Specifications
SKU: CUSTOM Data Logger Intelligent Water Checker IWC-6SD
Product overview
*2 Approximately 10 minutes (cancellable)
*3 SD card up to 16GB
This water quality measuring instrument combines the functions of a pH meter, oxidation-reduction potential (ORP) meter, conductivity meter, total dissolved impurities (TDS) meter, dissolved oxygen meter, and salinity meter into one unit.
- By connecting an optional sensor that matches the measurement item, you can measure and record data to an SD card.
- The recorded data is in CSV format, which can be used with commercially available spreadsheet software, and can be used for a variety of purposes.
- Data can be recorded automatically or manually. The sampling time for automatic recording can be selected from 11 levels between 1 and 3600 seconds.
- Comes with a green backlight function that makes it easy to see even in dark places.
- Equipped with a hold function that temporarily freezes the measurement value on the display.
- Comes with a "MAX/MIN memory function" that stores the maximum and minimum measurement values.
- To prevent battery drain, the device is equipped with an auto power off function that automatically turns off the power if the device is not operated for a certain period of time.
Measurement results can be recorded by inserting an SD card.
The data recorded on the SD card can be imported into spreadsheet software such as Excel as a CSV file.
Part names
- ①Display section
- ②POWER button (backlight button)
- ③HOLD button (ESC button/)
- ④REC button (ENTER button/B button)
- ⑤MODE button (▲ button)
- ⑥RANGE button (▼ button/Function button/Sampling check)
- ⑦TIME button
- ⑧Logger button (SET button/sampling check)
- ⑨ Stand
- ⑩Battery cover
- ⑪ Battery cover screws (2 pcs)
- ⑫Tripod screw hole
- ⑬Temperature sensor socket (socket for automatic pH temperature compensation)
- ⑭ Dissolved oxygen sensor socket
- ⑮Conductivity sensor socket
- ⑯ pH electrode socket (BNC socket)
- ⑰DC9VAC adapter input socket
- ⑱RESET button
- ⑳SD card socket
specification
Electrical Performance
pH
| range | resolution | Measurement accuracy |
|---|---|---|
| 0 to 14.00 pH | 0.01 pH | ±(0.02 pH + 2 dgt) |
- Sensor used: PE-03 (optional)
- The above measurement accuracy does not include the accuracy of optional sensors.
Oxidation-reduction potential (ORP)
| range | resolution | Measurement accuracy |
|---|---|---|
| -1999 to +1999 mV | 1 mV | ±(0.5% + 2dgt) |
- Sensor used: ORP-14 (optional)
- The above measurement accuracy does not include the accuracy of optional sensors.
conductivity
| range | Measurement range | resolution | Measurement accuracy |
|---|---|---|---|
| 200 μS | 0 to 200.0 μS/cm | 0.1 μS/cm | ± (2% FS + 1 dgt) / < 100 mS/cm |
| 2 mS | 0.2 to 2,000 mS/cm | 0.001 mS/cm | |
| 20 mS | 2 to 20.00 mS/cm | 0.01 mS/cm | |
| 200mS | 20 to 200.0 mS/cm | 0.1 mS/cm |
- Sensor used: CDPB-03 (optional)
- FS=full scale
- The above measurement accuracy does not include the accuracy of optional sensors.
Total Dissolved Solids (TDS)
| range | Measurement range | resolution | Measurement accuracy |
|---|---|---|---|
| 200 ppm | 0 to 132 ppm | 0.1 ppm | ±(2% FS +1dgt) / < 66,000ppm |
| 2,000 ppm | 132 to 1,320 ppm | 1 ppm | |
| 20,000 ppm | 1,320 to 13,200 ppm | 10 ppm | |
| 200,000 ppm | 13,200 to 132,000 ppm | 100 ppm |
- Sensor used: CDPB-03 (optional)
- The above measurement accuracy does not include the accuracy of optional sensors.
temperature
| Measurement range | resolution | Measurement accuracy |
|---|---|---|
| 0 to +60°C | 0.1℃ | ±0.8℃ |
- Sensor used: TP-07 (optional)
- The above measurement accuracy does not include the accuracy of optional sensors.
Salt content
| Measurement range | resolution | Measurement accuracy |
|---|---|---|
| 0 to 12.00% | 0.01% | ±0.5% FS |
- Sensor used: CDPB-03 (optional)
- FS=full scale
- Temperature range of the solution to be measured: 0 to +60°C
- The above measurement accuracy does not include the accuracy of optional sensors.
Dissolved oxygen (DO)
| Measurement range | resolution | Measurement accuracy |
|---|---|---|
| 0-20.0 mg/ℓ | 0.1 mg/ℓ | ±0.4 mg/ℓ |
- Sensor used: OXPB-11 (optional)
- The above measurement accuracy does not include the accuracy of optional sensors.
atmospheric oxygen
| Measurement range | resolution | Measurement accuracy |
|---|---|---|
| 0 to 100% | 0.1% O2 | ±0.7% O2 |
- Sensor used: OXPB-11 (optional)
- The above measurement accuracy does not include the accuracy of optional sensors.
General specifications
| Auto Power Off | Approximately 10 minutes (can be canceled) |
|---|---|
| Operating temperature and humidity | 0 to +50°C, 85% RH or less (no condensation) |
| Storage temperature and humidity | -10 to +60°C, 85% RH or less (no condensation) |
| size | Approximately W 68 × H 177 × D 45 mm |
| weight | Approximately 489 g (including batteries) |
| power supply | 6 AA batteries or AC adapter (sold separately) |
| battery life |
|
| accessories | operating instructions |
| proofreading | Not possible |
| option |
|
| JAN | 4983621991185 |
- Although commercially available SD cards (1-16GB) are compatible, operation has not been certified for all SD card products from all manufacturers. Please note that some SD cards may not be compatible.
- The SD card is not included and must be purchased separately.
- Other types of memory cards, such as SDXC cards and MMC cards, cannot be used.
- SDXC and SDHC are registered trademarks of the SD Association.
Inquiry - CUSTOM Data Logger Intelligent Water Checker IWC-6SD