How to Test a Used Total Station Before Payment Malaysia

A used total station should be tested before payment whenever possible. Turning on the instrument and displaying an angle is not enough to confirm that it is suitable for surveying or construction work.

A proper inspection should cover the physical condition, optics, EDM, angles, compensator, plummet, battery, memory and data transfer.

## Bring the Correct Test Equipment

Prepare:

- Stable tripod
- Known serviceable prism
- Prism pole
- Charged battery
- Compatible charger
- Suitable data cable
- Laptop with required software, if available
- Measuring tape
- Known reference points or distances
- Field notes for recording results

If the seller provides all accessories, test the supplied items rather than relying only on your own equipment.

## Step 1: Confirm the Model and Serial Number

Check:

- Complete model name
- Serial number
- Angle-accuracy version
- Reflectorless capability
- Communication-port type
- Battery model
- Charger model
- Included accessories

Compare the serial number on the instrument with any calibration or service documents provided.

## Step 2: Inspect the Instrument Body

Look for:

- Cracks
- Impact marks
- Loose handle
- Missing covers
- Damaged display
- Worn keyboard
- Corroded battery contacts
- Loose communication port
- Signs of water entry
- Evidence of previous opening or repair

Cosmetic wear may be acceptable, but impact damage can indicate a larger internal problem.

## Step 3: Check the Tribrach and Foot Screws

Mount the total station on the tripod.

Confirm that:

- The instrument locks securely
- All three foot screws move smoothly
- The tribrach does not shift
- The circular level responds normally
- There is no excessive mechanical play

A loose tribrach can affect centring, levelling and angle observations.

## Step 4: Inspect the Telescope

Look through the telescope and check:

- Image clarity
- Eyepiece adjustment
- Smooth focusing
- Crosshair condition
- Internal haze or fungus
- Telescope rotation
- Vertical and horizontal fine motion
- Clamping action

Focus on both near and distant targets. The image should remain usable throughout the focusing range.

## Step 5: Test the Plummet

Set up over a clearly marked point.

Rotate the instrument through different positions and observe whether the optical or laser plummet remains centred.

If the plummet moves significantly around the point, it may require adjustment.

A plummet error affects the actual station position.

## Step 6: Test the Electronic Level and Compensator

Level the instrument carefully.

Tilt it slightly within a safe range and observe:

- Electronic-level response
- Compensator response
- Tilt warning
- Return to a stable reading
- Repeatability after re-levelling

The compensator should not behave erratically or remain stuck.

## Step 7: Test Angle Measurement

Observe a clear target in Face Left and Face Right positions.

Check:

- Horizontal-angle consistency
- Vertical-angle consistency
- Collimation
- Vertical-index behaviour
- Smooth circle rotation
- Stable fine adjustment

A formal accuracy assessment requires the correct procedure and reference equipment, but a face comparison may reveal an obvious problem.

## Step 8: Test Prism Distance Measurement

Measure a known serviceable prism at more than one distance.

Confirm:

- Measurement starts normally
- Results repeat consistently
- No unusual delay occurs
- No EDM error appears
- The selected prism constant is correct
- Fine and tracking modes work, where applicable

Repeat several measurements without moving the prism.

Unstable or intermittent results require further investigation.

## Step 9: Test Reflectorless Measurement

If the model supports reflectorless EDM, test it separately on suitable non-reflective targets.

Use several surfaces and distances.

Do not judge reflectorless performance using:

- Glass
- Highly reflective metal
- Moving targets
- Wet or very dark surfaces
- Targets at an extreme angle

If prism mode works but reflectorless mode repeatedly fails on suitable targets, the EDM may require inspection.

## Step 10: Check the Battery

Leave the instrument operating while other tests are performed.

Watch for:

- Rapid battery-level drop
- Unexpected shutdown
- Restart during EDM measurement
- Loose battery contact
- Charging problem
- Swollen or damaged battery casing

An old battery may require replacement even when the instrument itself is serviceable.

## Step 11: Test the Display and Keyboard

Press every key and check all display areas.

Confirm that:

- Buttons respond
- Display segments are complete
- Backlight works
- Touchscreen responds, if fitted
- Menus can be accessed
- Job names can be entered
- Points can be stored
- Stored data can be reviewed

For dual-display instruments, test both sides.

## Step 12: Test Memory and Programs

Create a small test job and enter several points.

Try:

- Coordinate input
- Point storage
- Station setup
- Backsight entry
- Setting-out function
- Job deletion
- Data review

Available functions depend on the model. Confirm that the programs required by your team are present and working.

## Step 13: Test Data Transfer

Connect the instrument to a computer using the correct cable and software.

Verify:

- Cable compatibility
- Driver installation
- COM-port assignment
- Baud rate and communication settings
- Download from the instrument
- Upload to the instrument
- File format
- Coordinate order

A successful cable connection does not guarantee that the exported data will work correctly with your survey software.

## Step 14: Inspect the Accessories

Check the condition and compatibility of:

- Battery
- Charger
- Carrying case
- Tribrach
- Data cable
- Prism
- Prism pole
- Tripod
- Mini prism
- Memory card or USB device, where applicable

Missing accessories should be included when calculating the actual purchase cost.

## Step 15: Review Calibration and Repair History

Ask for:

- Latest calibration record
- Previous repair details
- Reason for sale
- Known faults
- Battery replacement history
- EDM repair history
- Evidence of water or impact damage

A previous repair is not automatically a problem, but the current performance should be verified.

## Warning Signs Before Payment

Pause the purchase when:

- The seller refuses measurement testing
- EDM cannot produce stable results
- Compensator behaves abnormally
- Serial number is missing or unclear
- Data cannot be stored or exported
- Battery shuts down quickly
- Calibration status is unknown
- The instrument has visible water damage
- The price does not account for required repair

Agree clearly whether the unit is sold as working, partially working or for parts.

## Consider a Pre-Purchase Inspection

An independent check can help determine whether the instrument is appropriate for its intended work.

Checking establishes its current condition. Calibration evaluates and adjusts measurement performance where appropriate. Repair addresses mechanical, optical or electronic faults.

MTM Precision provides pre-purchase checking, calibration and repair assessment for used Topcon, Sokkia, Nikon, Leica, Trimble and Pentax total stations.

Customers from Selangor, Kuala Lumpur, Johor and Penang may contact us with the model, serial number, photos and seller’s description before arranging an inspection at our Puchong Service Centre.

**MTM Precision Sdn Bhd**  
**Showroom & Service Centre:** No. 29-1 & 29-2, Jalan Bandar 18, Pusat Bandar Puchong, 47160 Puchong, Selangor, Malaysia  
**Tel:** 03-8080 7172  
**WhatsApp:** +6016-660 7346  
**Email:** mtmpre@yahoo.com  
**Website:** www.mtmpre.com.my

23 Sep 2026