphere immediately surrounding the detector. It does not automatically draw a remote sample through a long hose.

 

When the space must be tested remotely before entry, a pump-suction gas detector with an appropriate sampling hose or probe may be the better instrument.

 

Stage 2: Test at Different Levels

 

Some gases or vapours may accumulate differently according to their properties, temperature, airflow and the shape of the confined space.

 

Testing only at the opening may fail to represent conditions deeper inside.

 

The responsible gas tester may need to assess:

 

Upper area

 

Middle area

 

Lower area

 

Corners

 

Dead zones

 

Areas close to residue or leakage

 

Locations where work will be performed

 

The detector must be allowed sufficient time to receive and respond to the sample from each location.

 

Stage 3: Continuous Personal Monitoring During Entry

 

Atmospheric conditions can change after entry due to:

 

Work disturbing sludge or residue

 

Equipment releasing exhaust

 

Chemical reaction

 

Welding or cutting

 

Failure of ventilation

 

Gas leakage

 

Changing process conditions

 

Movement to another section of the space

 

A diffusion detector such as the G40S can be worn by the entrant for continuous personal monitoring, subject to the site procedure and instrument configuration.

 

This is where the G40S is particularly useful: it monitors the atmosphere close to the worker while providing audible, visual and vibration alarms.

 

Diffusion vs Pump-Suction Detector

 

RequirementG40S Diffusion DetectorPump-Suction DetectorPersonal monitoring while wornSuitablePossible, but normally largerMeasures air around the userYesYesRemote testing through a hoseNo internal pumpYesTesting before physical entryLimited unless safely lowered according to procedureBetter suitedSampling at a specific depthLimitedSuitable with correct hose and procedureContinuous wearable operationSuitableDepends on modelHose blockage or pump-failure warningNot applicableRequired on suitable pump models

 

A confined-space team may need both instruments: a pump detector for remote testing and a wearable diffusion detector for the entrant.

 

How to Prepare the G40S Before Use

 

1. Confirm the Sensor Configuration

 

Check that the delivered unit contains the sensors required for the job.

 

For common confined-space applications, the purchase order and instrument configuration should clearly state:

 

LEL

 

O₂

 

CO

 

H₂S

 

Do not assume that every unit carrying the G40S model name contains the same sensors. The platform can support different gas combinations.

 

2. Inspect the Detector

 

Check for:

 

Physical damage

 

Cracked casing

 

Blocked sensor openings

 

Contamination

 

Water ingress

 

Loose clips

 

Damaged charging port

 

Low battery

 

Fault indication

 

A damaged or contaminated detector should be assessed before use.

 

3. Switch On in Clean Air

 

Start the detector in an atmosphere known to be clean and safe, following the operating instructions.

 

Allow the instrument to complete:

 

Self-test

 

Sensor warm-up

 

Alarm test

 

Battery check

 

Zero check

 

Do not perform a clean-air zero inside a suspected hazardous area.

 

4. Check Alarm Settings

 

Confirm the required:

 

Low alarm

 

High alarm

 

TWA alarm

 

STEL alarm

 

Oxygen low alarm

 

Oxygen high alarm

 

Alarm thresholds should be established by the responsible safety personnel according to applicable requirements and the worksite risk assessment.

 

5. Perform the Required Bump Test

 

A bump test exposes the sensors to an appropriate known gas mixture to confirm:

 

Sensor response

 

Audible alarm

 

Visual alarm

 

Vibration alarm

 

Display response

 

Alarm activation

 

A detector that powers on is not necessarily responding correctly to gas.

 

If a sensor fails to respond correctly, the instrument should not be relied upon until it has been inspected and calibrated or repaired as appropriate.

 

6. Confirm Calibration Status

 

Calibration compares and adjusts the detector against known reference gas.

 

Check:

 

Last calibration date

 

Due date

 

Calibration record

 

Gas-cylinder expiry date

 

Regulator condition

 

Correct gas concentration

 

Flow arrangement

 

Calibration cap or adaptor

 

Calibration frequency depends on the sensor, manufacturer’s instructions, site policy, use conditions and applicable safety requirements.

 

How Should the G40S Be Worn?

 

For personal monitoring, the detector is generally positioned near the wearer’s breathing zone, such as the upper chest or collar.

 

The sensor openings should remain exposed.

 

Avoid placing the detector:

 

Inside a pocket

 

Under clothing

 

Behind protective equipment

 

Where mud or water can block the openings

 

In a tool bag

 

At a distant location that does not represent the user’s exposure

 

Instrument placement should follow the site procedure and guidance from the responsible safety personnel.

 

What Should the User Do When an Alarm Activates?

 

A gas-detector alarm must never be treated as an inconvenience or silenced without investigation.

 

The worker should follow the site emergency and evacuation procedure. This may include:

 

Stop work

 

Leave the affected area

 

Warn other personnel

 

Inform the standby person or supervisor

 

Prevent unauthorised entry

 

Review ventilation

 

Investigate the source from a safe location

 

Repeat atmospheric testing only under the approved procedure

 

Do not re-enter merely because the alarm has stopped. A falling reading does not by itself confirm that all hazards have been controlled.

 

How the G40S Fall Alarm Helps

 

The G40S includes a fall-detection function that can provide an additional alert if a worker falls or remains in an abnormal position.

 

This may be valuable where:

 

Visibility is poor

 

The entrant works alone within a section

 

Machinery noise makes voice communication difficult

 

Access is vertical

 

The space has several internal compartments

 

The worker could become incapacitated by gas exposure

 

Fall detection is an additional layer of protection, not a replacement for a standby person or rescue system.

 

The response procedure should define who receives the fall alarm and what action must follow.

 

Using G40S Data Records After Work

 

The G40S can store measurement and alarm records for subsequent review.

 

The stored information may help determine:

 

Which gas generated the alarm

 

When the event occurred

 

Peak concentration

 

Duration of exposure

 

Whether conditions changed during the work

 

Whether repeated alarms occurred

 

Whether additional controls are needed

 

Records can support internal safety reviews, but they must be associated with the correct detector, user, date, time and work location.

 

Before deployment, verify that the detector clock is correct and that the organisation can retrieve and understand the exported data.

 

Can G40S GPS Replace the Standby Person?

 

No.

 

Optional GPS or location information may help identify the general position of a detector, but it cannot replace direct confined-space supervision, communication or rescue readiness.

 

GPS performance may also be limited underground, inside tanks, beneath structures or within industrial buildings.

 

Customers requiring 4G or GPS should confirm:

 

Whether the module is installed

 

Mobile-network compatibility

 

Indoor and underground limitations

 

Platform access

 

Alert transmission

 

Data charges

 

Subscription fees

 

Offline data storage

 

Upload behaviour after reconnection

 

Common Mistakes to Avoid

 

Testing Only Once

 

The atmosphere may change after work begins. Continuous monitoring may be required.

 

Assuming One Reading Represents the Entire Space

 

Gas conditions may differ between the upper, middle and lower areas or between compartments.

 

Using a Diffusion Detector as a Remote Pump Detector

 

The G40S does not contain a pump for drawing samples through a long hose.

 

Skipping the Bump Test

 

Battery power and a working display do not prove that the sensors respond to gas.

 

Ignoring Sensor Life

 

Electrochemical and combustible-gas sensors have finite service lives and may be affected by exposure, contamination and storage conditions.

 

Blocking the Sensor Openings

 

Clothing, dirt, water or accessories can delay or prevent gas from reaching the sensors.

 

Entering to Rescue a Collapsed Worker Without Protection

 

An unplanned rescue can create additional victims. Rescue must follow the established procedure using trained personnel and suitable equipment.

 

G40S Confined-Space Checklist

 

Before use:

 

Confirm the required gases

 

Check the measuring ranges

 

Inspect the detector

 

Verify battery level

 

Start in clean air

 

Confirm alarm thresholds

 

Perform the required bump test

 

Verify calibration status

 

Check the date and time

 

Confirm data recording

 

Test communication functions if used

 

Confirm the entry and rescue procedures

 

During use:

 

Keep the detector in the breathing zone

 

Do not obstruct the sensors

 

Observe changing readings

 

Respond immediately to alarms

 

Maintain communication

 

Continue monitoring as required

 

After use:

 

Review alarm records

 

Report abnormal readings

 

Clean the detector correctly

 

Recharge the battery

 

Record the equipment condition

 

Arrange calibration or inspection if required

 

Is the OTYWELL G40S Right for Your Application?

 

The G40S is suitable for organisations seeking a wearable diffusion detector with:

 

Common four-gas monitoring

 

Sound, light and vibration alarms

 

TWA and STEL exposure warnings

 

Fall detection

 

Emergency-assistance functions

 

Measurement and alarm records

 

Computer data upload

 

Optional wireless connectivity

 

For work that requires remote sampling before entry, ask MTM Precision about a suitable pump-suction gas detector instead of assuming the G40S can perform every stage of the testing process.

 

Contact MTM Precision

 

MTM Precision Sdn Bhd

Website: www.mtmpre.com.my

Email: mtmpre@yahoo.com

WhatsApp: 016-660 7346

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Supply and support enquiries are welcome from Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sarawak, Sabah and throughout Malaysia.

 

Safety reference: Malaysia Department of Occupational Safety and Health – Authorised Gas Tester / Entry Supervisor and Industry Code of Practice for Safe Working in a Confined Space 2010.


 

13 Sep 2026