How to Clean Suspended Solids Meter Sample Bottles Malaysia
How to Clean Suspended Solids Meter Sample Bottles Malaysia
A clean sample bottle is essential when measuring suspended solids with a portable meter.
Fingerprints, stains, water droplets, scratches and residue from previous samples may affect the instrument’s optical path and cause inconsistent readings.
The Qiwei SS-500B Portable Suspended Solids Meter includes two measuring bottles. Correct cleaning and handling can improve repeatability and extend their usable life.
Why Does Bottle Condition Matter?
A suspended-solids meter evaluates particles inside a water sample using an optical measurement system.
Anything that changes the way light passes through the bottle may affect the result, including:
Fingerprints
Oil
Dust
Water droplets
Chemical stains
Dried wastewater residue
Scratches
Air bubbles
Cloudy glass or plastic surfaces
A bottle may appear clean from a distance but still have a thin film that influences the reading.
Clean the Bottle Immediately After Testing
Wastewater residue becomes more difficult to remove after it dries.
After completing a measurement:
Empty the sample safely.
Rinse the bottle promptly.
Repeat the rinse if particles remain.
Inspect the inside and outside surfaces.
Allow the bottle to dry in a clean location.
Store it where it is protected from dust and scratches.
Follow the facility’s safety procedure when handling contaminated, chemical or biological samples.
Rinse Between Different Samples
Residue from one sample may contaminate the next sample and produce a misleading result.
Before testing a new water source:
Rinse the bottle thoroughly.
Check that no sediment remains at the bottom.
Confirm that no foam or chemical film is present.
Avoid mixing wastewater from different sampling points.
Use separate bottles when cross-contamination is a concern.
For consistent testing, some facilities assign one bottle to a specific sampling location.
Avoid Abrasive Cleaning Materials
Rough brushes, abrasive powder and damaged cloth may scratch the measuring surface.
Scratches can scatter light and may cause a permanent difference between bottles.
Use a soft, clean material for the outside surface. Any internal cleaning procedure should be suitable for both the bottle material and the tested sample.
Do not use an aggressive chemical cleaner unless it is approved by the manufacturer or compatible with the bottle.
Remove Fingerprints Before Measurement
Operators frequently touch the bottle while filling or inserting it into the meter.
Fingerprints contain oil and moisture that may affect the optical measurement area.
Before inserting the bottle:
Hold it near the cap or upper section.
Inspect the outside under good lighting.
Wipe away fingerprints using a clean, lint-free material.
Confirm that the outside is dry.
Avoid touching the cleaned measurement area again.
This simple step can improve measurement consistency.
Check for External Water Droplets
Water droplets may remain after rinsing or filling the bottle.
Even clean water on the outside can change how light enters or leaves the sample.
Always dry the outside before measurement. Pay particular attention to the lower area of the bottle that sits inside the instrument.
Remove Air Bubbles from the Sample
Air bubbles inside the bottle can be mistaken for changes in suspended particles.
Bubbles may appear when:
The sample is shaken strongly
Water is poured quickly
Wastewater contains foam
A pump introduces air
Cleaning chemicals remain in the bottle
Mix samples gently and pour slowly. Allow large bubbles to disappear, but do not wait so long that suspended particles settle.
Inspect for Scratches and Permanent Cloudiness
Over time, repeated use may cause the measuring bottle to become scratched or cloudy.
Compare both bottles supplied with the SS-500B.
If one bottle repeatedly gives a different result using the same sample:
Clean both bottles again.
Fill them from the same well-mixed sample.
Measure them using identical timing.
Repeat the comparison.
Inspect the bottle producing unusual readings.
A damaged bottle should be replaced rather than continually corrected through measurement technique.
Use Consistent Bottle Orientation
Some sample bottles may produce small variations when rotated because of surface condition or manufacturing differences.
If the bottle has an orientation mark, align it consistently when inserting it into the instrument.
If no mark is provided, users may establish a consistent handling procedure without permanently marking the optical measurement area.
Avoid Mixing Bottle Sets
When several suspended-solids meters are used at one facility, sample bottles may be exchanged accidentally.
To reduce confusion:
Label the storage case
Keep each bottle with its assigned instrument
Record replacement dates
Separate damaged bottles immediately
Avoid placing labels over the measurement area
This is especially useful for wastewater contractors operating at multiple customer sites.
How to Store Sample Bottles
Store bottles:
Clean and dry
Inside a protective case
Away from dust
Away from direct sunlight
Away from abrasive tools
Away from chemical vapour
Where they cannot roll or collide
Do not store wastewater inside the bottle for extended periods.
Common Cleaning Mistakes
Leaving Samples Overnight
Dried residue may create stains or films that are difficult to remove.
Using Tissue That Leaves Fibres
Loose fibres may remain on the bottle surface and affect measurement.
Touching the Bottle After Cleaning
Handling the optical area reintroduces fingerprints.
Using One Quick Rinse
Sticky, oily or high-solids wastewater may require more thorough cleaning.
Continuing to Use Scratched Bottles
Cleaning cannot repair scratches or permanent surface damage.
Sample-Bottle Troubleshooting Test
If the SS-500B readings appear inconsistent:
Prepare one well-mixed sample.
Clean and dry both bottles.
Fill both bottles from the same sample.
Remove bubbles.
Measure both using the same timing.
Repeat each reading.
Compare the results.
If one bottle consistently produces different readings, inspect or replace that bottle.
If both bottles produce unstable results with a stable sample, review calibration, instrument condition and operating procedure.
About the Qiwei SS-500B
The Qiwei SS-500B Portable Suspended Solids Meter offers:
Measurement range: 0–500 mg/L
Displayed resolution: 1 mg/L
Battery or adaptor operation
Two measuring bottles
Portable field and indoor testing
It is suitable for routine wastewater screening, treatment-process comparison and industrial water monitoring.
Contact MTM Precision
For the Qiwei SS-500B Portable Suspended Solids Meter, replacement sample bottles, water-quality instruments, technical consultation or quotation, contact:
MTM Precision Sdn. Bhd.
Professional Test & Measurement Solutions Partner in Malaysia
Website: www.mtmpre.com.my
Email: mtmpre@yahoo.com
WhatsApp: +6016-660 7346
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.
Malaysia Coverage:
Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kedah, Perlis, Terengganu, Sabah and Sarawak.
15 Sep 2026