How to Set Low High TWA and STEL Alarms on OTYWELL G40S Malaysia
How to Set Low High TWA and STEL Alarms on OTYWELL G40S Malaysia
The OTYWELL G40S portable gas detector supports several alarm types, including low alarm, high alarm, TWA, STEL, oxygen deficiency and oxygen enrichment alarms.
Correct alarm settings help ensure that the instrument warns workers early enough to follow the site’s emergency procedure.
However, there is no single set of alarm values suitable for every Malaysian workplace.
Alarm settings should be selected according to:
Installed sensor type
Target gas
Applicable exposure limit
Site risk assessment
Confined-space entry procedure
Work activity
Emergency response plan
Detector limitations
Instructions from the responsible safety personnel
Do not change the detector simply to prevent frequent or inconvenient alarms.
The manufacturer describes the G40S as a natural-diffusion portable detector with high, low, TWA and STEL alarm functions. The exact available settings depend on the sensors and firmware installed in the delivered unit. OTYWELL G40S product information
Quick Explanation of G40S Alarm Types
Alarm TypeWhat It IndicatesTypical PurposeLow alarmGas has reached an initial warning levelEarly warning and corrective actionHigh alarmGas has reached a more serious levelImmediate emergency responseTWA alarmAverage exposure has become excessive over a defined periodManaging cumulative toxic-gas exposureSTEL alarmShort-term average exposure has become excessiveManaging brief elevated toxic-gas exposureOxygen-low alarmOxygen has fallen below the approved limitWarning of an oxygen-deficient atmosphereOxygen-high alarmOxygen has risen above the approved limitWarning of increased fire or process riskOver-range alarmGas exceeds the sensor’s measuring capabilityTreat as a serious or unknown concentration
Workers must understand the meaning of every alarm before entering a hazardous work area.
What Is a Low Alarm?
A low alarm is normally the detector’s first warning that a gas concentration has reached a predetermined level.
For toxic or combustible gases, the low alarm usually activates when the reading rises.
For oxygen, the low alarm activates when the oxygen reading falls below the approved minimum.
A low alarm may require the worker to:
Stop the current activity
Warn nearby personnel
Leave the affected area
Check ventilation
Inform the supervisor
Investigate the gas source
Reassess the work permit
Follow the site emergency procedure
The exact response must be decided before work begins.
A low alarm should not be treated as permission to continue working until the high alarm activates.
What Is a High Alarm?
A high alarm indicates that the concentration has reached a more serious preset level.
It should produce a clearly recognisable combination of:
Audible alarm
Visual alarm
Vibration alarm
Gas name or channel indication
Concentration reading
Alarm record
A high alarm commonly requires immediate withdrawal and emergency action according to the site procedure.
Do not remain in the area merely to observe whether the reading will decrease.
What Is TWA?
TWA means Time-Weighted Average.
It represents average exposure over a specified monitoring period.
TWA is mainly relevant to toxic-gas channels such as:
Carbon monoxide
Hydrogen sulphide
Ammonia
Sulphur dioxide
Other occupational toxic gases
A worker may encounter a gas concentration that does not activate the high alarm but remains elevated for a long period. The cumulative exposure may eventually become unacceptable.
The TWA function helps identify this type of exposure.
Simple TWA Example
Consider a worker who experiences:
A lower concentration during the first part of a shift
A higher concentration during cleaning
A lower concentration after ventilation improves
The TWA calculation combines exposure over time instead of judging only the latest reading.
This is why a worker can receive a TWA alarm even when the current concentration has fallen.
The instrument is responding to accumulated exposure history.
What Is STEL?
STEL means Short-Term Exposure Limit.
It addresses average exposure over a shorter defined period.
STEL is designed to identify elevated short-duration exposure that may not be adequately represented by a full-shift average.
It may be important during:
Opening a process line
Disturbing wastewater sludge
Tank cleaning
Chemical transfer
Maintenance near an exhaust source
Sampling a process
Entering a poorly ventilated compartment
Temporary ventilation failure
A brief high exposure should not be dismissed simply because the full-shift TWA remains low.
TWA and STEL Are Not the Same as Instantaneous Alarms
FunctionMeasurement BehaviourLow alarmResponds when the current reading crosses its setpointHigh alarmResponds when the current reading crosses a higher setpointTWAEvaluates accumulated average exposureSTELEvaluates a shorter moving or defined averagePeak readingRecords the highest detected valueOver-rangeIndicates the sensor’s measuring range has been exceeded
A peak reading, high alarm, TWA alarm and STEL alarm may therefore occur at different times.
Why Alarm Values Must Be Job-Specific
A four-gas detector may be used in many different environments:
Storage tanks
Wastewater plants
Palm oil mills
Sewers
Manholes
Factories
Pump rooms
Marine compartments
Petrochemical facilities
Construction sites
Utility chambers
The acceptable alarm configuration can vary according to:
Gas toxicity
Flammability
Exposure duration
Worker health requirements
Required personal protective equipment
Ventilation
Possibility of sudden release
Whether the instrument is for personal or area monitoring
The site’s emergency capabilities
Malaysia’s confined-space framework requires atmospheric hazards to be assessed and managed within a structured entry programme. Relevant responsibilities, testing requirements and permit controls should be checked against the DOSH Malaysia Industry Code of Practice for Safe Working in a Confined Space.
Who Should Approve the Alarm Settings?
Alarm values should be approved by an appropriate responsible or competent person under the organisation’s safety system.
Depending on the workplace, this may involve:
Safety and Health Officer
Industrial hygienist
Authorised Gas Tester
Entry Supervisor
Occupational-health professional
Process-safety engineer
Customer safety department
Instrument administrator
A salesperson, general technician or instrument user should not independently choose exposure limits without considering the approved safety requirements.
Setting the LEL Low and High Alarms
The LEL channel indicates combustible gas as a percentage of the Lower Explosive Limit.
Its alarm settings require careful review because sensor response can depend on:
Sensor technology
Calibration gas
Target gas
Correction factor
Oxygen concentration
Sensor contamination
Inhibiting substances
Environmental conditions
Important Questions Before Setting LEL Alarms
Confirm:
What combustible gas or vapour is expected?
Which gas was used for calibration?
Does the target gas require a correction factor?
Is the LEL sensor catalytic, infrared or another type?
Can the process create oxygen deficiency?
Are substances present that can poison or inhibit the sensor?
What action must occur at the first alarm?
What site limit has been approved?
A low LEL reading does not automatically prove that every possible vapour is absent.
Setting Oxygen Alarms
The oxygen channel usually requires both:
Oxygen-low alarm
Oxygen-high alarm
Clean outdoor air is normally approximately 20.9% oxygen by volume, but the allowable work range must follow the applicable procedure.
Oxygen Deficiency
Oxygen can be reduced by:
Rusting
Oxidation
Biological decomposition
Inert-gas purging
Combustion
Displacement by another gas
Chemical reaction
An oxygen-deficient atmosphere may be dangerous even when the toxic and combustible channels show low readings.
Oxygen Enrichment
Elevated oxygen can increase the ease and intensity of combustion.
Possible causes include:
Oxygen-line leakage
Process malfunction
Incorrect purging
Welding or cutting equipment
Medical or industrial oxygen release
Do not configure only an oxygen-low alarm and ignore possible oxygen enrichment when the process can release oxygen.
Setting CO Alarms
Carbon monoxide may arise from:
Vehicle exhaust
Generators
Engines
Boilers
Incomplete combustion
Hot work
Fires
Process heating
The CO channel may include:
Low instantaneous alarm
High instantaneous alarm
TWA alarm
STEL alarm
All four settings have different purposes.
If diesel- or petrol-powered equipment operates near a confined-space ventilation inlet, CO may enter even though the engine itself remains outside the space.
Setting H₂S Alarms
Hydrogen sulphide may be present in:
Sewers
Wastewater tanks
Sludge
Septic systems
Palm oil mill effluent
Sour petroleum operations
Organic decomposition
Contaminated process water
H₂S concentrations can change rapidly when sludge or residue is disturbed.
The H₂S configuration may require:
Low alarm
High alarm
TWA alarm
STEL alarm
Never rely on smell to determine whether H₂S is present. A worker’s ability to smell the gas cannot be used as a dependable safety indicator.
Why Factory-Default Alarm Values Need Review
Factory defaults can provide an initial instrument configuration, but they may not match:
Malaysian workplace requirements
Customer-specific limits
The exact target gas
A particular confined-space procedure
Corporate safety standards
The intended response plan
Updated occupational exposure limits
Before issuing the detector, record:
Original settings
Approved settings
Person authorising the change
Date of change
Reason for change
Detector serial number
Software or firmware version when relevant
This prevents unexplained differences between instruments.
Do Not Raise an Alarm Setting to Stop Nuisance Alarms
Repeated alarms may indicate:
A real gas hazard
Inadequate ventilation
A leaking process
Incorrect work practice
Cross-sensitivity
Sensor contamination
Incorrect calibration
An unsuitable sensor
An inappropriate original setting
The correct response is to investigate.
Raising the alarm threshold merely to silence the detector may remove an important warning without correcting the hazard.
How to Review the G40S Alarm Configuration
The precise menu sequence depends on the delivered G40S version, firmware and access-control settings.
A controlled review should generally include:
Confirm the detector model and serial number.
Record all installed gas channels.
Review the current low and high settings.
Review TWA and STEL settings for toxic gases.
Review oxygen-low and oxygen-high settings.
Compare the values with the approved site schedule.
Obtain authorisation for any required changes.
Enter the settings using the approved procedure.
Save and recheck every value.
Perform the required functional test.
Record the change.
Inform the assigned user.
Do not guess the password or enter an engineering menu without the appropriate instructions.
Test the Alarm After Changing a Setting
After any alarm-setting change, confirm that the detector operates correctly.
Testing should verify:
Correct gas channel
Correct alarm level
Display indication
Audible alarm
Visual alarm
Vibration alarm
Alarm latching behaviour
Data-record entry
Wireless transmission when enabled
Use suitable test gas and approved test equipment.
Changing a number in the menu does not prove that the complete alarm system works.
Should Every G40S Have Identical Settings?
Not necessarily.
Identical settings may be useful when all detectors:
Contain the same sensors
Serve the same application
Follow the same company procedure
Use the same calibration gas
Are assigned to workers under the same exposure policy
Different settings may be necessary when detectors serve different processes, gases or customer sites.
To prevent confusion, label and manage each configuration clearly.
What Happens to TWA When Workers Share a Detector?
TWA is associated with exposure over time.
If one detector is passed from one worker to another, the displayed TWA history may combine exposure from different individuals.
A controlled shift-handover procedure should define:
Whether TWA data is reset
Who may reset it
When it may be reset
How the previous data is saved
How each worker is identified
How records are assigned
Do not erase exposure information before it has been reviewed and saved where required.
What If the Current Reading Is Low but TWA Is Alarming?
This can happen because TWA reflects earlier accumulated exposure.
The user should:
Follow the alarm response procedure
Leave the affected area when required
Inform the supervisor
Review the exposure history
Check peak and stored readings
Investigate the source
Obtain occupational-safety guidance before resuming work
Do not assume the alarm is false because the current reading has decreased.
What If STEL Keeps Activating?
Repeated STEL alarms may indicate recurring short exposures caused by:
Opening valves
Disturbing sludge
Chemical dosing
Sampling
Cleaning
Poor local exhaust
Intermittent leaks
Ventilation cycling
The work method may need to be redesigned.
Possible controls include:
Isolation
Improved ventilation
Remote operation
Reduced exposure duration
Process modification
Different respiratory protection
Additional area monitoring
The detector warns of exposure; it does not remove the gas.
What Is an Over-Range Alarm?
An over-range condition means the concentration may have exceeded the sensor’s measurement capability.
Treat it seriously.
The actual concentration may be higher than the displayed maximum, and the sensor may require recovery, testing or calibration afterward.
After an over-range event:
Leave the area safely
Follow the emergency procedure
Do not interpret the maximum display as the exact concentration
Allow recovery in verified clean air
Inspect the instrument
Perform the required bump test
Calibrate when required
Arrange technical assessment if readings remain abnormal
Alarm Latching and Reset Behaviour
Some alarm configurations remain active until acknowledged or until readings return to an acceptable range.
Confirm whether the G40S is configured for:
Latching alarm
Non-latching alarm
Manual acknowledgement
Automatic clearing
Silence or mute function
Workers must understand that acknowledging an alarm does not make the atmosphere safe.
The site procedure should clearly state when an alarm may be reset and who is authorised to do so.
Alarm-Setting Control Sheet
ItemInformation to RecordDetector modelOTYWELL G40SSerial numberIndividual unit numberInstalled sensorsExact gases and rangesCalibration gasGas type and concentrationLow alarmApproved value for each channelHigh alarmApproved value for each channelTWA alarmApproved value where applicableSTEL alarmApproved value where applicableOxygen-low alarmApproved site valueOxygen-high alarmApproved site valueAlarm behaviourLatching or non-latchingApproved byResponsible personConfiguration dateDate settings were enteredFunctional testPass or failNext reviewScheduled review date
This record should remain traceable to the individual detector.
Common Alarm-Setting Mistakes
Copying Values from Another Brand
Different detectors may use different sensors, algorithms, ranges and averaging behaviour.
Setting Low and High to the Same Value
This removes the intended staged warning unless specifically required by an approved procedure.
Ignoring TWA and STEL
Instantaneous alarms alone may not control cumulative toxic-gas exposure.
Using One Configuration for Every Worksite
Different processes and customer requirements may need different settings.
Changing Settings Without Documentation
Undocumented changes make incident investigation and instrument control difficult.
Silencing Repeated Alarms
Frequent alarms should trigger investigation, not automatic threshold increases.
Resetting TWA During a Worker’s Shift
This may erase or interrupt the worker’s exposure history.
Ignoring Oxygen Enrichment
Excess oxygen can increase fire risk even when oxygen deficiency is not present.
Trusting the Menu Without a Gas Test
Configuration review must be supported by appropriate functional testing.
G40S Alarm-Setting Checklist
Before changing any value:
Confirm the exact sensors
Obtain the approved alarm schedule
Check applicable exposure requirements
Review the risk assessment
Record the existing settings
Confirm authorisation
After entering the settings:
Recheck every channel
Confirm units of measurement
Test audible, visual and vibration alarms
Apply suitable test gas
Verify data logging
Verify wireless notification when required
Record the new configuration
Brief the detector user
During use:
Respond to every alarm
Do not silence unexplained alarms
Record alarm events
Review TWA and STEL exposure
Stop work when acceptable conditions cannot be maintained
Frequently Asked Questions
Can MTM Provide One Universal Alarm Setting?
No single setting should be promoted as suitable for every workplace.
The customer’s responsible safety personnel must confirm the appropriate values for the gas, activity and site.
Can a User Change the Alarm Values?
Only authorised personnel should change safety-critical settings.
Unauthorised adjustments can create inconsistent protection between detectors.
Does a TWA Alarm Mean the Current Gas Level Is High?
Not necessarily. It may indicate that exposure accumulated earlier during the monitoring period has exceeded the approved average.
Does STEL Mean the Highest Instantaneous Reading?
No. STEL normally represents average exposure during a defined short period, while peak records the highest instantaneous reading detected.
Can TWA and STEL Be Used for LEL?
TWA and STEL are generally associated with occupational exposure to toxic gases. Combustible gas is normally managed using instantaneous LEL alarms and the site’s fire-and-explosion controls.
Should Alarm Settings Be Checked After Calibration?
Yes. Calibration and alarm configuration serve different purposes. Confirm the alarm values and perform the required functional test after service or configuration changes.
Can Wi-Fi or 4G Replace the Local Alarm?
No. Wireless transmission is an additional monitoring function. Signal availability may be unreliable inside tanks, underground chambers or steel structures.
OTYWELL G40S Supplier in Malaysia
MTM Precision supplies portable gas-detection instruments for industrial maintenance, confined-space work and personal gas monitoring.
We can help customers discuss:
G40S sensor configurations
Low- and high-alarm requirements
TWA and STEL functions
Four-gas detector selection
Diffusion and pump-suction models
Data-logging requirements
Fall-alarm functions
Wi-Fi, Bluetooth, 4G and GPS options
Bump-test accessories
Calibration-gas requirements
Please provide the target gases, application, required ranges and certification requirements when requesting a quotation.
Contact MTM Precision
MTM Precision Sdn Bhd
Website: www.mtmpre.com.my
Email: mtmpre@yahoo.com
WhatsApp: 016-660 7346
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.
Supply and support enquiries are welcome from Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sarawak, Sabah and throughout Malaysia.
13 Sep 2026