How to Set Low High TWA and STEL Alarms on OTYWELL G40S Malaysia

How to Set Low High TWA and STEL Alarms on OTYWELL G40S Malaysia

 

The OTYWELL G40S portable gas detector supports several alarm types, including low alarm, high alarm, TWA, STEL, oxygen deficiency and oxygen enrichment alarms.

 

Correct alarm settings help ensure that the instrument warns workers early enough to follow the site’s emergency procedure.

 

However, there is no single set of alarm values suitable for every Malaysian workplace.

 

Alarm settings should be selected according to:

 

Installed sensor type

 

Target gas

 

Applicable exposure limit

 

Site risk assessment

 

Confined-space entry procedure

 

Work activity

 

Emergency response plan

 

Detector limitations

 

Instructions from the responsible safety personnel

 

Do not change the detector simply to prevent frequent or inconvenient alarms.

 

The manufacturer describes the G40S as a natural-diffusion portable detector with high, low, TWA and STEL alarm functions. The exact available settings depend on the sensors and firmware installed in the delivered unit. OTYWELL G40S product information

 

Quick Explanation of G40S Alarm Types

 

Alarm TypeWhat It IndicatesTypical PurposeLow alarmGas has reached an initial warning levelEarly warning and corrective actionHigh alarmGas has reached a more serious levelImmediate emergency responseTWA alarmAverage exposure has become excessive over a defined periodManaging cumulative toxic-gas exposureSTEL alarmShort-term average exposure has become excessiveManaging brief elevated toxic-gas exposureOxygen-low alarmOxygen has fallen below the approved limitWarning of an oxygen-deficient atmosphereOxygen-high alarmOxygen has risen above the approved limitWarning of increased fire or process riskOver-range alarmGas exceeds the sensor’s measuring capabilityTreat as a serious or unknown concentration

 

Workers must understand the meaning of every alarm before entering a hazardous work area.

 

What Is a Low Alarm?

 

A low alarm is normally the detector’s first warning that a gas concentration has reached a predetermined level.

 

For toxic or combustible gases, the low alarm usually activates when the reading rises.

 

For oxygen, the low alarm activates when the oxygen reading falls below the approved minimum.

 

A low alarm may require the worker to:

 

Stop the current activity

 

Warn nearby personnel

 

Leave the affected area

 

Check ventilation

 

Inform the supervisor

 

Investigate the gas source

 

Reassess the work permit

 

Follow the site emergency procedure

 

The exact response must be decided before work begins.

 

A low alarm should not be treated as permission to continue working until the high alarm activates.

 

What Is a High Alarm?

 

A high alarm indicates that the concentration has reached a more serious preset level.

 

It should produce a clearly recognisable combination of:

 

Audible alarm

 

Visual alarm

 

Vibration alarm

 

Gas name or channel indication

 

Concentration reading

 

Alarm record

 

A high alarm commonly requires immediate withdrawal and emergency action according to the site procedure.

 

Do not remain in the area merely to observe whether the reading will decrease.

 

What Is TWA?

 

TWA means Time-Weighted Average.

 

It represents average exposure over a specified monitoring period.

 

TWA is mainly relevant to toxic-gas channels such as:

 

Carbon monoxide

 

Hydrogen sulphide

 

Ammonia

 

Sulphur dioxide

 

Other occupational toxic gases

 

A worker may encounter a gas concentration that does not activate the high alarm but remains elevated for a long period. The cumulative exposure may eventually become unacceptable.

 

The TWA function helps identify this type of exposure.

 

Simple TWA Example

 

Consider a worker who experiences:

 

A lower concentration during the first part of a shift

 

A higher concentration during cleaning

 

A lower concentration after ventilation improves

 

The TWA calculation combines exposure over time instead of judging only the latest reading.

 

This is why a worker can receive a TWA alarm even when the current concentration has fallen.

 

The instrument is responding to accumulated exposure history.

 

What Is STEL?

 

STEL means Short-Term Exposure Limit.

 

It addresses average exposure over a shorter defined period.

 

STEL is designed to identify elevated short-duration exposure that may not be adequately represented by a full-shift average.

 

It may be important during:

 

Opening a process line

 

Disturbing wastewater sludge

 

Tank cleaning

 

Chemical transfer

 

Maintenance near an exhaust source

 

Sampling a process

 

Entering a poorly ventilated compartment

 

Temporary ventilation failure

 

A brief high exposure should not be dismissed simply because the full-shift TWA remains low.

 

TWA and STEL Are Not the Same as Instantaneous Alarms

 

FunctionMeasurement BehaviourLow alarmResponds when the current reading crosses its setpointHigh alarmResponds when the current reading crosses a higher setpointTWAEvaluates accumulated average exposureSTELEvaluates a shorter moving or defined averagePeak readingRecords the highest detected valueOver-rangeIndicates the sensor’s measuring range has been exceeded

 

A peak reading, high alarm, TWA alarm and STEL alarm may therefore occur at different times.

 

Why Alarm Values Must Be Job-Specific

 

A four-gas detector may be used in many different environments:

 

Storage tanks

 

Wastewater plants

 

Palm oil mills

 

Sewers

 

Manholes

 

Factories

 

Pump rooms

 

Marine compartments

 

Petrochemical facilities

 

Construction sites

 

Utility chambers

 

The acceptable alarm configuration can vary according to:

 

Gas toxicity

 

Flammability

 

Exposure duration

 

Worker health requirements

 

Required personal protective equipment

 

Ventilation

 

Possibility of sudden release

 

Whether the instrument is for personal or area monitoring

 

The site’s emergency capabilities

 

Malaysia’s confined-space framework requires atmospheric hazards to be assessed and managed within a structured entry programme. Relevant responsibilities, testing requirements and permit controls should be checked against the DOSH Malaysia Industry Code of Practice for Safe Working in a Confined Space.

 

Who Should Approve the Alarm Settings?

 

Alarm values should be approved by an appropriate responsible or competent person under the organisation’s safety system.

 

Depending on the workplace, this may involve:

 

Safety and Health Officer

 

Industrial hygienist

 

Authorised Gas Tester

 

Entry Supervisor

 

Occupational-health professional

 

Process-safety engineer

 

Customer safety department

 

Instrument administrator

 

A salesperson, general technician or instrument user should not independently choose exposure limits without considering the approved safety requirements.

 

Setting the LEL Low and High Alarms

 

The LEL channel indicates combustible gas as a percentage of the Lower Explosive Limit.

 

Its alarm settings require careful review because sensor response can depend on:

 

Sensor technology

 

Calibration gas

 

Target gas

 

Correction factor

 

Oxygen concentration

 

Sensor contamination

 

Inhibiting substances

 

Environmental conditions

 

Important Questions Before Setting LEL Alarms

 

Confirm:

 

What combustible gas or vapour is expected?

 

Which gas was used for calibration?

 

Does the target gas require a correction factor?

 

Is the LEL sensor catalytic, infrared or another type?

 

Can the process create oxygen deficiency?

 

Are substances present that can poison or inhibit the sensor?

 

What action must occur at the first alarm?

 

What site limit has been approved?

 

A low LEL reading does not automatically prove that every possible vapour is absent.

 

Setting Oxygen Alarms

 

The oxygen channel usually requires both:

 

Oxygen-low alarm

 

Oxygen-high alarm

 

Clean outdoor air is normally approximately 20.9% oxygen by volume, but the allowable work range must follow the applicable procedure.

 

Oxygen Deficiency

 

Oxygen can be reduced by:

 

Rusting

 

Oxidation

 

Biological decomposition

 

Inert-gas purging

 

Combustion

 

Displacement by another gas

 

Chemical reaction

 

An oxygen-deficient atmosphere may be dangerous even when the toxic and combustible channels show low readings.

 

Oxygen Enrichment

 

Elevated oxygen can increase the ease and intensity of combustion.

 

Possible causes include:

 

Oxygen-line leakage

 

Process malfunction

 

Incorrect purging

 

Welding or cutting equipment

 

Medical or industrial oxygen release

 

Do not configure only an oxygen-low alarm and ignore possible oxygen enrichment when the process can release oxygen.

 

Setting CO Alarms

 

Carbon monoxide may arise from:

 

Vehicle exhaust

 

Generators

 

Engines

 

Boilers

 

Incomplete combustion

 

Hot work

 

Fires

 

Process heating

 

The CO channel may include:

 

Low instantaneous alarm

 

High instantaneous alarm

 

TWA alarm

 

STEL alarm

 

All four settings have different purposes.

 

If diesel- or petrol-powered equipment operates near a confined-space ventilation inlet, CO may enter even though the engine itself remains outside the space.

 

Setting H₂S Alarms

 

Hydrogen sulphide may be present in:

 

Sewers

 

Wastewater tanks

 

Sludge

 

Septic systems

 

Palm oil mill effluent

 

Sour petroleum operations

 

Organic decomposition

 

Contaminated process water

 

H₂S concentrations can change rapidly when sludge or residue is disturbed.

 

The H₂S configuration may require:

 

Low alarm

 

High alarm

 

TWA alarm

 

STEL alarm

 

Never rely on smell to determine whether H₂S is present. A worker’s ability to smell the gas cannot be used as a dependable safety indicator.

 

Why Factory-Default Alarm Values Need Review

 

Factory defaults can provide an initial instrument configuration, but they may not match:

 

Malaysian workplace requirements

 

Customer-specific limits

 

The exact target gas

 

A particular confined-space procedure

 

Corporate safety standards

 

The intended response plan

 

Updated occupational exposure limits

 

Before issuing the detector, record:

 

Original settings

 

Approved settings

 

Person authorising the change

 

Date of change

 

Reason for change

 

Detector serial number

 

Software or firmware version when relevant

 

This prevents unexplained differences between instruments.

 

Do Not Raise an Alarm Setting to Stop Nuisance Alarms

 

Repeated alarms may indicate:

 

A real gas hazard

 

Inadequate ventilation

 

A leaking process

 

Incorrect work practice

 

Cross-sensitivity

 

Sensor contamination

 

Incorrect calibration

 

An unsuitable sensor

 

An inappropriate original setting

 

The correct response is to investigate.

 

Raising the alarm threshold merely to silence the detector may remove an important warning without correcting the hazard.

 

How to Review the G40S Alarm Configuration

 

The precise menu sequence depends on the delivered G40S version, firmware and access-control settings.

 

A controlled review should generally include:

 

Confirm the detector model and serial number.

 

Record all installed gas channels.

 

Review the current low and high settings.

 

Review TWA and STEL settings for toxic gases.

 

Review oxygen-low and oxygen-high settings.

 

Compare the values with the approved site schedule.

 

Obtain authorisation for any required changes.

 

Enter the settings using the approved procedure.

 

Save and recheck every value.

 

Perform the required functional test.

 

Record the change.

 

Inform the assigned user.

 

Do not guess the password or enter an engineering menu without the appropriate instructions.

 

Test the Alarm After Changing a Setting

 

After any alarm-setting change, confirm that the detector operates correctly.

 

Testing should verify:

 

Correct gas channel

 

Correct alarm level

 

Display indication

 

Audible alarm

 

Visual alarm

 

Vibration alarm

 

Alarm latching behaviour

 

Data-record entry

 

Wireless transmission when enabled

 

Use suitable test gas and approved test equipment.

 

Changing a number in the menu does not prove that the complete alarm system works.

 

Should Every G40S Have Identical Settings?

 

Not necessarily.

 

Identical settings may be useful when all detectors:

 

Contain the same sensors

 

Serve the same application

 

Follow the same company procedure

 

Use the same calibration gas

 

Are assigned to workers under the same exposure policy

 

Different settings may be necessary when detectors serve different processes, gases or customer sites.

 

To prevent confusion, label and manage each configuration clearly.

 

What Happens to TWA When Workers Share a Detector?

 

TWA is associated with exposure over time.

 

If one detector is passed from one worker to another, the displayed TWA history may combine exposure from different individuals.

 

A controlled shift-handover procedure should define:

 

Whether TWA data is reset

 

Who may reset it

 

When it may be reset

 

How the previous data is saved

 

How each worker is identified

 

How records are assigned

 

Do not erase exposure information before it has been reviewed and saved where required.

 

What If the Current Reading Is Low but TWA Is Alarming?

 

This can happen because TWA reflects earlier accumulated exposure.

 

The user should:

 

Follow the alarm response procedure

 

Leave the affected area when required

 

Inform the supervisor

 

Review the exposure history

 

Check peak and stored readings

 

Investigate the source

 

Obtain occupational-safety guidance before resuming work

 

Do not assume the alarm is false because the current reading has decreased.

 

What If STEL Keeps Activating?

 

Repeated STEL alarms may indicate recurring short exposures caused by:

 

Opening valves

 

Disturbing sludge

 

Chemical dosing

 

Sampling

 

Cleaning

 

Poor local exhaust

 

Intermittent leaks

 

Ventilation cycling

 

The work method may need to be redesigned.

 

Possible controls include:

 

Isolation

 

Improved ventilation

 

Remote operation

 

Reduced exposure duration

 

Process modification

 

Different respiratory protection

 

Additional area monitoring

 

The detector warns of exposure; it does not remove the gas.

 

What Is an Over-Range Alarm?

 

An over-range condition means the concentration may have exceeded the sensor’s measurement capability.

 

Treat it seriously.

 

The actual concentration may be higher than the displayed maximum, and the sensor may require recovery, testing or calibration afterward.

 

After an over-range event:

 

Leave the area safely

 

Follow the emergency procedure

 

Do not interpret the maximum display as the exact concentration

 

Allow recovery in verified clean air

 

Inspect the instrument

 

Perform the required bump test

 

Calibrate when required

 

Arrange technical assessment if readings remain abnormal

 

Alarm Latching and Reset Behaviour

 

Some alarm configurations remain active until acknowledged or until readings return to an acceptable range.

 

Confirm whether the G40S is configured for:

 

Latching alarm

 

Non-latching alarm

 

Manual acknowledgement

 

Automatic clearing

 

Silence or mute function

 

Workers must understand that acknowledging an alarm does not make the atmosphere safe.

 

The site procedure should clearly state when an alarm may be reset and who is authorised to do so.

 

Alarm-Setting Control Sheet

 

ItemInformation to RecordDetector modelOTYWELL G40SSerial numberIndividual unit numberInstalled sensorsExact gases and rangesCalibration gasGas type and concentrationLow alarmApproved value for each channelHigh alarmApproved value for each channelTWA alarmApproved value where applicableSTEL alarmApproved value where applicableOxygen-low alarmApproved site valueOxygen-high alarmApproved site valueAlarm behaviourLatching or non-latchingApproved byResponsible personConfiguration dateDate settings were enteredFunctional testPass or failNext reviewScheduled review date

 

This record should remain traceable to the individual detector.

 

Common Alarm-Setting Mistakes

 

Copying Values from Another Brand

 

Different detectors may use different sensors, algorithms, ranges and averaging behaviour.

 

Setting Low and High to the Same Value

 

This removes the intended staged warning unless specifically required by an approved procedure.

 

Ignoring TWA and STEL

 

Instantaneous alarms alone may not control cumulative toxic-gas exposure.

 

Using One Configuration for Every Worksite

 

Different processes and customer requirements may need different settings.

 

Changing Settings Without Documentation

 

Undocumented changes make incident investigation and instrument control difficult.

 

Silencing Repeated Alarms

 

Frequent alarms should trigger investigation, not automatic threshold increases.

 

Resetting TWA During a Worker’s Shift

 

This may erase or interrupt the worker’s exposure history.

 

Ignoring Oxygen Enrichment

 

Excess oxygen can increase fire risk even when oxygen deficiency is not present.

 

Trusting the Menu Without a Gas Test

 

Configuration review must be supported by appropriate functional testing.

 

G40S Alarm-Setting Checklist

 

Before changing any value:

 

Confirm the exact sensors

 

Obtain the approved alarm schedule

 

Check applicable exposure requirements

 

Review the risk assessment

 

Record the existing settings

 

Confirm authorisation

 

After entering the settings:

 

Recheck every channel

 

Confirm units of measurement

 

Test audible, visual and vibration alarms

 

Apply suitable test gas

 

Verify data logging

 

Verify wireless notification when required

 

Record the new configuration

 

Brief the detector user

 

During use:

 

Respond to every alarm

 

Do not silence unexplained alarms

 

Record alarm events

 

Review TWA and STEL exposure

 

Stop work when acceptable conditions cannot be maintained

 

Frequently Asked Questions

 

Can MTM Provide One Universal Alarm Setting?

 

No single setting should be promoted as suitable for every workplace.

 

The customer’s responsible safety personnel must confirm the appropriate values for the gas, activity and site.

 

Can a User Change the Alarm Values?

 

Only authorised personnel should change safety-critical settings.

 

Unauthorised adjustments can create inconsistent protection between detectors.

 

Does a TWA Alarm Mean the Current Gas Level Is High?

 

Not necessarily. It may indicate that exposure accumulated earlier during the monitoring period has exceeded the approved average.

 

Does STEL Mean the Highest Instantaneous Reading?

 

No. STEL normally represents average exposure during a defined short period, while peak records the highest instantaneous reading detected.

 

Can TWA and STEL Be Used for LEL?

 

TWA and STEL are generally associated with occupational exposure to toxic gases. Combustible gas is normally managed using instantaneous LEL alarms and the site’s fire-and-explosion controls.

 

Should Alarm Settings Be Checked After Calibration?

 

Yes. Calibration and alarm configuration serve different purposes. Confirm the alarm values and perform the required functional test after service or configuration changes.

 

Can Wi-Fi or 4G Replace the Local Alarm?

 

No. Wireless transmission is an additional monitoring function. Signal availability may be unreliable inside tanks, underground chambers or steel structures.

 

OTYWELL G40S Supplier in Malaysia

 

MTM Precision supplies portable gas-detection instruments for industrial maintenance, confined-space work and personal gas monitoring.

 

We can help customers discuss:

 

G40S sensor configurations

 

Low- and high-alarm requirements

 

TWA and STEL functions

 

Four-gas detector selection

 

Diffusion and pump-suction models

 

Data-logging requirements

 

Fall-alarm functions

 

Wi-Fi, Bluetooth, 4G and GPS options

 

Bump-test accessories

 

Calibration-gas requirements

 

Please provide the target gases, application, required ranges and certification requirements when requesting a quotation.

 

Contact MTM Precision

 