Share:


SK SATO No.8152-60 Digital Panel Meter SK-EM-20

Specifications

SKU: SK SATO No.8152-60 Digital Panel Meter SK-EM-20

JAN code : 4974425801862

Calibration : May be available depends on probes

JCSS Accredited Calibration : May be available depends on probes

A digital indicator with highly visible LCD of the size of 96 x 48 mm applies to DIN standards. Thermocouples, thermistor, and analog voltage input are available.

Sensor probes need to be purchased separately.

Feature

 Large, bright, highly visible LCD

 Thermocouples, thermistor, and analog voltage input available

 A compact LCD with a size of 96 x 48 mm applies to DIN standards

 LCD can automatically turn off after a preset time period

 Ideal for Temperature/Humidity Transmitter SK-RHC series

 °C/°F switchable

Description

 Measuring ranges and codes

Input method Code Range (°C) Range (°F) Remarks
Thermo-
couple
B 01 0 to 1800 0 to 3300 *2
R 02 0 to 1700 0 to 3100  
S 03 0 to 1700 0 to 3100  
K 04 −199.9 to 800.0 −300 to 1500  
05 0 to 1200 0 to 2200  
E 06 0 to 700 0 to 1300  
J 07 0 to 600 0 to 1100  
T 08 −199.9 to 300.0 −300 to 600 *3
N 09 0 to 1300 0 to 2300  
U 10 −199.9 to 300.0 −300 to 600 *3
L 11 0 to 600 0 to 1100  
C (WRe5-26) 12 0 to 2300 0 to 4200  
Therm-
istor
Pt 31 −199.9 to 600.0 −300 to 1100 *4
32 −100.0 to 100.0 −150.0 to 200.0  
JPt 33 −199.9 to 500.0 −300 to 1000 *4
34 −100.0 to 100.0 −150.0 to 200.0  
Analog
voltage
0–10 mV 71 Scaling to 0.0 to 100.0 available
Scaling supported range: −1999 to 9999 Unit
Span: 10 to 10000 Unit
0–5 V 81
1–5 V 82
0–10 V 83

*1 Factory setting: type-K thermocouple (code 05, 0 to 1200°C)

*2 No accuracy is warranted for type-B thermocouple below 400°C.

*3 Accuracy of type-T and U thermocouple: ±(0.5% F.S.+ 1 digit) (−100 to 0°C), ±(1% F.S.+ 1 digit) (< −100°C)

*4 Pt (code 31) and JPt (code 33) scale out at temperature below −240.0°C (−400°F)

Specifications

Cat. No. 8152-60
Product name Digital Panel Meter
Model name SK-EM-20
Measuring range See "Measuring ranges and codes"
Display accuracy ± (0.5% F.S. + 1 digit) (type-T and U thermocouples −100 to 0°C)
± (1% F.S. + 1 digit) (type-T and U below −100°C)
± (0.3% F.S. + 1 digit) (other)
Out of range (type-B thermocouple below 400°C)
Resolution Depends on ranges
Ingress protection Only the front panel is IP66 dustproof/waterproof
Operating ambient −10 to 50°C, less than 90%rh without condensing, altitude below 2000 m, overvoltage category II, pollution degree 2 (IEC60664)
Complied regulations IEC61010-1, IEC61010-2-030, EN61010-1, EN61010-2-030, EMC EN61326-1
Dimensions 96 (W) x 48 (H) x 111 (D) mm with depth behind the panel 100 mm
Spacing of mounting holes: 92 (W) x 45 (H) mm
Weight 250 g
Power 100–240 VAC
Available accessory Shunt Resistor