How to Use UNI-T UT345A Ultrasonic Thickness Gauge Malaysia

How to Use UNI-T UT345A Ultrasonic Thickness Gauge Malaysia

 

The UNI-T UT345A Ultrasonic Thickness Gauge is designed to measure the thickness of suitable materials using ultrasonic technology.

 

It can be used for applications such as pipe wall thickness, steel plate measurement, storage tank inspection, corrosion monitoring and factory maintenance.

 

However, obtaining a reliable thickness reading involves more than simply placing the probe on the material.

 

Correct material identification, sound velocity, surface preparation, couplant and probe contact are all important.

 

This guide explains the basic process.

 

Step 1: Know What You Are Measuring

 

Before switching settings, identify the test object.

 

Ask:

 

What is the material?

 

What thickness do I expect?

 

Is the surface flat or curved?

 

Is it rusty?

 

Is it painted?

 

Can I access only one side?

 

Am I measuring material thickness or coating thickness?

 

The last question is especially important.

 

If you want to measure the actual pipe or steel wall, the UT345A may be the appropriate type of instrument.

 

If you want to measure paint or protective coating thickness, you normally need a coating thickness gauge, such as an appropriate UNI-T UT343-series model.

 

Step 2: Inspect the Measurement Surface

 

Look at the area where the probe will be placed.

 

Check for:

 

Loose rust

 

Scale

 

Dirt

 

Heavy corrosion

 

Flaking paint

 

Severe roughness

 

Irregular geometry

 

Good probe contact is important for ultrasonic measurement.

 

Where necessary, prepare the surface appropriately according to the applicable measurement procedure.

 

Step 3: Switch On the UT345A

 

Turn on the UNI-T UT345A and check that the instrument and probe appear to be operating normally.

 

Before relying on measurements, inspect the probe and cable for obvious damage.

 

If the instrument has not been used for some time, verification before measurement is good practice.

 

Step 4: Confirm the Material

 

Ultrasonic thickness measurement depends on the sound velocity of the material.

 

Different materials transmit ultrasonic waves at different speeds.

 

Therefore, the instrument must be configured appropriately for the material being tested.

 

The UT345A supports adjustable sound velocity of approximately:

 

1000–9999 m/s

 

Do not simply change the sound velocity until the displayed thickness matches the number you expect.

 

The setting should correspond to the actual material and appropriate procedure.

 

Step 5: Verify or Calibrate the Instrument

 

Before taking important measurements, follow the UNI-T manufacturer's recommended verification or calibration procedure.

 

This helps confirm that the instrument, probe and settings are functioning appropriately before inspecting an unknown component.

 

For critical work, follow the applicable inspection procedure and quality requirements.

 

Step 6: Apply Ultrasonic Couplant

 

Apply suitable ultrasonic couplant to the measurement location.

 

This step is important.

 

Even when the probe appears to touch the metal, microscopic air gaps may exist between the probe and surface.

 

Couplant helps provide an acoustic path:

 

Probe → Couplant → Material

 

Without suitable coupling, you may experience:

 

No reading

 

Unstable reading

 

Intermittent reading

 

Poor repeatability

 

Step 7: Place the Probe on the Surface

 

Place the probe against the measurement location with appropriate contact.

 

Try to keep the probe stable.

 

On a flat plate, probe positioning may be relatively straightforward.

 

On a curved pipe, correct positioning can require more care.

 

Do not continuously move the probe while trying to record a precise value.

 

Step 8: Wait for a Stable Reading

 

Do not immediately record the first number that appears.

 

Allow the measurement to stabilise.

 

For example, if the display shows:

 

5.82 mm → 5.81 mm → 5.82 mm

 

the measurement appears relatively repeatable.

 

If it shows:

 

5.8 mm → 4.1 mm → 6.3 mm → No Reading

 

something should be checked before accepting the result.

 

Possible causes include:

 

Couplant

 

Surface condition

 

Probe position

 

Curvature

 

Incorrect settings

 

Heavy corrosion

 

Step 9: Remove and Repeat

 

A useful basic check is to remove the probe and take another measurement at the same location.

 

For example:

 

Reading 1: 5.82 mm

 

Reading 2: 5.81 mm

 

Reading 3: 5.83 mm

 

Repeatability gives more confidence in the measurement than recording a single instantaneous value.

 

The acceptable repeatability for a particular job should follow the applicable measurement procedure.

 

Step 10: Measure Multiple Locations

 

When checking corrosion or wall thinning, do not rely on one measurement.

 

Corrosion may not be uniform.

 

For example:

 

P01: 6.10 mm

P02: 6.02 mm

P03: 5.28 mm

P04: 6.05 mm

 

P03 shows a lower measurement than the surrounding points.

 

Verify the result and investigate further according to the applicable inspection procedure.

 

How to Measure a Pipe with UT345A

 

For a suitable metal pipe:

 

1. Identify the pipe material

 

Confirm the material before selecting measurement settings.

 

2. Select the measurement area

 

Choose locations according to the maintenance or inspection objective.

 

3. Prepare the surface

 

Remove loose contamination that prevents proper probe contact.

 

4. Apply couplant

 

Ensure suitable acoustic coupling.

 

5. Position the probe

 

Take care with pipe curvature.

 

6. Obtain a stable reading

 

Do not record a rapidly fluctuating value.

 

7. Repeat

 

Verify measurement repeatability.

 

8. Measure additional points

 

Check surrounding or predefined locations where appropriate.

 

How to Measure a Steel Plate

 

A relatively clean, flat steel plate is often an easier application for learning basic ultrasonic thickness measurement.

 

After configuring and verifying the instrument:

 

Apply couplant → Place probe → Stabilise → Record → Repeat

 

If you are learning to use the UT345A, practising on a suitable material with reasonably known thickness can help you understand how probe contact and couplant affect the reading.

 

How to Check a Corroded Surface

 

Corrosion makes measurement more challenging.

 

Do not simply place the probe on top of loose rust.

 

Check for:

 

Loose corrosion products

 

Pitting

 

Scale

 

Severe roughness

 

Prepare the measurement location appropriately and take repeat measurements.

 

If one location shows an unexpectedly low value, verify it and check nearby points.

 

How to Measure a Painted Surface

 

Paint and protective coatings can affect ultrasonic thickness measurement depending on the coating, instrument and measurement method.

 

Do not automatically assume:

 

Displayed reading = base-metal thickness only

 

when measuring through an unknown coating.

 

For important measurements, follow UNI-T's instructions and the applicable inspection procedure.

 

How to Record UT345A Measurements

 

Do not record only:

 

5.62 mm

 

A better record is:

 

Asset: Cooling Water Pipe CW-01

Location: Pump Room

Point: UT-03

Material: Carbon Steel

Thickness: 5.62 mm

Instrument: UNI-T UT345A

Date: Inspection date

Surface: Prepared

Remarks: Reading repeated

 

This information makes future measurements much more useful.

 

Use the Same Measurement Points Next Time

 

One of the best uses of an ultrasonic thickness gauge is thickness trending.

 

For example:

 

2026: 6.2 mm

 

2027: 5.9 mm

 

2028: 5.6 mm

 

When measurements are obtained appropriately at repeatable locations, the history can help maintenance personnel identify changes in thickness over time.

 

Common UT345A User Mistakes

 

Mistake 1: No Couplant

 

Poor acoustic contact may result in no reading or unstable measurement.

 

Mistake 2: Wrong Sound Velocity

 

An apparently stable reading can still be wrong if the material setting is incorrect.

 

Mistake 3: Measuring Over Loose Rust

 

Poor surface contact can affect measurement quality.

 

Mistake 4: Recording One Reading Only

 

Repeatability should be checked.

 

Mistake 5: Checking Only One Point for Corrosion

 

Localised thinning may be missed.

 

Mistake 6: Confusing Coating Thickness with Metal Thickness

 

UT345A and coating thickness gauges serve different measurement purposes.

 

Mistake 7: Treating the Displayed Number as a Safety Decision

 

The UT345A measures thickness. It does not independently determine whether critical equipment is safe for continued operation.

 

What If UT345A Cannot Get a Reading?

 

Check in this order:

 

Material suitability

 

Surface condition

 

Couplant

 

Probe contact

 

Sound velocity

 

Curvature

 

Instrument verification

 

Measurement repeatability

 

If possible, test the instrument on a suitable reference material.

 

If it works on the reference but not on the actual component, the issue may relate to the application rather than an instrument fault.

 

UT345A for Preventive Maintenance

 

The UT345A can be incorporated into routine factory maintenance for suitable:

 

Pipe systems

 

Storage tanks

 

Steel components

 

Fabricated equipment

 

Selected machinery components

 

Instead of measuring randomly, create repeatable inspection points.

 

For example:

 

UT-01 – Cooling Water Pipe

 

UT-02 – Process Pipe Elbow

 

UT-03 – Storage Tank

 

UT-04 – Utility Pipe

 

The same locations can then be checked during future maintenance rounds.

 

Important Safety Note

 

For pressure vessels, pressure piping and other safety-critical equipment, thickness measurement should be performed and interpreted according to applicable procedures, engineering requirements and standards.

 

A UT345A thickness reading by itself does not establish whether equipment is safe or unsafe.

 

Where required, involve appropriately qualified inspection or engineering personnel.

 

Need Help with Your UNI-T UT345A?

 

If you are unsure how to measure your particular material, send MTM Precision:

 

Photo of the object

 

Material type

 

Expected thickness

 

Pipe diameter if applicable

 

Surface condition

 

Photo/video of the UT345A reading

 

This is much more useful than simply saying:

 

“The meter cannot measure.”

 

Our team can perform an initial assessment of the application and measurement setup.

 

UNI-T UT345A Ultrasonic Thickness Gauge Malaysia

 

MTM Precision Sdn Bhd supplies UNI-T UT345A Ultrasonic Thickness Gauge for pipe-wall measurement, metal-thickness inspection, corrosion monitoring and factory-maintenance applications in Malaysia.

 

MTM Precision Sdn Bhd

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Tel: 03-8080 7172

WhatsApp: +6016-660 7346

Email: mtmpre@yahoo.com

Website: www.mtmpre.com.my

 

For UNI-T UT345A Malaysia, product enquiries, application checking and ultrasonic thickness measurement support, contact MTM Precision.


 

22 Sep 2026