UNI-T UT345A Reading Unstable or Cannot Get Reading Malaysia

UNI-T UT345A Reading Unstable or Cannot Get Reading Malaysia

 

Is your UNI-T UT345A Ultrasonic Thickness Gauge showing an unstable reading, jumping between different thickness values or not displaying a measurement at all?

 

This does not automatically mean the instrument is faulty.

 

Ultrasonic thickness measurement depends on several factors including couplant, probe contact, surface condition, material, sound velocity, geometry and instrument setup.

 

Before sending the UT345A for checking, there are several basic troubleshooting steps you can try.

 

Common UT345A Problems

 

Users may experience symptoms such as:

 

No thickness reading

 

Reading appears and disappears

 

Thickness value keeps changing

 

Measurement is much higher than expected

 

Measurement is much lower than expected

 

Same location gives different readings

 

Works on one material but not another

 

Works on a flat plate but has difficulty on a pipe

 

Each symptom can have more than one possible cause.

 

Problem 1: No Reading

 

You place the UT345A probe on the metal, but no thickness value appears.

 

First check the coupling condition.

 

Ultrasonic energy needs to travel from the probe into the material.

 

If there is an air gap between the probe and surface, the ultrasonic signal may not transmit effectively.

 

What to Check

 

Is suitable ultrasonic couplant being used?

 

Is there enough couplant?

 

Is the probe making proper contact?

 

Is the surface excessively rough?

 

Is loose rust preventing contact?

 

Is the material suitable for ultrasonic measurement?

 

Apply suitable couplant and try again on a clean, accessible measurement location.

 

Problem 2: Reading Keeps Jumping

 

For example:

 

5.82 mm → 4.71 mm → 6.05 mm → No Reading

 

Do not immediately record one of these values.

 

A good thickness measurement should be sufficiently stable and repeatable for the intended procedure.

 

Possible causes include:

 

Poor probe contact

 

Insufficient couplant

 

Rough surface

 

Heavy corrosion

 

Pipe curvature

 

Probe movement

 

Incorrect instrument settings

 

Remove the probe, check the surface, reapply couplant and repeat the measurement.

 

Problem 3: Same Point Gives Different Measurements

 

Suppose you measure the same location three times:

 

5.8 mm

 

5.1 mm

 

6.0 mm

 

This level of variation should prompt troubleshooting before the result is accepted.

 

Check whether you are placing the probe at exactly the same location and with consistent contact.

 

On a rough or corroded surface, even a small change in probe position can affect the ultrasonic signal.

 

Problem 4: No Couplant

 

This is one of the first things to check.

 

The UT345A is an ultrasonic thickness gauge.

 

It needs effective acoustic coupling between the probe and material.

 

Placing the probe dry against a metal surface can leave microscopic air gaps.

 

The measurement path should effectively be:

 

Probe → Couplant → Material → Back Wall → Echo → Probe

 

Without good coupling, obtaining a reliable back-wall echo can become difficult.

 

Problem 5: Too Much Rust

 

A rusty pipe may look like an ideal application for a thickness gauge because corrosion is exactly what you want to investigate.

 

However, severe surface corrosion can also make ultrasonic measurement more difficult.

 

Problems can include:

 

Loose rust

 

Scale

 

Pitting

 

Uneven surface

 

Poor probe contact

 

The measurement area may need appropriate surface preparation according to the inspection procedure.

 

Do not simply press harder against loose corrosion and assume the displayed value is correct.

 

Problem 6: Wrong Sound Velocity

 

Ultrasonic thickness measurement depends on the speed of sound through the material.

 

Different materials have different sound velocities.

 

The UT345A provides adjustable sound velocity of approximately:

 

1000–9999 m/s

 

If the wrong velocity is selected, the instrument may display an incorrect thickness even if the reading appears stable.

 

Therefore, always confirm the material and appropriate sound-velocity setting.

 

Problem 7: Measuring a Small Pipe

 

A flat steel plate provides a relatively simple probe-contact surface.

 

A pipe is curved.

 

As pipe diameter becomes smaller, obtaining good probe contact can become more challenging depending on the probe and application.

 

If the UT345A works correctly on a suitable flat reference but gives unstable results on a particular pipe, consider:

 

Pipe diameter

 

Probe orientation

 

Surface curvature

 

Couplant

 

Surface condition

 

Do not immediately assume the meter has failed.

 

Problem 8: Measuring Through Paint

 

A painted surface can complicate ultrasonic thickness measurement.

 

Depending on the coating, instrument and measurement method, the coating may influence the result.

 

If accurate base-metal thickness is required, follow the manufacturer's instructions and the applicable inspection procedure.

 

Do not automatically assume:

 

Displayed thickness = steel thickness only

 

when measuring through an unknown coating.

 

Problem 9: Reading Is Much Higher Than Expected

 

Suppose you expect approximately:

 

6 mm

 

but the instrument shows a substantially different value.

 

Possible areas to investigate include:

 

Incorrect sound velocity

 

Material identification

 

Measurement setup

 

Coating

 

Surface condition

 

Probe coupling

 

Instrument verification

 

Do not simply adjust settings until the number looks correct.

 

The settings should be based on the actual material and appropriate procedure.

 

Problem 10: Reading Is Much Lower Than Expected

 

A low reading can be important because it may indicate material loss.

 

But first make sure the measurement is valid.

 

For example:

 

Nearby points:

 

6.0 mm

 

5.9 mm

 

6.1 mm

 

One point:

 

3.8 mm

 

Before concluding that severe wall thinning exists:

 

Reapply couplant.

 

Repeat the measurement.

 

Check probe positioning.

 

Verify instrument settings.

 

Check the surface.

 

Measure around the location.

 

If the low reading remains stable and repeatable, it may require further investigation.

 

Quick UT345A Troubleshooting Sequence

 

When the UNI-T UT345A cannot obtain a reliable reading, work through these checks systematically.

 

1. Check the Material

 

Is the material suitable for the ultrasonic thickness measurement method?

 

2. Check the Surface

 

Is it clean enough for good probe contact?

 

3. Check Couplant

 

Is suitable ultrasonic couplant being used?

 

4. Check Probe Contact

 

Is the probe positioned properly?

 

5. Check Sound Velocity

 

Does the setting correspond to the material?

 

6. Check Curvature

 

Are you measuring a small or highly curved component?

 

7. Repeat the Reading

 

Can the result be reproduced?

 

8. Verify the Instrument

 

Follow UNI-T's recommended verification or calibration procedure.

 

This sequence is more useful than immediately assuming the instrument is defective.

 

Simple Test Before Sending the UT345A for Checking

 

If possible, test the UT345A on a suitable material or reference whose thickness is reasonably known.

 

Apply suitable couplant and take several measurements at the same location.

 

If the meter produces stable and repeatable results there but struggles on your actual component, the problem may be related to the application or surface condition rather than the instrument.

 

If it cannot produce appropriate measurements on a suitable reference under correct operating conditions, further checking may be required.

 

UT345A Works on Steel Plate but Not on Corroded Pipe

 

This is a particularly useful diagnostic clue.

 

A clean flat steel surface provides:

 

Better probe contact

 

More uniform geometry

 

Less surface interference

 

A corroded pipe may introduce:

 

Curvature

 

Rust

 

Pitting

 

Roughness

 

Coating

 

Uneven probe contact

 

Therefore, successful measurement on a suitable flat reference does not guarantee that every corroded pipe will be equally easy to measure.

 

UT345A Works on One Material but Not Another

 

Do not assume every solid material behaves identically under ultrasonic measurement.

 

Material properties can affect ultrasonic propagation.

 

Before measuring a new material, check whether it is suitable and confirm the appropriate measurement parameters.

 

The UT345A should not be treated as a universal meter that can measure every object simply because the object has thickness.

 

Record Your Troubleshooting

 

If you need technical support, avoid sending only:

 

“UT345A cannot measure.”

 

Provide more information.

 

For example:

 

Material: Carbon steel

Expected thickness: Approximately 6 mm

Surface: Painted / rusty

Object: 2-inch pipe

Problem: Reading changes between 4–7 mm

Couplant: Used

Reference test: Stable / unstable

 

A short video showing the probe placement and instrument display can also be very useful.

 

When Should the Instrument Be Checked?

 

Further checking may be appropriate if the UT345A:

 

Cannot obtain a measurement on a suitable reference

 

Gives consistently abnormal results under correct conditions

 

Has a damaged probe or cable

 

Shows physical damage

 

Behaves inconsistently after setup and application factors have been checked

 

This helps separate an instrument problem from an application problem.

 

UNI-T UT345A Technical Support Malaysia

 

If you purchased or are using a UNI-T UT345A Ultrasonic Thickness Gauge and are unsure why it cannot obtain a stable measurement, you can send MTM Precision:

 

Photo of the UT345A

 

Photo of the probe

 

Photo of the test material

 

Material type

 

Approximate expected thickness

 

Pipe diameter if applicable

 

Surface condition

 

Short video showing the measurement problem

 

This gives us much more information for an initial assessment.

 

UNI-T UT345A Ultrasonic Thickness Gauge Malaysia

 

MTM Precision Sdn Bhd supplies UNI-T UT345A Ultrasonic Thickness Gauge and industrial measuring instruments for pipe-wall inspection, corrosion monitoring and factory-maintenance applications in Malaysia.

 

MTM Precision Sdn Bhd

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Tel: 03-8080 7172

WhatsApp: +6016-660 7346

Email: mtmpre@yahoo.com

Website: www.mtmpre.com.my

 

For UNI-T UT345A troubleshooting Malaysia, WhatsApp MTM Precision with a photo or short video of the instrument and measurement application for an initial assessment.


 

22 Sep 2026