Temtop S1-UP PM2.5 Reading Too High or Keeps Changing Troubleshooting Guide Malaysia

Temtop S1-UP PM2.5 Reading Too High or Keeps Changing Troubleshooting Guide Malaysia

 

Your Temtop S1-UP is showing a high PM2.5 reading.

 

The number keeps increasing and decreasing.

 

The reading changes after moving the monitor.

 

It does not match your air purifier.

 

Two PM2.5 monitors show different numbers.

 

Does this mean the S1-UP is faulty?

 

Not necessarily.

 

PM2.5 conditions can change quickly because of:

 

Outdoor haze.

 

Cooking.

 

Smoking.

 

Burning incense.

 

Cleaning.

 

Open windows.

 

Door movement.

 

Air purifiers.

 

Air conditioning.

 

Monitor placement.

 

Before deciding that the instrument requires inspection or repair, check the environment, operating mode and measurement position.

 

This guide explains the common reasons for unstable or unexpected S1-UP readings and the safe checks you can perform.

 

What Does the Temtop S1-UP Measure?

 

The S1-UP displays:

 

PM2.5

 

AQI

 

Temperature

 

Relative humidity

 

It is designed for general air-quality monitoring in locations such as:

 

Homes.

 

Bedrooms.

 

Offices.

 

Classrooms.

 

Clinics.

 

Shops.

 

Meeting rooms.

 

The S1-UP does not measure:

 

CO2

 

PM10

 

TVOC

 

Formaldehyde or HCHO

 

Carbon monoxide

 

Combustible gas

 

Make sure the displayed parameter is being interpreted correctly before troubleshooting the monitor.

 

Problem 1: PM2.5 Reading Is Higher Than Expected

 

A room may look clean but still contain airborne fine particles.

 

Possible sources include:

 

Malaysian haze.

 

Road traffic.

 

Construction activity.

 

Cooking smoke.

 

Cigarette smoke.

 

Burning incense.

 

Candles.

 

Cleaning.

 

Dust disturbance.

 

Particles entering when doors open.

 

Air flowing from another room.

 

A high reading does not automatically indicate a faulty sensor.

 

First investigate what is happening around the monitor.

 

Check for Outdoor Haze

 

Outdoor PM2.5 can enter through:

 

Open windows.

 

Open external doors.

 

Balcony doors.

 

Ventilation openings.

 

Gaps and leakage points.

 

Frequent movement between indoor and outdoor areas.

 

During haze, close or record the condition of the openings and observe whether the reading changes over time.

 

Do not assume that indoor PM2.5 will immediately become low after closing the windows.

 

Particles already inside the room may require time to settle or be filtered.

 

Check for Cooking Activity

 

Cooking can produce airborne particles.

 

PM2.5 may travel from the kitchen into:

 

Living rooms.

 

Dining areas.

 

Bedrooms.

 

Home offices.

 

Other connected spaces.

 

The reading may remain high even after cooking stops because particles are still suspended or moving through the home.

 

Do not position the S1-UP directly above a stove or expose it to heavy smoke, oil, heat or steam.

 

Check for Smoking, Incense or Candles

 

Cigarette smoke, incense and candles may produce particles that cause the reading to increase.

 

Particles can travel beyond the immediate source.

 

Even if the activity occurs in another room, the S1-UP may respond when the air moves through the building.

 

The monitor can show particle changes, but it does not analyse every gas or chemical contained in smoke.

 

Check for Cleaning and Dust Disturbance

 

Cleaning may temporarily increase airborne particles.

 

Possible activities include:

 

Sweeping.

 

Dusting.

 

Vacuuming.

 

Changing bedding.

 

Cleaning curtains.

 

Moving furniture.

 

Handling dusty boxes.

 

Allow the room condition to settle before deciding that the monitor is reading incorrectly.

 

Problem 2: PM2.5 Reading Keeps Going Up and Down

 

A changing PM2.5 reading can be normal.

 

Air is continuously moving.

 

Particle sources may start and stop.

 

The purifier may remove particles while new particles enter.

 

Possible causes include:

 

The door opening and closing.

 

The window being opened.

 

Air-conditioner airflow.

 

Purifier airflow.

 

People walking nearby.

 

Cooking activity.

 

Changing outdoor haze.

 

Cleaning.

 

Moving the monitor.

 

A fluctuating reading does not automatically mean the S1-UP is unstable or damaged.

 

Look at the Trend, Not Only One Number

 

Instead of focusing on every small change, ask:

 

Is the overall reading increasing?

 

Is it decreasing?

 

Does it rise when the door opens?

 

Does it fall after the purifier starts?

 

Does it increase during cooking?

 

Does it stabilise when the room remains closed?

 

The pattern can be more useful than one momentary number.

 

Problem 3: Reading Changes After Moving the S1-UP

 

This is expected.

 

When the S1-UP is moved, it may experience changes in:

 

Particle concentration.

 

Temperature.

 

Humidity.

 

Air movement.

 

Distance from windows.

 

Distance from the purifier.

 

Distance from pollution sources.

 

Allow the monitor time to respond to the new location.

 

Do not compare the first reading immediately after movement with a stable reading from the previous room.

 

How to Compare Two Rooms Properly

 

When comparing rooms:

 

Use a similar monitor height.

 

Avoid direct airflow.

 

Keep away from windows and purifier outlets.

 

Allow the reading to settle.

 

Record the room conditions.

 

Record door and window positions.

 

Record occupancy.

 

Record whether the air conditioner and purifier are operating.

 

Without this context, differences may be difficult to explain.

 

Problem 4: S1-UP Shows a Different Reading from the Air Purifier

 

This is a common situation.

 

The purifier and S1-UP may use:

 

Different particle sensors.

 

Different internal algorithms.

 

Different sampling methods.

 

Different update intervals.

 

Different measurement positions.

 

The purifier sensor is normally installed inside or near the purifier.

 

It may sample air close to the purifier’s intake.

 

The S1-UP may be positioned somewhere else in the room.

 

Therefore, the two instruments are not necessarily measuring identical air at the same moment.

 

How to Make a Better Comparison

 

Place the S1-UP in a stable position away from the purifier outlet.

 

Allow both instruments to operate under steady room conditions.

 

Avoid:

 

Cooking.

 

Smoking.

 

Cleaning.

 

Opening windows.

 

Moving the devices.

 

Repeated door movement.

 

Observe whether both instruments generally show the same direction of change.

 

For example:

 

Do both readings increase during cooking?

 

Do both decrease after the purifier operates?

 

A difference in the exact number does not automatically prove that one instrument is faulty.

 

Problem 5: S1-UP Reading Does Not Decrease After the Purifier Starts

 

Possible causes include:

 

The purifier is too small for the room.

 

The fan speed is too low.

 

The filter is dirty.

 

The filter is incorrectly installed.

 

Protective packaging remains on the filter.

 

The purifier intake or outlet is blocked.

 

Windows or doors are open.

 

Outdoor haze keeps entering.

 

Cooking or smoking continues.

 

The monitor is near a pollution source.

 

The room connects to a larger space.

 

The purifier may be working while new particles continuously enter the room.

 

Check the Purifier Setup

 

Confirm:

 

The filter is installed correctly.

 

All filter packaging has been removed.

 

The air intake is not blocked.

 

The clean-air outlet is not blocked.

 

The purifier has sufficient clearance.

 

The selected speed is suitable.

 

The purifier is appropriate for the room size.

 

Follow the purifier manufacturer’s operating and maintenance instructions.

 

The S1-UP can show the room PM2.5 trend, but it cannot diagnose the purifier’s internal components.

 

Problem 6: Reading Is Very Low Beside the Purifier

 

The S1-UP may be positioned directly in the purifier’s clean-air outlet.

 

This can produce a low local reading that does not represent the entire room.

 

Move the S1-UP to a representative occupied area.

 

Do not place it:

 

On top of the purifier.

 

Directly in front of the outlet.

 

Immediately beside the intake.

 

The objective is to monitor the room, not only the purified air stream.

 

Problem 7: Temperature Reading Looks Too Low

 

Check whether the S1-UP is positioned:

 

Under an air-conditioner outlet.

 

Beside a fan.

 

Against a cold wall.

 

Near an open window.

 

Directly in a strong air stream.

 

A monitor placed under the air conditioner may show a local temperature rather than the average occupied-room condition.

 

Move it to a more representative position and allow the reading to stabilise.

 

Problem 8: Temperature Reading Looks Too High

 

Check whether the monitor is:

 

In direct sunlight.

 

Near a window heated by sunlight.

 

Beside computer equipment.

 

Near a lamp.

 

Close to hot machinery.

 

Inside a poorly ventilated enclosure.

 

Near a kitchen.

 

Local heat can affect the displayed temperature.

 

Move the instrument away from the heat source and observe the reading again.

 

Problem 9: Humidity Reading Looks Too High

 

Possible causes include:

 

The monitor is near a bathroom.

 

A humidifier is operating nearby.

 

Clothes are drying indoors.

 

The room has poor moisture control.

 

Wet weather is affecting the building.

 

The monitor is near steam.

 

The instrument has been moved from a different environment.

 

Place the S1-UP away from direct moisture sources and allow it to stabilise.

 

One humidity reading does not identify the cause of a building moisture problem.

 

Problem 10: Humidity Reading Looks Too Low

 

Possible causes include:

 

Direct air-conditioner airflow.

 

A strongly air-conditioned room.

 

The monitor is near a dry-air outlet.

 

The monitor has recently been moved.

 

Compare the reading in a representative location away from the air-conditioning vent.

 

Problem 11: Two S1-UP Units Show Different Readings

 

Two units may not show exactly the same number at every moment.

 

Possible reasons include:

 

Slightly different positions.

 

Different airflow.

 

Different operating modes.

 

Different response timing.

 

Normal sensor variation.

 

One unit being closer to a particle source.

 

Different battery or operating conditions.

 

Even small differences in location can matter, especially near:

 

Purifier outlets.

 

Windows.

 

Doors.

 

Air conditioners.

 

Cooking areas.

 

How to Compare Two Units

 

Place both units:

 

Close together.

 

At the same height.

 

Away from direct airflow.

 

Away from windows and doors.

 

Away from purifier outlets.

 

In the same operating mode.

 

Allow both units to stabilise.

 

Observe the trend over a reasonable period.

 

Do not compare one monitor in ECO Mode with another in a more active mode without considering the different measurement behaviour.

 

Problem 12: S1-UP Does Not Match an Outdoor AQI App

 

Outdoor AQI and the S1-UP are not necessarily measuring the same environment.

 

An outdoor monitoring station may be:

 

Several kilometres away.

 

At a different height.

 

Exposed to different traffic.

 

Affected by different wind conditions.

 

Located in an outdoor environment.

 

The S1-UP measures the air around the instrument inside your room.

 

Therefore:

 

Outdoor AQI and indoor S1-UP readings should not be expected to match exactly.

 

The indoor reading is useful because it reflects the specific room being monitored.

 

Problem 13: Reading Changes When Air Conditioning Starts

 

The air conditioner changes:

 

Air movement.

 

Temperature.

 

Humidity.

 

Particle circulation.

 

The airflow may move particles toward or away from the monitor.

 

The reading may also change if the S1-UP is directly under the outlet.

 

For a meaningful comparison:

 

Position the S1-UP away from direct airflow.

 

Keep the location fixed.

 

Record when the air conditioner starts.

 

Observe the trend over time.

 

Problem 14: Reading Changes When the Door Opens

 

Opening the door connects the room to another air environment.

 

Air entering may contain more or fewer particles.

 

The effect depends on:

 

Conditions outside the room.

 

How long the door remains open.

 

Air pressure.

 

Air movement.

 

Nearby activities.

 

Outdoor haze.

 

Kitchen smoke.

 

Repeated door opening may explain fluctuating readings in:

 

Shops.

 

Clinics.

 

Offices.

 

Classrooms.

 

Bedrooms.

 

Problem 15: Reading Changes When the Window Opens

 

Opening the window may:

 

Introduce outdoor PM2.5.

 

Change air movement.

 

Change temperature.

 

Change humidity.

 

During haze or heavy traffic, PM2.5 may increase.

 

During cleaner outdoor conditions, the response may be different.

 

Record the window condition whenever comparing readings.

 

Problem 16: S1-UP Does Not Show CO2

 

This is not a fault.

 

The S1-UP is designed to display:

 

PM2.5.

 

AQI.

 

Temperature.

 

Humidity.

 

It does not include CO2 measurement.

 

If you require bedroom, meeting-room or classroom CO2 monitoring, choose a suitable CO2-capable model.

 

Do not treat PM2.5 as a substitute for CO2.

 

Problem 17: S1-UP Does Not Show PM10

 

This is also not a fault.

 

The S1-UP focuses on PM2.5.

 

Customers requiring PM10 should select an instrument with dedicated PM10 capability.

 

Always verify the measurement parameters before purchasing.

 

Problem 18: S1-UP Does Not Connect to an App

 

The Temtop S1-UP is a standalone model without the Bluetooth App connectivity offered by the S1+.

 

Failure to find an App connection is not a product fault.

 

Choose the Temtop S1+ when Bluetooth App connectivity is required.

 

Choose the S1-UP when direct on-screen monitoring is sufficient.

 

Problem 19: ECO Mode Responds Differently from A Mode

 

The S1-UP provides two modes.

 

ECO Mode

 

ECO Mode prioritises reduced power consumption and longer standby operation.

 

According to Temtop, standby duration may reach approximately 60 days, depending on use and battery condition.

 

A Mode

 

A Mode is intended for more immediate data.

 

According to Temtop, standby duration may reach approximately 20 days, depending on use and battery condition.

 

For active checking of changes after starting an air purifier or opening a window, A Mode may be more suitable.

 

For longer general observation, ECO Mode may be preferred.

 

Different mode behaviour should not automatically be interpreted as a sensor problem.

 

Problem 20: The Reading Appears Frozen

 

First check whether:

 

The operating mode affects the update behaviour.

 

The surrounding air is stable.

 

The monitor has sufficient battery power.

 

The display responds to other changes.

 

The instrument has been positioned correctly.

 

Move the monitor only if necessary and follow the manufacturer’s operating instructions.

 

Do not expose it directly to smoke, steam, water or excessive dust in an attempt to force a reading change.

 

Safe Basic Troubleshooting Procedure

 

Step 1: Check the Measurement Requirement

 

Confirm that you need PM2.5, AQI, temperature and humidity.

 

The S1-UP does not measure CO2, PM10, TVOC or HCHO.

 

Step 2: Check the Position

 

Move the monitor away from:

 

Purifier outlets.

 

Air-conditioner outlets.

 

Open windows.

 

External doors.

 

Steam.

 

Direct sunlight.

 

Hot equipment.

 

Step 3: Stabilise the Environment

 

Avoid cooking, smoking, cleaning and unnecessary door movement.

 

Step 4: Select the Appropriate Mode

 

Use A Mode for more immediate active checking.

 

Use ECO Mode when longer battery endurance is the priority.

 

Step 5: Observe the Trend

 

Record the readings over time rather than relying on one number.

 

Step 6: Compare Under Controlled Conditions

 

If comparing with another monitor, place both instruments close together under the same conditions.

 

Step 7: Document the Symptom

 

If the problem continues, record:

 

Instrument model.

 

Operating mode.

 

Displayed reading.

 

Location.

 

Room activity.

 

Photos or video.

 

Comparison conditions.

 

When the symptom occurs.

 

Good information makes initial assessment easier.

 

When Should You Request Inspection?

 

Consider requesting further assessment when:

 

The display does not operate normally.

 

The instrument does not respond after basic checks.

 

The reading remains clearly abnormal under controlled conditions.

 

Physical damage is visible.

 

The monitor has been exposed to water.

 

The battery or charging behaviour appears abnormal.

 

The unit behaves very differently from another comparable instrument under the same controlled conditions.

 

Do not open or modify the instrument unless authorised.

 

Unauthorised disassembly may cause additional damage.

 

Checking vs Calibration vs Repair

 

These terms should not be treated as identical.

 

Checking

 

A basic comparison helps determine whether the instrument responds reasonably under controlled conditions.

 

Calibration

 

Calibration compares the instrument against an appropriate reference using a defined procedure.

 

It establishes measurement performance under the calibration conditions.

 

Calibration does not automatically repair a damaged sensor.

 

Repair

 

Repair addresses an identified hardware or functional problem.

 

An inspection may be required before confirming whether repair is practical.

 

The correct next step depends on the symptom and application.

 

Information to Send for Initial Assessment

 

When contacting MTM Precision, send:

 

A clear photo of the instrument

 

A photo of the model label

 

The displayed reading

 

The operating mode

 

A description of the room

 

A photo showing the monitor position

 

The comparison instrument, if any

 

A short video if the reading freezes or changes abnormally

 

Also explain whether:

 

The air purifier is operating.

 

The air conditioner is operating.

 

Windows or doors are open.

 

Cooking or cleaning is taking place.

 

The problem occurs in every room.

 

This helps separate possible instrument issues from environmental and placement effects.

 

Temtop S1-UP Troubleshooting Summary

 

If the PM2.5 reading is too high, check:

 

Outdoor haze.

 

Cooking.

 

Smoking.

 

Incense.

 

Cleaning.

 

Open windows and doors.

 

Nearby construction.

 

Monitor placement.

 

If the reading keeps changing, check:

 

Air movement.

 

Purifier operation.

 

Air conditioning.

 

Door movement.

 

Window condition.

 

Indoor particle sources.

 

Operating mode.

 

If two instruments disagree:

 

Place them together.

 

Use the same conditions.

 

Avoid direct airflow.

 

Allow the readings to stabilise.

 

Compare the trend.

 

Most importantly:

 

A changing PM2.5 reading does not automatically mean the S1-UP is faulty.

 

Fine-particle conditions can change continuously.

 

Perform basic controlled checks before deciding whether inspection, calibration or repair is required.

 

Contact MTM Precision

 

For initial assessment, WhatsApp MTM Precision with a clear instrument photo, model label, displayed reading and description of the visible symptom.

 

MTM Precision Sdn Bhd

 

Showroom & Service Centre:

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia

 

Tel: 03-8080 7172

WhatsApp: +6016-660 7346

Email: mtmpre@yahoo.com

Website: www.mtmpre.com.my

03 Oct 2026