Common Push-Pull Force Testing Mistakes and How to Avoid Them Malaysia

Common Push-Pull Force Testing Mistakes and How to Avoid Them Malaysia

 

Push-pull force testing may appear straightforward: install the sample, apply a load and record the result. In practice, incorrect range selection, poor alignment, unsuitable grips and inconsistent loading can significantly affect the measurement.

 

The Landtek FM-207-500K Digital Force Gauge measures tension and compression force up to ±500 kgf, ±4900 N or ±1100 lbf.

 

Its real-time and peak-value modes support proof-load, pull-out, breaking-force and compression tests. However, reliable results still depend on the complete test setup and operating method.

 

Mistake 1: Choosing the Wrong Force Range

 

A force gauge should be selected according to the expected minimum and maximum test force.

 

A gauge with insufficient capacity may be overloaded. A gauge with a much higher capacity than required may provide less useful resolution for a small-force test.

 

For example, the FM-207-500K may be suitable for tests reaching several hundred kilograms-force, but it may not be the best choice for accurately evaluating a force of only 1 kgf or 2 kgf.

 

How to Avoid It

 

Before selecting a model, determine:

 

Normal operating force

 

Expected maximum force

 

Possible breaking force

 

Required resolution

 

Desired safety margin

 

Choose a measuring range that matches the actual application.

 

Mistake 2: Applying Force at an Angle

 

A force sensor should normally be loaded along its intended measurement axis.

 

Angular pulling or uneven compression can introduce:

 

Side loading

 

Bending

 

Twisting

 

Unstable readings

 

Premature sample failure

 

Sensor damage

 

How to Avoid It

 

Align the sensor, fixture and sample before beginning the test.

 

Use suitable adapters or locating fixtures to keep the applied force centred. Stop the test if the sample or fixture moves out of alignment.

 

Mistake 3: Using an Unstable Test Stand

 

A moving or bending test stand can affect the reading and create a safety risk.

 

A stand may appear strong enough at a low load but deform as the force increases.

 

How to Avoid It

 

Confirm the rated capacity of:

 

Test frame

 

Base

 

Crosshead

 

Sensor mount

 

Loading mechanism

 

Connection points

 

The stand should be securely mounted and rated for the intended test force.

 

Mistake 4: Using an Incorrect Grip

 

The grip transfers force to the sample. If it does not match the product, the sample may slip, become damaged or fail at the wrong location.

 

For example:

 

A cable may slip from a smooth clamp

 

A sharp grip may cut a soft sample

 

A hook may rotate under load

 

A thin component may be crushed before the test begins

 

How to Avoid It

 

Select the grip based on:

 

Sample shape

 

Material

 

Surface

 

Diameter or thickness

 

Test direction

 

Expected force

 

Intended failure mode

 

Possible solutions include wedge grips, capstan grips, cable clamps, hooks, clevises, compression plates and customised fixtures.

 

Mistake 5: Ignoring the Capacity of Small Connectors

 

The main test stand and sensor may support 500 kgf, but a small bolt, hook, clamp or adapter may not.

 

The complete testing system is only as strong as its weakest component.

 

How to Avoid It

 

Check every load-bearing component, including:

 

Bolts

 

Nuts

 

Hooks

 

Shackles

 

Links

 

Plates

 

Adapters

 

Mounting threads

 

Do not use a connector simply because it fits physically. Its rated load must also be suitable.

 

Mistake 6: Forgetting to Zero the Gauge

 

The weight of a fixture, preload from the sample or residual tension may appear on the display before the test begins.

 

If the instrument is not zeroed, this initial force may be included in the final result.

 

How to Avoid It

 

After installing the sensor and fixture:

 

Remove any unintended load

 

Allow the reading to stabilise

 

Press the zero function

 

Confirm that the display remains near zero

 

Start the test

 

Repeat the zero check before every new series of measurements.

 

Mistake 7: Using the Wrong Measurement Mode

 

Real-time and peak modes serve different purposes.

 

Using real-time mode for a fast breaking event may cause the operator to miss the maximum force. Using peak mode for a holding test may show the maximum value but not the current load.

 

How to Avoid It

 

Use real-time mode for:

 

Proof-load testing

 

Holding a specified force

 

Force adjustment

 

Monitoring gradual changes

 

Compression at a defined position

 

Use peak mode for:

 

Breaking-force testing

 

Pull-out testing

 

Release-force testing

 

Maximum compression

 

Destructive testing

 

Clear the previous peak value before starting the next test.

 

Mistake 8: Applying the Load Too Quickly

 

Rapid or jerking movement can create an impact force that is higher than the intended test load.

 

It can also reduce repeatability and cause unpredictable failure.

 

How to Avoid It

 

Apply the force gradually and use a consistent loading speed.

 

If a test standard specifies a crosshead or loading speed, use a suitable motorised or controlled test system.

 

Do not strike, drop or suddenly pull the sensor.

 

Mistake 9: Changing the Loading Speed Between Samples

 

Some materials respond differently when loaded quickly or slowly.

 

This is particularly relevant for:

 

Rubber

 

Plastic

 

Adhesives

 

Textiles

 

Flexible cables

 

Packaging materials

 

Soft components

 

How to Avoid It

 

Use the same loading speed for all samples being compared.

 

Document the test speed or loading method so another operator can repeat the procedure.

 

Mistake 10: Allowing the Sample to Slip

 

If a cable, rope or component slips from its grip, the recorded peak may represent grip failure rather than product strength.

 

How to Avoid It

 

Before performing the full test:

 

Check the grip surface

 

Apply a small preload

 

Confirm sample alignment

 

Watch for movement

 

Adjust or replace the grip if necessary

 

Do not increase clamping pressure excessively if doing so damages the sample.

 

Mistake 11: Misinterpreting the Failure Mode

 

Two samples may produce the same peak force but fail in different ways.

 

For example:

 

One wire breaks in the centre

 

Another wire pulls out of its terminal

 

One fastener breaks

 

Another pulls the surrounding material out

 

One cable fails normally

 

Another slips from the grip

 

How to Avoid It

 

Record both:

 

Maximum force

 

Type and location of failure

 

Photograph the sample after the test where appropriate.

 

Mistake 12: Treating One Test as Conclusive

 

A single result does not show production variation.

 

One sample may be unusually strong or weak.

 

How to Avoid It

 

Test multiple samples under the same conditions.

 

Compare:

 

Average result

 

Minimum result

 

Maximum result

 

Variation

 

Failure mode

 

Production batch

 

Follow the required sampling plan where a standard or customer specification applies.

 

Mistake 13: Treating Breaking Force as Safe Working Load

 

Breaking force is the force recorded when a sample fails. It is not automatically the maximum safe force for normal use.

 

A product may require a safety factor based on:

 

Applicable standard

 

Material

 

Repeated loading

 

Shock loading

 

Temperature

 

Corrosion

 

Wear

 

Application risk

 

How to Avoid It

 

Do not assign a safe working load based only on one breaking-force result.

 

Consult the relevant engineering requirement, safety factor and certification procedure.

 

Mistake 14: Continuing After an Overload Warning

 

The FM-207-500K provides an overload warning when excessive force is applied.

 

Continuing to increase the load can permanently affect the sensor.

 

How to Avoid It

 

If an overload warning appears:

 

Stop applying force

 

Remove the load gradually

 

Inspect the fixture

 

Confirm the expected force

 

Select a higher-capacity system if necessary

 

Do not treat the sensor’s safe-load specification as an additional measurement range.

 

Mistake 15: Pulling or Trapping the Sensor Cable

 

The electrical cable connecting the external sensor to the display is not designed to carry the test load.

 

A damaged or trapped cable may cause unstable readings or communication failure.

 

How to Avoid It

 

Route the cable away from:

 

Moving fixtures

 

Sharp edges

 

Pinch points

 

Hydraulic parts

 

Hot surfaces

 

The sample’s failure path

 

Provide enough cable movement without allowing it to interfere with the test.

 

Mistake 16: Ignoring Environmental Conditions

 

Vibration, unstable surfaces and corrosive environments may affect the test system or damage components.

 

How to Avoid It

 

Install the equipment on a stable structure away from unnecessary vibration.

 

Keep the instrument and sensor within the manufacturer’s specified operating conditions and protect them from corrosive media.

 

Mistake 17: Failing to Check Calibration

 

A force gauge may continue to display readings even when its measurement performance has changed.

 

Possible causes include:

 

Overload

 

Impact

 

Long-term use

 

Sensor damage

 

Harsh operating conditions

 

Previous repair

 

How to Avoid It

 

Establish a calibration or verification schedule based on:

 

Usage frequency

 

Application risk

 

Customer requirements

 

Quality procedure

 

History of overload

 

Applicable standards

 

If results become inconsistent or the sensor has been overloaded, arrange inspection before continuing critical tests.

 

Basic Troubleshooting Guide

 

ProblemPossible causeRecommended checkReading does not return to zeroResidual load or fixture bindingRemove the load and inspect alignmentReading fluctuatesVibration, loose mounting or sample movementStabilise the setup and tighten fixturesPeak result is unusually highImpact or sudden loadingRepeat using controlled loading speedPeak result is unusually lowSample slipped or grip failedInspect and replace the gripResults vary between operatorsDifferent speed or sample positioningStandardise the test procedureOverload warning appearsForce exceeds the selected rangeStop and use a higher-capacity systemNo sensor readingLoose or damaged sensor connectionSwitch off and inspect the connectionSample breaks beside gripExcessive clamp pressure or sharp edgeChange the grip or clamping method

 

Do not dismantle the sensor or indicator unless servicing is performed by qualified personnel.

 

How to Improve Test Repeatability

 

For more consistent results:

 

Use the same sample preparation method

 

Use the same fixture

 

Maintain the same sensor alignment

 

Apply the same loading speed

 

Use the correct measurement mode

 

Zero the gauge before each test

 

Follow the same stopping criterion

 

Record the failure mode

 

Inspect fixtures regularly

 

Maintain calibration records

 

A written procedure reduces differences between operators and production shifts.

 

Frequently Asked Questions

 

Why does my reading fluctuate?

 

Possible causes include vibration, hand movement, loose mounting, sample slippage, sensor misalignment or an unstable stand.

 

Why is the peak value higher than expected?

 

The load may have been applied too quickly, creating an impact peak. Check the loading method and repeat the test gradually.

 

Why does the gauge not return to zero?

 

Check for residual tension, fixture binding, sensor misalignment or damage caused by overload.

 

Can I continue using the gauge after an overload?

 

Remove the load and check its performance. If zero, repeatability or accuracy appears abnormal, arrange inspection or calibration before critical use.

 

Why do identical samples produce different results?

 

Differences in material, production, sample preparation, grip position, loading speed and failure mode can all affect the result.

 

Can calibration correct a poor test fixture?

 

No. Calibration verifies the measurement system, but it cannot correct sample slippage, misalignment or an unstable test stand.

 

Get Support for Force Measurement in Malaysia

 

When requesting assistance, send MTM Precision:

 

Force-gauge model

 

Photo or video of the setup

 

Sample details

 

Expected force

 

Actual reading

 

Measurement mode

 

Error or overload indication

 

Test-stand and fixture information

 

Calibration history

 

This information helps distinguish between an instrument issue, fixture problem and incorrect testing method.

 

MTM Precision supplies the Landtek FM-207-500K Digital Force Gauge for high-capacity push, pull, tension and compression-force applications.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd. (744811-A)

 

Telephone: +603-8080 7172

WhatsApp: +6019-665 7346 / +6016-660 7346

Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my

Website: www.mtmpre.com.my

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar Lapan Belas,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Supply and technical support are available throughout Malaysia, including Selangor, Kuala Lumpur, Johor, Penang, Perak, Melaka, Negeri Sembilan, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sarawak and Sabah.


 

16 Sep 2026