How to Check Power Supply Ripple with FNIRSI DSO-TC3 Plus Malaysia

How to Check Power Supply Ripple with FNIRSI DSO-TC3 Plus Malaysia

A digital multimeter shows that a power supply output is approximately 5 V, 12 V or 24 V.

Does that mean the power supply output is clean and stable?

Not necessarily.

A DC power rail can show approximately the correct average voltage while still containing unwanted voltage variation known as ripple.

For electronics technicians and PCB repair users in Malaysia, an oscilloscope can provide information about a power supply that a simple DC voltage reading may not reveal.

The FNIRSI DSO-TC3 Plus can support basic waveform investigation on suitable low-voltage circuits within its measurement capability.

What Is Power Supply Ripple?

Ideally, a DC power supply output would remain perfectly constant.

In a real circuit, the voltage may contain small variations.

Instead of:

Perfect Flat DC

the oscilloscope may show:

DC + Repeating Voltage Variation

This unwanted variation is commonly referred to as ripple.

Where Does Ripple Come From?

Ripple can be associated with several parts of a power supply.

Possible contributors include:

Rectification

Switching operation

Output filtering

Capacitors

Inductors

Load changes

Control-loop behaviour

Some ripple is normal.

The important question is whether the amount and waveform are appropriate for the circuit.

Why Can a Multimeter Miss Ripple?

A digital multimeter is excellent for checking the DC voltage level.

For example:

12.0 V

But that single number does not necessarily show everything happening over time.

The actual voltage may be varying around the nominal DC level.

An oscilloscope allows you to observe this changing component as a waveform.

This is why a multimeter and oscilloscope complement each other.

Symptoms of Excessive Ripple

Depending on the equipment, excessive power supply ripple may contribute to symptoms such as:

Equipment resetting

Unstable operation

Audio noise

Display problems

Communication errors

Sensor instability

Power supply overheating

Intermittent faults

However, these symptoms can have many other causes.

Ripple measurement is part of the diagnosis, not proof of one specific fault.

Capacitors and Ripple

Output capacitors are important parts of many power supplies.

They help smooth the output and reduce unwanted voltage variation.

As an electrolytic capacitor ages, its performance may deteriorate.

Possible changes can involve:

Capacitance

ESR

Leakage

Performance under ripple current

A capacitor may therefore contribute to excessive ripple even when its capacitance value initially appears reasonable.

Step 1 – Identify the Power Rail

Before measuring, identify exactly what you are testing.

For example:

3.3 V rail

5 V rail

12 V rail

24 V rail

or another DC output.

Know the expected voltage before connecting the oscilloscope.

Do not probe an unknown power rail blindly.

Step 2 – Determine Whether the Measurement Is Safe

This is critical.

The DSO-TC3 Plus should only be connected where the signal and measurement arrangement are suitable for the instrument.

Extra care is required with:

AC mains

SMPS primary circuits

High-voltage DC

Inverters

UPS systems

Motor drives

Floating power circuits

Do not assume that every power supply point is safe simply because you want to measure ripple.

Primary Side vs Secondary Side

In a mains-powered switching power supply, the primary and secondary sides have very different safety considerations.

The primary side can contain dangerous voltage even after power is removed because capacitors may retain charge.

Basic portable oscilloscope measurements should not be attempted on high-voltage primary-side circuits without the appropriate knowledge, equipment and measurement method.

For basic troubleshooting, technicians often begin with suitable low-voltage secondary outputs.

Step 3 – Connect to the Correct Reference

The oscilloscope needs an appropriate reference connection.

For a suitable low-voltage DC output, identify the correct circuit reference before connecting the probe.

Incorrect reference connections can cause:

Incorrect measurements

Noise

Short circuits

Equipment damage

and potentially dangerous conditions.

Understand the circuit before making the connection.

Step 4 – Check the DC Voltage First

Before analysing ripple, check whether the basic output voltage is approximately correct using an appropriate instrument.

For example:

Expected:

12 V DC

Measured:

Approximately 12 V DC

The supply may still have ripple, but this confirms that the main DC level is present.

If the output is completely missing, you may need to troubleshoot a different fault first.

Step 5 – Adjust the Oscilloscope Settings

Ripple is usually much smaller than the main DC voltage.

Therefore, the oscilloscope needs to be configured so that the small variation can be seen clearly.

The exact setup depends on the circuit, signal and instrument capabilities.

The important objective is to display the small changing component without losing sight of safe input conditions.

Why Does the Ripple Look Like a Flat Line?

Possible reasons include:

Ripple is genuinely small

Vertical scale is not sensitive enough

Time base is unsuitable

Wrong test point

Poor probe connection

Adjust the settings appropriately before concluding that there is no ripple.

Why Does the Ripple Look Huge?

A very large waveform may indicate an actual power supply problem.

But first check the measurement technique.

False or exaggerated noise can be caused by:

Long probe ground connection

Poor reference connection

Electromagnetic interference

Nearby switching circuits

Incorrect probe arrangement

Always improve the probing method before condemning the power supply.

Probe Technique Is Very Important

When measuring relatively small ripple on a DC rail, the measurement setup can significantly affect the result.

A long loop formed by the probe and reference connection can pick up electromagnetic noise.

This may make a good power supply look worse than it actually is.

Keep the measurement connection appropriate and minimise unnecessary loop area where possible.

Step 6 – Adjust the Time Base

Ripple is a waveform, so the horizontal time setting matters.

If the time base is too slow, many ripple cycles may be compressed together.

If it is too fast, you may see only a small section of one cycle.

Adjust the time base until the waveform pattern becomes clear.

What Should Normal Ripple Look Like?

There is no universal ripple waveform or acceptable value for every power supply.

The correct limit depends on factors such as:

Power supply design

Output voltage

Load

Switching frequency

Application

Manufacturer specification

Do not use one arbitrary ripple number for every circuit.

Compare the measurement with the equipment or power supply specification where available.

Ripple Under No Load vs Under Load

A power supply may behave differently when loaded.

For example:

No load → appears stable

Normal load → ripple increases

This can reveal problems that are not obvious during an unloaded test.

However, load testing must be performed using an appropriate load and safe procedure.

Do not intentionally overload a power supply simply to see what happens.

Ripple That Increases with Load

If ripple becomes excessive as load increases, possible areas to investigate include:

Output capacitors

Power supply regulation

Inductor

Rectifier components

Excessive load

Current capability

Control-loop behaviour

The exact diagnosis depends on the circuit.

Example: 5 V Power Supply Causes Equipment Reset

Suppose a control board repeatedly resets.

A multimeter shows approximately:

5 V

So the supply initially appears normal.

But the oscilloscope shows that the 5 V rail periodically drops or contains significant variation.

Now the technician has a stronger reason to investigate:

Power supply capacitors

Regulator

Load current

Connections

Upstream supply

This is more useful than relying on the average voltage alone.

Example: Audio Equipment Has Hum or Noise

Power supply problems can sometimes contribute to unwanted audio noise.

An oscilloscope may help determine whether unwanted periodic variation is present on a supply rail.

However, audio hum can also come from:

Grounding

Signal cables

Amplifier circuits

External interference

Do not assume every audio-noise problem is caused by a capacitor.

Example: LED Driver Flickers

An unstable LED driver output may produce visible flicker.

Possible causes can include:

Power supply instability

Capacitor deterioration

Driver-control problems

Load problems

LED failure

Waveform analysis can provide additional information, but the correct measurement point and safe measurement method must be selected.

Example: Controller Repeatedly Starts and Stops

A switching power supply may enter a repeated startup cycle.

You may observe:

Voltage rises

↓

Voltage falls

↓

Voltage rises again

This could indicate that the power supply is attempting to start and then entering protection.

Possible causes may include:

Overload

Short circuit

Feedback fault

Controller supply problem

Output capacitor problem

Power component failure

The waveform pattern helps guide the next troubleshooting step.

Ripple vs Switching Noise

Not every high-frequency feature visible on a power rail should automatically be described as simple ripple.

A switching power supply can also contain:

Switching spikes

High-frequency noise

Transient events

These may require more advanced measurement techniques and higher-performance equipment to analyse accurately.

A portable oscilloscope is useful for basic troubleshooting, but it does not replace specialised power-electronics measurement equipment.

Can a Bad Capacitor Cause Ripple?

Yes, a degraded filtering capacitor can contribute to excessive ripple.

But do not jump directly from:

“Ripple is high”

to

“Replace capacitor.”

Other possible causes include:

Excessive load

Incorrect switching operation

Rectifier problem

Inductor problem

Regulation problem

Poor measurement technique

Use additional evidence before replacing components.

Test the Capacitor Separately

If a capacitor is suspected:

Power down safely

↓

Confirm capacitor is discharged

↓

Remove or isolate it if required

↓

Measure the capacitor

↓

Evaluate relevant parameters

↓

Compare with specification

This links the DSO-TC3 Plus component tester function with oscilloscope troubleshooting.

Capacitor Tests Good but Ripple Is Still High

This situation is possible.

The capacitor may appear reasonable during a basic component test but behave differently under:

Operating voltage

Temperature

Ripple current

Load

Alternatively, the ripple may not be caused by the capacitor at all.

Continue troubleshooting the complete power supply.

Compare Input and Output

In suitable low-voltage circuits, comparing different power rails can help identify where instability begins.

For example:

Input supply stable

but

Regulator output unstable

This directs attention toward the regulator stage and its surrounding components.

Comparison measurements can often be more informative than analysing one waveform alone.

Compare with a Known-Good Board

If two identical pieces of equipment are available, a known-good board can provide valuable reference information.

Compare:

DC voltage

Ripple waveform

Startup behaviour

Load behaviour

Component measurements

This is especially useful when manufacturer waveform specifications are unavailable.

Do Not Chase Every Small Waveform

No real electronic circuit is perfectly noise-free.

Seeing a small waveform on top of a DC rail does not automatically mean the power supply is defective.

The important questions are:

How large is it?

What is its frequency?

Does it change under load?

Does it correspond with the equipment fault?

Is it within the manufacturer's specification?

Troubleshooting requires interpretation, not simply finding any visible variation.

DSO-TC3 Plus Power Supply Ripple Troubleshooting Workflow

A practical workflow is:

1. Understand the power supply circuit

2. Identify the correct low-voltage test point

3. Confirm safe measurement conditions

4. Measure the DC voltage

5. Connect the oscilloscope correctly

6. Adjust voltage scale and time base

7. Observe the ripple waveform

8. Improve probe technique if the waveform looks noisy

9. Compare under appropriate operating conditions

10. Investigate capacitors and other power components if required

11. Confirm the repair by repeating the measurement

This provides a much stronger diagnosis than simply replacing capacitors at random.

When Should You Use a Bench Oscilloscope?

More demanding ripple and power-electronics measurements may require:

Higher bandwidth

Better waveform analysis

Multiple channels

Differential probes

High-voltage probes

Advanced triggering

Specialised measurement accessories

The DSO-TC3 Plus is better positioned for suitable basic low-voltage troubleshooting and portable electronics work.

FNIRSI DSO-TC3 Plus Power Supply Testing Malaysia

The FNIRSI DSO-TC3 Plus can support a useful troubleshooting sequence:

Measure component

  •  

Observe waveform

For power supply repair, this means technicians can investigate a suspected capacitor and then examine suitable low-voltage circuit behaviour using the oscilloscope.

The key is to understand that:

Correct DC voltage does not always mean clean DC power.

And:

Visible ripple does not automatically prove a bad capacitor.

Use the waveform together with component tests and circuit symptoms.

FNIRSI DSO-TC3 Plus Malaysia

MTM Precision supplies FNIRSI electronic testing instruments in Malaysia for PCB repair, power supply troubleshooting, electronics repair, component testing, maintenance and technical education.

Customers from Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kelantan, Terengganu, Kedah, Perlis, Sabah and Sarawak may contact us for product enquiries and availability.

If you need to check power supply ripple or an unstable DC output, send us a photo of the power supply PCB, output voltage information and the waveform through WhatsApp.

If possible, also tell us:

Expected output voltage

Equipment type

Fault symptom

This makes it easier to understand what measurement approach may be suitable.