How to Measure Pour Over and Filter Coffee TDS with PAL-401 Malaysia

How to Measure Pour Over and Filter Coffee TDS with PAL-401

 

Pour-over and filter-coffee recipes can change significantly when the grind size, coffee dose, water quantity, pouring method or brewing time varies.

 

The Qi Wei PAL-401 Coffee Concentration Meter provides an entry-level method for comparing filter-coffee strength using a small sample of approximately 0.3mL.

 

It can help cafés, baristas and coffee trainers check whether different brews are producing similar concentration results.

 

Quick Measurement Procedure

 

To measure pour-over coffee TDS:

 

Record the dry coffee dose.

 

Record the brewing-water quantity.

 

Weigh the final coffee beverage.

 

Mix the complete beverage gently.

 

Allow the sample temperature to stabilise.

 

Check that the PAL-401 sample stage is clean and dry.

 

Apply approximately 0.3mL of coffee.

 

Measure the sample two or three times.

 

Record the TDS result.

 

Compare the reading with the recipe and taste.

 

Why Measure Filter-Coffee TDS?

 

Two cups prepared using the same coffee beans may taste different because of changes in the brewing process.

 

TDS provides a numerical way to compare beverage concentration.

 

It can help answer questions such as:

 

Is today’s coffee weaker than yesterday’s?

 

Did the grinder adjustment change the beverage strength?

 

Are different baristas producing similar results?

 

Did the new pouring method increase extraction?

 

Is the coffee-to-water ratio consistent?

 

Does the same recipe work with a different roast?

 

TDS does not replace tasting, but it helps identify changes that may not be obvious from the recipe alone.

 

PAL-401 Specifications for Filter Coffee

 

SpecificationPAL-401Measurement range0–22.0%Resolution0.1%Sample volumeApproximately 0.3mLMeasurement timeApproximately 3 secondsPower supplyTwo AAA batteriesApproximate weight100gEnvironmental temperature stated by manufacturer20–23°C

 

The PAL-401 is best used for general comparison and larger concentration differences.

 

Because filter coffee normally has a relatively low TDS, users requiring detailed two-decimal-place analysis should consider a higher-resolution coffee refractometer.

 

Step 1: Document the Brewing Recipe

 

Before brewing, record:

 

Coffee bean

 

Roast date

 

Dry coffee dose

 

Grind setting

 

Water quantity

 

Water temperature

 

Brewing method

 

Filter type

 

Pouring sequence

 

Total brewing time

 

Without this information, it may be difficult to identify why the TDS reading has changed.

 

Step 2: Weigh the Final Beverage

 

Brewing-water quantity and final beverage weight are not always the same.

 

Some water remains inside the coffee grounds and filter after brewing.

 

Weighing the final beverage provides a better value for extraction-yield calculation.

 

For example:

 

Brewing water: 320g

 

Dry coffee dose: 20g

 

Final beverage: 285g

 

Use the 285g beverage weight, not the original 320g water quantity, when applying the simplified extraction-yield formula.

 

Step 3: Mix the Complete Beverage

 

Filter coffee may not be completely uniform immediately after brewing.

 

The coffee collected at the beginning of extraction is normally more concentrated than the liquid collected near the end.

 

Mix the complete beverage gently before taking the sample. This helps ensure that the sample represents the entire brew.

 

Do not measure only:

 

The first coffee leaving the dripper

 

The final dripping liquid

 

Coffee taken only from the top of the server

 

Step 4: Allow the Sample Temperature to Stabilise

 

The manufacturer’s product information states an environmental temperature of approximately 20–23°C.

 

Very hot samples may continue changing as the temperature stabilises.

 

For consistent comparison:

 

Transfer a small amount into a clean sample container.

 

Allow it to cool under similar conditions.

 

Measure every recipe at approximately the same sample temperature.

 

Avoid comparing one hot sample with another sample measured at room temperature.

 

Step 5: Prepare the PAL-401

 

Before adding the coffee:

 

Clean the optical sample stage.

 

Remove old coffee oils and residue.

 

Make sure no cleaning water remains.

 

Check the zero setting according to the operating instructions.

 

Avoid touching the optical surface.

 

Any remaining water can dilute the sample, while dried coffee residue may increase the displayed result.

 

Step 6: Apply the Coffee Sample

 

Using a clean dropper or pipette, place approximately 0.3mL of filter coffee onto the sample stage.

 

Ensure that:

 

The optical surface is fully covered.

 

No large air bubbles are present.

 

The sample does not contain visible coffee grounds.

 

The liquid does not overflow into the instrument body.

 

Filter coffee prepared with a paper filter generally contains fewer suspended particles than coffee prepared with a metal filter.

 

Step 7: Repeat the Measurement

 

Take two or three readings from the same mixed coffee sample.

 

Example:

 

First reading: 1.2%

 

Second reading: 1.3%

 

Third reading: 1.3%

 

The coffee may be recorded as approximately 1.3% for general comparison.

 

A small change between adjacent values can occur because the PAL-401 displays results in 0.1% increments.

 

Step 8: Calculate Extraction Yield

 

A simplified formula is:

 

Extraction Yield (%) = Beverage Weight × TDS (%) ÷ Dry Coffee Dose

 

Example:

 

Dry coffee dose: 20g

 

Final beverage weight: 300g

 

PAL-401 reading: 1.3%

 

Calculation:

 

300 × 1.3% ÷ 20 = 19.5%

 

The estimated extraction yield is 19.5%.

 

How Brewing Variables Affect TDS

 

Brewing ChangePossible TDS EffectHigher coffee dose with the same waterMay increase concentrationLower coffee dose with the same waterMay decrease concentrationFiner grindMay increase extraction and concentrationCoarser grindMay reduce extractionLonger brewing timeMay increase extractionShorter brewing timeMay reduce extractionLess final beverageMay produce a stronger drinkMore final beverageMay produce a more diluted drinkIncreased agitationMay increase extractionUneven pouringMay produce inconsistent results

 

These are general tendencies. The actual result depends on how the variables interact.

 

Example: Comparing Three Pour-Over Recipes

 

RecipeCoffee DoseBeverage WeightTDSEstimated ExtractionRecipe A20g280g1.2%16.8%Recipe B20g300g1.3%19.5%Recipe C20g320g1.2%19.2%

 

Recipe B has the highest concentration, while Recipes B and C produce relatively similar estimated extraction yields.

 

The preferred recipe should be selected by combining the measurement results with taste.

 

What If the Filter Coffee TDS Is Too Low?

 

A low reading may indicate a weak beverage or low extraction.

 

Check:

 

Coffee dose

 

Final beverage weight

 

Grind size

 

Brewing time

 

Water temperature

 

Pouring speed

 

Coffee-bed saturation

 

Channeling around the filter

 

Measurement procedure

 

Do not immediately change several variables together. Adjust one variable at a time and measure again.

 

What If the TDS Is Too High?

 

A high reading may indicate a concentrated beverage, but it does not automatically mean the coffee is over-extracted.

 

Check:

 

Whether the correct coffee dose was used

 

Whether the beverage weight is lower than expected

 

Whether the grind size is too fine

 

Whether the brew time is longer than normal

 

Whether residue remained on the PAL-401 sample stage

 

Taste the coffee before deciding whether the recipe requires adjustment.

 

Paper Filter vs Metal Filter Measurement

 

Paper-filter coffee normally contains fewer oils and suspended fine particles.

 

Metal-filter coffee may contain more:

 

Coffee oils

 

Fine particles

 

Sediment

 

These components can reduce measurement repeatability.

 

When comparing brewing methods, use a consistent sampling and filtration procedure. Do not assume that an unfiltered metal-filter sample can be directly compared with a cleaner paper-filter sample without considering the preparation difference.

 

Common Measurement Mistakes

 

Measuring Without Mixing

 

The sample may not represent the complete beverage.

 

Using Brewing-Water Weight in the Formula

 

Use the final beverage weight for the simplified extraction calculation.

 

Measuring at Different Temperatures

 

Changing temperature conditions may affect comparison consistency.

 

Leaving Water on the Sample Stage

 

Cleaning water can dilute the next sample.

 

Expecting Two-Decimal-Place Detail

 

The PAL-401 resolution is 0.1%. It is designed mainly for basic comparison.

 

Relying Only on TDS

 

A target number cannot replace sensory evaluation.

 

Frequently Asked Questions

 

Can PAL-401 measure pour-over coffee?

 

Yes. It can be used for basic concentration comparison of pour-over and other filter-coffee samples.

 

Must filter coffee be filtered again before measurement?

 

Paper-filter coffee normally requires less additional preparation. Metal-filter or immersion coffee may contain particles that affect repeatability.

 

What is a typical filter-coffee TDS?

 

A commonly referenced range is approximately 1.15–1.35%, but the preferred value depends on the recipe, coffee and intended taste.

 

Why do I get both 1.2% and 1.3%?

 

The actual sample may be near the boundary between the two displayed values. PAL-401 has a 0.1% resolution.

 

Can I adjust the grinder based only on TDS?

 

No. Consider TDS, extraction yield, brewing time, beverage weight and taste before changing the grinder.

 

PAL-401 Coffee Concentration Meter Available in Malaysia

 

The Qi Wei PAL-401 Coffee Concentration Meter is available from MTM Precision for cafés, coffee roasters, barista training centres and food and beverage users throughout Malaysia.

 

Customers may visit our Puchong showroom to view the instrument and discuss suitable measurement procedures for pour-over, filter coffee or espresso.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd.

 

Professional Test & Measurement Instruments • Technical Consultation • Product Demonstration • After-Sales Support

 

MTM Precision supplies professional testing and measurement instruments for industrial, laboratory, engineering, water-quality and food and beverage applications throughout Malaysia.

 

Website: www.mtmpre.com.my

WhatsApp: +6016-660 7346

Email: mtmpre@yahoo.com

 

Showroom & Service Centre

 

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Customers are welcome to visit our Puchong showroom to view selected products, discuss their applications and identify a suitable testing solution with our team.


 

15 Sep 2026