2

Share:


CUSTOM Digital logger illuminance meter LX-2000SD

Specifications

SKU: CUSTOM Digital logger illuminance meter LX-2000SD

Product overview

[Measurement item] Illuminance
[Function] Hold
[Function] MAX MIN
[Features] Equipped with backlight
[Features] AC adapter compatibleSold separately
[Function] Stand
*1 Sensor sold separately
*2 Approximately 10 minutes (cancellable)
*3 SD card up to 16GB

A light meter that can also record data to an SD card.

  • The recorded data can be used with commercially available spreadsheet software, making it suitable for a variety of purposes.
  • Data recording can be selected from automatic recording or manual recording.
  • The sampling time for the automatic data logger can be set to any of the following: 1, 2, 5, 10, 30, 60, 120, 300, 600, 1800, or 3600 seconds.
  • Equipped with a hold function that temporarily freezes the measurement value on the display.
  • Comes with a MAX (maximum value)/MIN (minimum value) display function for measurement values.
  • To prevent battery drain, the device is equipped with an auto power off function that automatically turns off the power if the device is not operated for a certain period of time.
  • Temperature measurement is also possible by connecting a K/J type thermocouple sensor (sold separately).
  • Comes with a green backlight function that makes it easy to see even in dark places.

Measurement results can be recorded by inserting an SD card.

The data recorded on the SD card can be imported into spreadsheet software such as Excel as a CSV file.
CT-05SD SD card installation exampleLX-2000SD CSV file exampleThe image shows a model in the same series with the same functions.
Microsoft Excel is a registered trademark or trademark of Microsoft Corporation in the United States and/or other countries.

Up to four thermocouple sensors (sold separately) can be connected to measure temperature.

LX-2000SD sensor connection imageCompatible thermocouples: K/J/T/E/R/S types
*Thermocouple sensor is not included in the product.

Part names

LX-2000SD Part names
  • ①Display section
  • ②POWER button (ESC button/backlight button)
  • ③FUNC. button (HOLD button/NEXT button)
  • ④REC button (ENTER button/Unit button)
  • ⑤SET button (▼ button/Time check button)
  • ⑥LOGGER buttons (▲ button/Zero button/Sampling check button)
  • ⑦ Stand
  • ⑧Battery cover
  • 9. Battery cover screws (2 pcs)
  • ⑩Tripod screw hole
  • ⑪K/J type thermocouple sensor connector socket
  • ⑫ Probe socket
  • ⑬DC9V AC adapter input socket
  • 14. RESET button
  • ⑮SD card socket
  • ⑯ Probe plug
  • ⑰Illuminance sensor head
  • ⑱Illuminance sensor
  • ⑲Sensor cover
▲Back to top of page

specification

Electrical Performance

illuminance

range Display range resolution Measurement accuracy
2,000 lux 0 to 1,999 lux 1 lux ±(4%+2dgt)
20,000 lux 1,800 to 19,990 lux 10 lux
100,000 lux 18,000 to 99,900 lux 100 lux
  • Accuracy was tested by comparison with a standard parallel light tungsten lamp at an absolute temperature of 2856K.

temperature

kinds Measurement range resolution Measurement accuracy
K-type -50.0 to +1300.0℃ 0.1℃ ±(0.4%+0.5℃)
-100.0 to -50.1°C ±(0.4%+1℃)
J-type -50.0 to +1200.0℃ 0.1℃ ±(0.4%+0.5℃)
-100.0 to -50.1°C ±(0.4%+1℃)
  • The above measurement accuracy does not include the accuracy of optional sensors.
  • We do not carry J-type thermocouple sensors.
  • Using the sensor for purposes other than its intended use or handling it roughly may result in damage to the sensor or injury, so please select a sensor that is suitable for your purpose.
  • Depending on the object being measured and how it is used, the sensor may be damaged.
  • Please use the sensor within a temperature range with sufficient margin.

General specifications

sensor Silicon photodiode
sampling Approximately 1 time/second
Auto Power Off Approximately 10 minutes (can be canceled)
Operating temperature and humidity 0 to +50°C, 85% RH or less (no condensation)
Storage temperature and humidity -20 to +60°C, 85% RH or less (no condensation)
size
  • Body: Approx. W 68 x H 177 x D 45 mm
  • Sensor: Approx. W 58 x H 88.5 mm
weight Approximately 430 g (including battery and light sensor)
power supply 6 AA batteries or AC adapter (sold separately)
battery life
  • Normal measurement: Approximately 300 hours
  • Data logger function operating: Approximately 70 hours (backlight off, alkaline batteries used)
accessories Instruction manual, light sensor, soft case
proofreading Possible
option
JAN 4983621201017
  • The above performance is at 23 ± 5°C
Precautions when using a light meter
  • Illuminance meters are calibrated using a filament-type light bulb known as a standard light bulb as a standard light source. Fluorescent lamps and LED lighting, which are commonly used in general workplaces, have different wavelengths of light compared to standard light bulbs, so measurements may differ.
    In such cases, it may be necessary to take measures such as measuring with multiple illuminance meters (different models from different manufacturers) and taking the average value.
About SD cards
  • Although commercially available SD cards (1-16GB) are compatible, operation has not been certified for all SD card products from all manufacturers. Please note that some SD cards may not be compatible.
  • The SD card is not included and must be purchased separately.
  • Other types of memory cards, such as SDXC cards and MMC cards, cannot be used.
  • SDXC and SDHC are registered trademarks of the SD Association.


 Inquiry - CUSTOM Digital logger illuminance meter LX-2000SD