Suspended Solids Meter Reading Unstable? Causes and Solutions Malaysia
Suspended Solids Meter Reading Unstable? Causes and Solutions Malaysia
An unstable suspended-solids reading does not always mean that the meter is faulty.
Wastewater particles may settle, float or move continuously inside the sample. Air bubbles, dirty measuring bottles and inconsistent sample preparation can also cause readings to change.
Before sending the instrument for inspection, check the sample and measuring procedure systematically.
Why Do Suspended-Solids Readings Change?
A suspended-solids meter evaluates particles present in a water sample. Unlike a completely dissolved solution, suspended particles may not remain evenly distributed.
The reading may change when:
-
Heavy particles settle
-
Light material floats
-
Particles gather together
-
Air bubbles move through the sample
-
The sample bottle is rotated
-
The sample changes with time
Correct sample handling is therefore essential for repeatable results.
Cause 1: Particles Are Settling
Particle settling is one of the most common reasons for falling or inconsistent readings.
Large and heavy particles may begin settling immediately after the sample is collected. If the first test is performed quickly but the second test is delayed, the two results may differ.
Recommended Action
-
Mix the sample gently before measurement.
-
Avoid vigorous shaking.
-
Transfer the sample into the measuring bottle promptly.
-
Measure within a consistent time.
-
Repeat the same procedure for every sample.
If particles settle very quickly, record the time between mixing and measurement.
Cause 2: The Sample Contains Air Bubbles
Air bubbles may scatter light and affect optical measurements.
Bubbles are commonly introduced when:
-
The sample is shaken strongly
-
Water is poured too quickly
-
The bottle is filled from a height
-
Wastewater contains foam
-
A pump introduces air into the sample
Recommended Action
Mix the sample gently and allow large bubbles to disappear without waiting long enough for suspended particles to settle.
Do not tap or shake the measuring bottle aggressively immediately before testing.
Cause 3: The Measuring Bottle Is Dirty
Fingerprints, oil, stains, chemical residue and water droplets on the measuring bottle may interfere with the reading.
Recommended Action
-
Clean the bottle after every test.
-
Rinse it using an appropriate procedure for the sample.
-
Wipe the outside with a clean, soft material.
-
Make sure the outside is dry.
-
Avoid touching the optical measurement area.
-
Replace permanently stained or scratched bottles.
The Qiwei SS-500B is supplied with two measuring bottles, allowing users to compare the condition of the bottles when troubleshooting.
Cause 4: The Bottle Is Scratched
Repeated use, unsuitable cleaning tools or rough handling may scratch the bottle.
Even small scratches can affect the way light passes through the sample.
Recommended Action
Inspect the bottle under good lighting. If one bottle produces consistently different readings, replace it and compare the result using a clean bottle.
Do not use abrasive cleaning materials.
Cause 5: The Sample Is Not Representative
Wastewater conditions may vary significantly within the same tank or drain.
A sample collected from the surface may contain fewer heavy particles, while a sample collected near the bottom may contain more sediment.
Recommended Action
Use an established sampling point and sampling depth.
When comparing results over time, keep the following consistent:
-
Sampling location
-
Sampling depth
-
Sampling time
-
Container type
-
Mixing procedure
-
Delay before measurement
Without consistent sampling, the meter may correctly display differences between samples that appear visually similar.
Cause 6: The Water Has Strong Colour
Strongly coloured industrial water may influence some optical measurement methods.
Potential examples include wastewater from:
-
Dyeing and textile processes
-
Food and beverage production
-
Chemical manufacturing
-
Metal treatment
-
Printing operations
Recommended Action
Compare the portable reading with a reference sample or laboratory result. Provide sample information to the supplier when selecting the instrument.
Do not assume that every coloured sample will behave like clear wastewater.
Cause 7: The Reading Is Outside the Instrument Range
The Qiwei SS-500B measures suspended solids from 0 to 500 mg/L.
A highly concentrated sample may exceed the instrument’s range and produce an over-range, maximum or unreliable reading.
Recommended Action
Confirm the expected concentration using historical laboratory results. If the sample commonly exceeds 500 mg/L, consult the supplier regarding a more suitable measuring range or an approved sample-preparation procedure.
Do not dilute compliance samples without following the required test method.
Cause 8: Calibration or Verification Is Required
If clean reference samples also produce unexpected results, the instrument may require calibration, verification or inspection.
Possible signs include:
-
Zero sample does not return the expected reading
-
Results remain unstable with a stable reference sample
-
Two clean bottles give significantly different results
-
Readings change after impact or long storage
-
Results differ substantially from laboratory findings
Follow the manufacturer’s calibration instructions and use the correct standards.
Quick Troubleshooting Checklist
Before requesting service, check the following:
-
Is the sample mixed evenly?
-
Are particles settling quickly?
-
Are there air bubbles?
-
Is the bottle clean and dry?
-
Is the bottle scratched?
-
Was the same sampling point used?
-
Is the sample strongly coloured?
-
Is the concentration within 0–500 mg/L?
-
Has calibration or verification been performed?
-
Can the result be compared with a known sample?
How to Check Whether the Problem Is the Sample or Meter
Use a simple comparison procedure:
-
Prepare one well-mixed sample.
-
Fill both clean measuring bottles from the same sample.
-
Remove bubbles and clean the bottle surfaces.
-
Measure both bottles using the same timing.
-
Repeat the measurement.
-
Compare the results.
If both bottles produce similar results but field samples vary, the variation may be caused by the sample.
If a stable reference sample repeatedly produces inconsistent results, the meter, bottle or calibration procedure should be inspected.
About the Qiwei SS-500B
The Qiwei SS-500B Portable Suspended Solids Meter provides:
-
Measurement range: 0–500 mg/L
-
Displayed resolution: 1 mg/L
-
Battery or adaptor operation
-
Portable field and indoor use
-
Two measuring bottles
It is suitable for routine suspended-solids screening, wastewater-process comparison and industrial water monitoring.
Contact MTM Precision
For the Qiwei SS-500B Portable Suspended Solids Meter, troubleshooting assistance, replacement measuring bottles, water-quality instruments or quotation, contact:
MTM Precision Sdn. Bhd.
Professional Test & Measurement Solutions Partner in Malaysia
Website: www.mtmpre.com.my
Email: mtmpre@yahoo.com
WhatsApp: +6016-660 7346
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.
Malaysia Coverage:
Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kedah, Perlis, Terengganu, Sabah and Sarawak.
15 Sep 2026