Why Does My FNIRSI 1014D Show the Wrong Frequency or Waveform Malaysia?

Why Does My FNIRSI 1014D Show the Wrong Frequency or Waveform Malaysia?

You connect the FNIRSI 1014D Oscilloscope to a circuit, but something does not look right.

The frequency is different from what you expected.

The square wave looks rounded.

The amplitude seems too low.

Or the waveform looks noisy and distorted.

Does this mean the oscilloscope is inaccurate?

Not necessarily.

Before deciding that the instrument has a problem, check the probe, oscilloscope settings, signal characteristics and measurement technique.

1. Check the Probe Setting First

One of the easiest mistakes to make is using the wrong probe attenuation setting.

For example, an oscilloscope probe may be configured for:

1X

or

10X

The oscilloscope setting needs to correspond with the actual probe configuration.

If the probe and oscilloscope settings do not match, the displayed voltage may be incorrect.

So if the waveform shape looks reasonable but the amplitude appears wrong, check the probe setting first.

2. Check the Probe Compensation

A poorly compensated probe can make a square wave look incorrect.

Instead of a clean square-wave shape, you may see edges that appear:

Rounded

or

Over-peaked

This does not automatically mean the circuit signal is bad.

The probe itself may need to be checked and adjusted using an appropriate reference signal and the manufacturer's recommended procedure.

Probe condition is an important part of oscilloscope measurement.

3. Check the Time Base

If the horizontal time scale is unsuitable, the waveform may be difficult to interpret.

For example:

Too slow → Many cycles compressed together

Too fast → Only a small section of the waveform visible

Adjust the time base until several clear cycles are visible.

This makes it easier to check frequency and waveform shape.

4. Check the Trigger Setting

An unstable trigger can make a good repetitive signal look confusing.

The waveform may:

  • Move continuously

  • Jump between positions

  • Appear inconsistent

  • Be difficult to measure

Check the:

Trigger source

Trigger level

Trigger edge

For a repetitive waveform, correct triggering should normally make the display easier to observe.

5. Is the Signal Frequency Actually Stable?

Do not assume that the circuit is producing exactly the frequency you expect.

The oscilloscope may be revealing a real circuit problem.

For example, an oscillator or PWM source may have:

  • Changing frequency

  • Intermittent operation

  • Missing pulses

  • Timing instability

  • Unexpected modulation

If possible, compare the measurement with the circuit specification or a known-good unit.

6. Why Does a Square Wave Look Rounded?

This is a very common oscilloscope question.

A square wave contains much more high-frequency content than its fundamental frequency alone suggests.

The sharp edges require higher-frequency components.

As the signal approaches the bandwidth limitations of the:

Oscilloscope

Probe

or

measurement setup

the displayed edges may become more rounded.

Therefore:

Signal frequency below 100MHz

does not automatically mean:

Perfect square-wave reproduction on a 100MHz oscilloscope.

7. Understand the 100MHz Bandwidth Specification

The FNIRSI 1014D is specified as a 100MHz-bandwidth oscilloscope.

This should not be interpreted as meaning every waveform up to 100MHz will be reproduced perfectly.

As the frequency approaches the oscilloscope bandwidth limit, measurement response becomes increasingly important.

If your work requires accurate high-frequency waveform shape or fast-edge analysis, a higher-bandwidth professional oscilloscope may be required.

8. Sampling Rate Also Matters

The oscilloscope converts an analogue waveform into digital samples.

If the signal and oscilloscope settings do not provide suitable waveform acquisition, the displayed result may not represent the signal as clearly as expected.

The FNIRSI 1014D is specified with a sampling rate of up to 1GSa/s under specified conditions.

However, sampling rate should not be considered in isolation.

Oscilloscope performance also depends on factors such as:

  • Bandwidth

  • Memory depth

  • Time-base setting

  • Channel configuration

  • Trigger

  • Probe

  • Signal characteristics

9. Check Whether You Are Measuring Noise

Sometimes the waveform looks bad because the signal really contains noise.

Possible sources include:

  • Switching power supplies

  • Digital circuits

  • Motors

  • Poor grounding

  • Long wires

  • Nearby electrical equipment

  • PCB switching activity

The oscilloscope may simply be showing what is actually present on the circuit.

Before deciding the oscilloscope is wrong, compare the signal at different points and operating conditions.

10. Your Ground Lead Can Affect the Waveform

Probe grounding is especially important when observing fast or small signals.

A long ground connection can introduce additional inductance and pick up unwanted electrical noise.

This may produce:

Ringing

Spikes

Noise

or other waveform artifacts.

For sensitive measurements, use an appropriate probing technique and keep connections suitable for the signal being measured.

11. Check the Measurement Point

Two points that look electrically similar on a schematic may not always produce exactly the same waveform in a real circuit.

PCB traces, components, grounding and loading can affect what you see.

Make sure you are probing the correct test point.

If service information is available, compare your measurement location with the recommended test point.

12. The Probe Can Load the Circuit

Connecting an oscilloscope probe changes the circuit slightly.

For many measurements this effect may be small.

But sensitive oscillator circuits and higher-frequency signals can be affected by probe loading.

If the circuit changes behaviour when you connect the probe, consider whether the probe impedance and capacitance are affecting the measurement.

More specialized probes may be required for demanding applications.

13. Why Is the Measured Amplitude Too Low?

Possible reasons include:

Incorrect probe attenuation setting

Oscilloscope bandwidth limitation

Probe bandwidth limitation

Circuit loading

Wrong test point

Signal source impedance

Actual circuit fault

Do not immediately adjust the reading mentally to match what you expect.

Instead, determine why the measured result is different.

14. Why Is the Frequency Reading Jumping?

If the frequency value changes continuously, check whether:

The waveform is stable

The trigger is stable

The signal amplitude is sufficient

There is excessive noise

The probe connection is secure

The waveform is actually periodic

An automatically calculated frequency value is only as useful as the waveform being measured.

15. Use the Built-In Signal Generator as a Basic Check

The FNIRSI 1014D includes a built-in signal generator.

For suitable troubleshooting and learning, a known generated waveform can help you understand whether your oscilloscope settings are reasonable.

Generate a suitable test signal and observe it.

If the known signal displays normally but the PCB waveform does not, the issue may be related to the circuit, test point or measurement method rather than the oscilloscope itself.

16. Compare with a Known-Good Signal

When troubleshooting measurement problems, always try to reduce the number of unknowns.

Instead of asking:

“Why does this complicated PCB waveform look wrong?”

start with:

“Can I display a known stable waveform correctly?”

Then move back to the circuit under test.

This simple approach can save a lot of troubleshooting time.

Is the FNIRSI 1014D Faulty?

Do not conclude that the instrument is faulty based on one unexpected waveform.

First check:

  • Probe attenuation

  • Probe compensation

  • Time base

  • Vertical scale

  • Trigger settings

  • Ground connection

  • Measurement point

  • Signal frequency

  • Signal stability

  • Bandwidth requirements

If a known signal still produces obviously incorrect results after correct setup, then further checking of the probe or oscilloscope may be necessary.

Good Oscilloscope Measurement Is More Than Connecting a Probe

An oscilloscope is not like a simple voltage indicator.

Correct waveform measurement depends on the complete measurement system:

Signal + Probe + Connection + Oscilloscope Settings + User Technique

Understanding this is especially important when moving from low-frequency beginner experiments into faster electronic signals.

Looking for FNIRSI 1014D in Malaysia?

For FNIRSI 1014D 100MHz Dual-Channel Oscilloscope with Signal Generator enquiries in Malaysia, contact MTM Precision Sdn Bhd.

If your waveform or frequency reading does not look correct, you can WhatsApp us a photo of the oscilloscope screen, probe setting and equipment being tested for an initial assessment.

MTM Precision Sdn Bhd

Showroom & Service Centre:
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.

Tel: 03-8080 7172
WhatsApp: +6016-660 7346
Email: mtmpre@yahoo.com
Website: www.mtmpre.com.my

Serving customers across Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kelantan, Terengganu, Kedah, Perlis, Sabah and Sarawak.

05 Oct 2026