Common FNIRSI 1014D Oscilloscope Measurement Mistakes and How to Avoid Them Malaysia

Common FNIRSI 1014D Oscilloscope Measurement Mistakes and How to Avoid Them Malaysia

You connect the FNIRSI 1014D Oscilloscope to a PCB, but the waveform looks wrong.

The voltage is ten times different from what you expected.

The waveform keeps moving.

A square wave looks rounded.

The screen shows excessive noise.

Does this mean the oscilloscope is faulty?

Not necessarily.

Many oscilloscope problems are actually caused by probe setup, grounding, trigger settings or measurement technique.

Here are some common mistakes to check before deciding there is a problem with the FNIRSI 1014D or the circuit under test.

Mistake 1: Wrong 1X / 10X Probe Setting

This is one of the easiest mistakes to make.

If the physical probe is set to:

10X

but the oscilloscope is configured for:

1X

the displayed voltage may be incorrect.

The opposite mismatch can also cause incorrect readings.

Whenever the waveform amplitude appears approximately ten times too high or too low, check the probe attenuation settings first.

Mistake 2: Assuming the Ground Clip Is a Floating Multimeter Lead

This is a much more serious mistake.

A beginner may think:

Probe Tip = Red Multimeter Lead

Ground Clip = Black Multimeter Lead

But a bench oscilloscope probe should not automatically be treated like two independent floating multimeter leads.

The oscilloscope grounding arrangement must be understood before making measurements.

Incorrect connection can create a short circuit.

Mistake 3: Connecting Ground to the Wrong Point

Suppose you are measuring a mains-powered or bridge circuit.

If you attach the oscilloscope ground clip to a point that is not suitable for the oscilloscope's ground reference, you may create a dangerous current path.

Possible consequences include:

Damaged PCB

Damaged probe

Damaged oscilloscope

Electrical shock or other serious hazards

Never guess when working on mains or high-energy circuits.

Use the appropriate probe and measurement method.

Mistake 4: Using a Long Ground Lead for Fast Signals

A long ground connection may pick up noise and add unwanted inductance.

When measuring faster signals, this can make the waveform appear to contain:

Ringing

Spikes

Noise

You may then mistakenly blame the PCB.

For sensitive measurements, use a suitable probing technique and keep the measurement loop appropriate for the signal.

Mistake 5: Wrong Time-Base Setting

Sometimes the waveform is correct, but the display setting makes it look wrong.

If the time base is too slow:

Many waveform cycles may be compressed together.

If the time base is too fast:

You may see only a small part of one cycle.

Adjust the horizontal scale until the important part of the waveform is clearly visible.

Mistake 6: Wrong Vertical Scale

The same problem can happen vertically.

If the voltage scale is too large, a small signal may look almost flat.

If it is too small, the waveform may extend beyond the screen.

Set the vertical scale so the waveform occupies a useful portion of the display.

Mistake 7: Incorrect Trigger Source

Suppose your signal is connected to CH1.

But the oscilloscope is triggering from another source that does not contain a suitable repetitive signal.

The waveform may refuse to stabilize.

Check:

Trigger source

Trigger level

Trigger edge

before assuming the measured signal is unstable.

Mistake 8: Trigger Level Outside the Signal

The trigger level needs to interact appropriately with the waveform.

If your signal moves between approximately:

0V and 5V

but the trigger level is positioned outside the relevant signal range, stable triggering may become difficult.

Move the trigger level to a suitable point within the waveform.

Mistake 9: Assuming Every Moving Waveform Is a Trigger Problem

Sometimes the waveform keeps changing because the signal itself is changing.

Examples include:

Variable PWM

Communication data

Sensor output

Startup sequence

Intermittent fault

Noise

Trigger adjustment cannot make a genuinely changing signal become perfectly repetitive.

The oscilloscope may be showing real circuit behaviour.

Mistake 10: Expecting a Perfect Square Wave

A real square wave may not look perfectly square.

Its appearance depends on:

Signal rise time

Oscilloscope bandwidth

Probe bandwidth

Probe loading

Measurement connection

Circuit impedance

Signal frequency

A 100MHz oscilloscope specification does not mean every square wave below 100MHz will be reproduced with perfect vertical edges.

Mistake 11: Ignoring Probe Bandwidth

The oscilloscope is only one part of the measurement system.

Think of the complete chain:

Circuit → Probe → Oscilloscope

If the probe cannot reproduce the signal properly, a high oscilloscope bandwidth alone cannot fix the measurement.

Always consider the probe specification when measuring faster waveforms.

Mistake 12: Not Checking Probe Compensation

A passive 10X probe may require proper compensation.

If the probe is not compensated correctly, a square-wave reference may appear:

Rounded

or

Over-peaked

You may then incorrectly conclude that the circuit waveform is distorted.

Check the probe using an appropriate reference signal and follow the recommended compensation procedure.

Mistake 13: Measuring the Wrong Test Point

On a complicated PCB, probing the wrong side of a component can produce a completely different waveform.

Before making conclusions, confirm:

Where is the signal generated?

Where should it travel?

Which point are you actually measuring?

A schematic, service manual or known-good PCB can be very useful.

Mistake 14: Ignoring Probe Loading

Connecting a probe changes the circuit slightly.

For many ordinary measurements this may not create a major problem.

But sensitive circuits such as oscillator nodes can be affected by probe capacitance and impedance.

If an oscillator changes behaviour when the probe is connected, consider whether the measurement itself is influencing the circuit.

Mistake 15: Assuming Frequency Alone Tells the Whole Story

Suppose two signals both measure:

20kHz

They are not necessarily identical.

One may have:

Correct amplitude

Clean edges

Stable duty cycle

while the other may contain:

Noise

Incorrect amplitude

Distortion

Missing pulses

Unstable duty cycle

Always look at the waveform, not only the frequency number.

Mistake 16: Confusing Peak, Peak-to-Peak and RMS

Voltage can be described in different ways.

For example:

Peak voltage

Peak-to-peak voltage

RMS voltage

These values are not interchangeable.

When comparing the oscilloscope result with a specification or another instrument, make sure both values refer to the same measurement type.

Mistake 17: Using Auto Measurement Without Looking at the Waveform

Automatic measurements are convenient.

But do not look only at the numbers.

If the waveform is noisy, unstable or incorrectly triggered, an automatically calculated value may not tell the whole story.

First look at the waveform.

Ask:

Does this signal make sense?

Then interpret the measurement values.

Mistake 18: Starting with a Complicated PCB

If you are learning the FNIRSI 1014D, do not make your first measurement on an unknown industrial control board.

Start with a simple, known low-voltage waveform.

Practice:

Voltage scale

Time base

Trigger

Frequency measurement

1X / 10X probe settings

Dual-channel operation

Then move to real PCB troubleshooting.

Mistake 19: Assuming the Oscilloscope Automatically Finds the Fault

An oscilloscope does not normally tell you:

“Replace capacitor C25.”

It provides information about what the electrical circuit is doing.

The technician still needs to interpret:

Where the signal should be

What the signal should look like

Where the behaviour becomes abnormal

The oscilloscope is a diagnostic tool.

Mistake 20: Randomly Replacing Components Before Signal Tracing

If a PCB has a signal problem, avoid immediately replacing multiple ICs and capacitors without diagnosis.

A better approach may be:

Check power

→ Check clock

→ Check reset

→ Check control signal

→ Check driver

→ Check output

Find the first point where the expected behaviour disappears.

This can make PCB repair much more systematic.

Mistake 21: Ignoring the Second Channel

If you have a dual-channel oscilloscope, use the second channel when it provides useful information.

For example:

CH1 → Input

CH2 → Output

or:

CH1 → PWM Control

CH2 → Driver Output

Seeing both signals at the same time can reveal relationships that are difficult to identify with separate measurements.

Mistake 22: Forgetting the Built-In Signal Generator

The FNIRSI 1014D also includes a signal generator.

For suitable low-voltage circuits, a known test signal can help with:

Amplifier testing

Filter experiments

Signal tracing

Electronics education

A known signal can make troubleshooting easier because you already know what is being applied to the circuit.

Mistake 23: Using the Oscilloscope on High Voltage Without Understanding the Circuit

This is one mistake that should never be treated casually.

Do not connect the FNIRSI 1014D directly to:

AC mains

High-voltage inverter sections

Motor drives

Unknown switching nodes

High-energy power circuits

without understanding the voltage, grounding arrangement and required probe system.

An appropriate high-voltage differential probe, isolated measurement solution or other approved method may be required depending on the application.

A Better FNIRSI 1014D Troubleshooting Routine

Before blaming the oscilloscope or PCB, check these points:

1. Correct test point

2. Correct probe

3. Correct 1X / 10X setting

4. Correct grounding

5. Correct vertical scale

6. Correct time base

7. Correct trigger source

8. Correct trigger level

9. Signal frequency within suitable measurement capability

10. Compare with a known-good signal if possible

Following the same routine each time reduces measurement mistakes.

Is the FNIRSI 1014D Suitable for PCB Troubleshooting?

For general electronics work, the FNIRSI 1014D can be useful for:

  • Power supply ripple

  • PWM

  • Clock signals

  • Oscillator circuits

  • Amplifier waveforms

  • Input/output comparison

  • Signal tracing

  • Arduino and MCU troubleshooting

  • Basic FFT analysis

  • Electronics education

The most important factor is not simply owning an oscilloscope.

It is knowing how to connect and interpret it correctly.

Looking for FNIRSI 1014D in Malaysia?

For FNIRSI 1014D Oscilloscope with Signal Generator enquiries in Malaysia, contact MTM Precision Sdn Bhd.

If you already have a measurement problem and are unsure whether the FNIRSI 1014D is suitable, WhatsApp us a photo of the PCB, equipment, measurement point or waveform you need to check.

MTM Precision Sdn Bhd

Showroom & Service Centre:
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.

Tel: 03-8080 7172
WhatsApp: +6016-660 7346
Email: mtmpre@yahoo.com
Website: www.mtmpre.com.my

Serving customers across Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kelantan, Terengganu, Kedah, Perlis, Sabah and Sarawak.

05 Oct 2026