Why Does a Four-Gas Detector Show 20.9% Oxygen Malaysia
Why Does a Four-Gas Detector Show 20.9% Oxygen Malaysia
When the Smart Sensor ST8904 or another four-gas detector is switched on in fresh air, its oxygen channel will normally display approximately 20.9% VOL.
Many first-time users ask whether this means the detector has found an oxygen leak or whether the reading should be zero.
The answer is simple:
Unlike toxic-gas channels, the oxygen channel is expected to show the percentage of oxygen naturally present in the surrounding air.
Why Is the Normal Oxygen Reading About 20.9%?
Normal dry air at sea level contains approximately:
78% nitrogen
20.9% oxygen
Small amounts of argon, carbon dioxide and other gases
Therefore, a correctly functioning portable gas detector will generally show an oxygen reading near 20.9% in suitable fresh air.
The other ST8904 channels may normally display approximately:
CO: 0 ppm
H₂S: 0 ppm
LEL: 0%
O₂: Approximately 20.9% VOL
Oxygen is different because it is naturally present in the atmosphere.
Is 20.9% an Oxygen Leak?
No. A reading around 20.9% in normal fresh air is expected.
An oxygen leak or oxygen-enriched atmosphere would cause the reading to increase above the normal atmospheric level.
Oxygen enrichment can increase fire and combustion risks. It should not be treated as harmless simply because oxygen is necessary for breathing.
What Does a Low Oxygen Reading Mean?
A lower-than-expected oxygen reading may indicate that oxygen has been:
Displaced by another gas
Consumed by combustion
Consumed by biological activity
Reduced by oxidation or corrosion
Affected by a chemical reaction
Diluted inside an enclosed space
Potential locations include:
Storage tanks
Manholes
Sewers
Underground chambers
Process vessels
Silos
Wastewater facilities
Poorly ventilated rooms
A low oxygen reading can also result from sensor drift, incorrect zeroing or sensor deterioration. Always investigate before assuming the reading is safe or incorrect.
Can a Normal Oxygen Reading Prove the Area Is Safe?
No.
A reading near 20.9% oxygen does not prove that the atmosphere is free from every hazard.
For example:
Carbon monoxide may still be present
Hydrogen sulphide may still be present
A combustible gas may still be present
A specialised toxic gas may be present
The atmosphere may change after work starts
This is why the Smart Sensor ST8904 monitors O₂ together with CO, H₂S and LEL.
The detector must also be suitable for the gases expected at the site.
Why Does the ST8904 Show 20.8% or 21.0% Instead?
A small difference from 20.9% does not automatically indicate a fault.
The displayed value can be influenced by:
Sensor tolerance
Calibration condition
Temperature
Humidity
Atmospheric pressure
Altitude
Sensor age
Display resolution
Actual environmental conditions
However, a reading that is significantly different, unstable or unable to return to its expected fresh-air value should be investigated.
Why Does the Oxygen Reading Drop When Another Gas Is Present?
When another gas displaces normal air, the percentage of oxygen may decrease.
Possible examples include:
Nitrogen entering a space
Carbon dioxide accumulation
Fuel gas leakage
Process-gas release
Decomposition gas inside a tank or sewer
The oxygen sensor may show that oxygen has been displaced, but it does not identify the gas responsible.
A normal-looking reading on the CO, H₂S and LEL channels does not exclude a different gas that the detector cannot measure.
Should the Oxygen Channel Be Zeroed to 0%?
No. Do not attempt to make the oxygen channel display zero in fresh air.
The fresh-air reference for oxygen is normally around the atmospheric oxygen concentration, while the toxic and combustible channels generally use a zero baseline in verified clean air.
Incorrect zeroing can produce misleading readings and unsafe decisions.
Follow the operating procedure for the specific detector.
Where Should Fresh-Air Zeroing Be Performed?
Perform fresh-air zeroing only in an area known to contain clean, uncontaminated air.
Do not zero the detector:
Inside the space being tested
Near vehicle exhaust
Beside an open chemical container
Near fuel or solvent vapour
Inside a poorly ventilated room
Where the atmosphere is already uncertain
Zeroing in contaminated air may cause the detector to treat a hazardous concentration as its baseline.
What If the ST8904 Shows 19% Oxygen in Open Air?
Move the detector to a location independently confirmed to contain suitable fresh air and allow the reading to stabilise.
Then check:
Was the detector switched on in contaminated air?
Was it zeroed incorrectly?
Is the sensor opening blocked?
Has the detector been exposed to water or condensation?
Is the oxygen sensor ageing?
Has the detector passed its bump test?
Is calibration overdue?
Does another detector show the same result?
Do not use the detector for safety-critical work if the oxygen channel cannot be verified.
Why Does the Reading Change Inside a Manhole or Tank?
A confined space may have different atmospheric conditions at different depths.
The oxygen concentration near the opening may differ from the concentration at the middle or bottom of the space.
Possible reasons include:
Gas density
Inadequate ventilation
Connected pipelines
Decomposition
Chemical reactions
Sludge disturbance
Process-gas entry
Testing only at the opening may not represent the complete atmosphere.
Does Ventilation Immediately Make the Space Safe?
Not necessarily.
Ventilation may improve the atmosphere, but the space should be retested according to the approved procedure.
The team must also consider:
Whether ventilation reaches the entire space
Whether gas continues entering
Whether work will create new hazards
Whether ventilation could fail
Whether continuous monitoring is required
A single acceptable reading before work begins does not guarantee that conditions will remain safe.
Bump Test and Calibration of the Oxygen Sensor
A bump test exposes the oxygen sensor to an appropriate test condition to confirm that it responds and activates the required alarm.
Calibration compares the detector with a known reference and adjusts the reading when necessary.
The oxygen channel may require further inspection when:
It fails a bump test
It cannot be calibrated
Response becomes slow
Reading fluctuates excessively
Fresh-air reading remains abnormal
A sensor fault appears
The detector has experienced impact, water or chemical exposure
The oxygen sensor is a consumable component and may eventually require replacement.
Smart Sensor ST8904 Oxygen Measuring Range
The ST8904 oxygen channel measures:
0–30% VOL O₂
The instrument also monitors:
CO: 0–999 ppm
H₂S: 0–500 ppm
Combustible gas: 0–100% LEL
This combination is commonly used for confined-space, wastewater, manhole and factory-maintenance applications.
Need Help with an Abnormal Oxygen Reading?
MTM Precision supplies the Smart Sensor ST8904 four-in-one gas detector in Malaysia.
When requesting assistance, provide:
Actual oxygen reading
Location of the test
Fresh-air reading
Last bump-test result
Last calibration date
Sensor age, if known
Photos or video of the display
Whether the detector was exposed to water, impact or high gas concentration
Do not rely on an instrument for confined-space entry when its oxygen reading cannot be verified.
Contact MTM Precision
MTM Precision Sdn. Bhd.
Office: +603-8080 7172
Desmond: +6019-220 7346
Fifi Ng: +6016-660 7346
WhatsApp: +6019-220 7346
Email: enquiry@mtmpre.com.my
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.
Serving customers throughout Selangor, Kuala Lumpur, Johor, Penang, Perak, Melaka, Negeri Sembilan, Pahang, Kedah, Perlis, Terengganu, Sabah and Sarawak.
16 Sep 2026