Suspended Solids Meter Reading Too Low or Zero? Malaysia

Suspended Solids Meter Reading Too Low or Zero?

 

A suspended-solids meter may display a result that appears lower than expected—or even zero—because of the sample, measuring bottle, procedure or instrument condition.

 

Before assuming that the meter is faulty, check whether particles have settled, the correct sample was collected and the instrument is being operated within its specified range.

 

The following troubleshooting guide applies to portable instruments such as the Qiwei SS-500B Suspended Solids Meter.

 

Cause 1: Suspended Particles Have Settled

 

Many wastewater particles do not remain evenly distributed.

 

Heavy particles may settle at the bottom of the collection container while the clearer water near the top is transferred into the measuring bottle.

 

This can produce a reading that is much lower than expected.

 

Recommended Action

 

Mix the original sample gently.

 

Avoid vigorous shaking.

 

Transfer it immediately into the measuring bottle.

 

Measure using consistent timing.

 

Repeat the test.

 

If particles settle very quickly, document the time between mixing and measurement.

 

Cause 2: The Wrong Part of the Sample Was Collected

 

A surface sample may contain fewer heavy particles than water collected from the middle of the flow.

 

Similarly, a sample taken from a stagnant corner may not represent the main process.

 

Recommended Action

 

Use a fixed and representative sampling location.

 

When comparing results, keep the following consistent:

 

Sampling point

 

Sampling depth

 

Container

 

Collection method

 

Time of testing

 

Process condition

 

Cause 3: The Sample Is Below the Practical Measuring Level

 

The sample may contain very little suspended material.

 

If the actual concentration is close to the instrument’s lower measurement capability, the reading may remain at or near zero.

 

The Qiwei SS-500B displays readings in 1 mg/L increments.

 

Customers requiring finer displayed readings at low concentrations may need to consider a lower-range or higher-resolution model, subject to the complete accuracy specification.

 

Cause 4: The Sample Is Not Suspended Solids

 

Clear water may contain dissolved substances without visible or measurable suspended particles.

 

A water sample can show:

 

High conductivity

 

High TDS

 

Dissolved salts

 

Chemical contamination

 

while still having a low suspended-solids reading.

 

A TDS or conductivity result should not be expected to match a suspended-solids result.

 

Cause 5: The Wrong Instrument Was Selected

 

Check whether the required parameter is:

 

Suspended solids in mg/L

 

Turbidity in NTU

 

TDS in ppm

 

Conductivity in µS/cm

 

Another water-quality parameter

 

A suspended-solids meter will not reproduce the reading from a turbidity, TDS or conductivity meter.

 

Cause 6: The Measuring Bottle Was Not Filled Correctly

 

Incorrect sample volume, leakage or positioning may affect the measurement.

 

Recommended Action

 

Follow the manufacturer’s filling instructions.

 

Use the correct measuring bottle.

 

Check that the cap and bottle are in good condition.

 

Keep the outside clean and dry.

 

Insert it correctly into the meter.

 

Use a consistent orientation where applicable.

 

Cause 7: The Bottle Is Dirty in a Way That Affects the Baseline

 

Dirt often causes high or unstable readings, but residue, stains or incorrect cleaning may also affect the instrument’s baseline and produce unpredictable results.

 

Inspect for:

 

Chemical film

 

Detergent residue

 

Permanent staining

 

Scratches

 

Cloudiness

 

Condensation

 

External water droplets

 

Rinse and clean the bottle properly before repeating the measurement.

 

Cause 8: The Instrument Was Zeroed Incorrectly

 

If the wrong blank or an unsuitable reference was used during zeroing, sample results may appear too low.

 

Recommended Action

 

Follow the manufacturer’s zeroing and calibration procedure.

 

Do not assume that any clear-looking liquid is suitable as a blank. Use only the required reference material or procedure.

 

Cause 9: Incorrect Calibration

 

An unsuitable standard, expired reference material or incorrect calibration step may shift the instrument response.

 

Check:

 

Correct calibration material

 

Reference value

 

Expiry date

 

Sample bottle condition

 

Calibration sequence

 

Instrument settings

 

Manufacturer instructions

 

Do not use a turbidity standard unless the manufacturer specifically states that it is suitable for the suspended-solids meter.

 

Cause 10: Strong Sample Characteristics Are Affecting the Reading

 

Unusual industrial samples may not behave like ordinary wastewater.

 

Potential influences include:

 

Strong colour

 

Oil

 

Foam

 

Floating particles

 

Very large particles

 

Rapid sedimentation

 

Chemical reaction

 

Biological material

 

Compare the portable result with a laboratory method when the sample composition is unusual.

 

Cause 11: Weak Battery or Power Problem

 

Unstable or insufficient power may affect instrument operation.

 

Check:

 

Battery condition

 

Battery installation

 

Battery contacts

 

Power adaptor

 

Cable connection

 

Display warnings

 

Replace batteries with the correct type when required.

 

Cause 12: The Optical Compartment Is Obstructed

 

Dust, moisture or residue inside the measuring compartment may interfere with the instrument.

 

Do not pour water directly into the meter or insert a wet bottle.

 

If liquid has entered the instrument, stop using it and arrange an inspection.

 

Quick Troubleshooting Test

 

Use this procedure to determine whether the low reading is caused by the sample or meter:

 

Inspect and clean both measuring bottles.

 

Prepare one well-mixed sample.

 

Fill both bottles from the same sample.

 

Remove air bubbles.

 

Clean and dry the outside surfaces.

 

Measure both bottles using the same timing.

 

Repeat each reading.

 

Compare with a known or laboratory-tested sample where available.

 

If the instrument repeatedly produces low results using a suitable reference sample, calibration or inspection may be required.

 

When Should You Stop Using the Meter?

 

Arrange further checking when:

 

A known reference repeatedly reads zero

 

Verification fails

 

The instrument was dropped

 

Water entered the meter

 

Different bottles produce unexplained results

 

Calibration cannot be completed

 

The meter differs significantly from reference data

 

The display or controls are malfunctioning

 

Document the problem and recent measurement history before requesting technical assistance.

 

About the Qiwei SS-500B

 

The SS-500B provides:

 

Measurement range: 0–500 mg/L

 

Displayed resolution: 1 mg/L

 

AC adaptor or AA-battery operation

 

Two measuring bottles

 

Portable field and indoor use

 

It is intended for rapid suspended-solids screening and treatment-process comparison.

 

Contact MTM Precision

 

For the Qiwei SS-500B Portable Suspended Solids Meter, troubleshooting assistance, water-quality instruments or quotation, contact:

 

MTM Precision Sdn. Bhd.

Professional Test & Measurement Solutions Partner in Malaysia

 

Website: www.mtmpre.com.my

Email: mtmpre@yahoo.com

WhatsApp: +6016-660 7346

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Malaysia Coverage:

Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kedah, Perlis, Terengganu, Sabah and Sarawak.


 

15 Sep 2026