How to Calibrate QiWei WGZ-1B Turbidity Meter and Use Turbidity Standards Malaysia

How to Calibrate QiWei WGZ-1B Turbidity Meter Malaysia

 

Why does a turbidity meter need calibration, and which turbidity standard should be used?

 

The QiWei WGZ-1B Portable Turbidity Meter measures turbidity from 0 to 200 NTU using the 90° scattered-light principle.

 

Correct calibration and verification help users identify instrument drift, contaminated sample vials, deteriorated standards and measurement errors before testing actual water samples.

 

โญ QiWei WGZ-1B Main Specifications

 

Measurement parameter: Turbidity

 

Measuring unit: NTU

 

Measuring range: 0–200 NTU

 

Minimum display: 0.1 NTU

 

Measurement principle: 90° scattered light

 

Repeatability: ≤1%

 

Zero drift: ±1%

 

Power supply: 5 × AA batteries or AC adapter

 

Portable design

 

LCD display

 

Automatic power-off

 

Sample vials and Formazine turbidity standard supplied

 

๐Ÿงช What Is a Turbidity Standard?

 

A turbidity standard is a liquid with a known reference value used to check or adjust the response of a turbidity meter.

 

Common purposes include:

 

Initial instrument setup

 

Routine performance checks

 

Calibration

 

Verification after transport

 

Troubleshooting unusual readings

 

Checking performance after long-term storage

 

Quality-control documentation

 

A standard is not an ordinary water sample. It must be handled carefully to protect its assigned value.

 

๐Ÿ” Calibration vs Verification

 

These two terms are often confused.

 

Verification

 

Verification checks whether the instrument reads an appropriate reference standard within an acceptable tolerance.

 

It answers:

 

Is the instrument still producing an acceptable result?

 

Calibration

 

Calibration compares the instrument’s response with one or more known reference values and may involve adjusting the instrument.

 

It answers:

 

How does the instrument respond to known turbidity values, and does it require adjustment?

 

Accredited Calibration

 

When a customer requires a traceable or accredited certificate, this is different from the user’s routine adjustment procedure.

 

The customer should confirm:

 

Required standard

 

Required calibration points

 

Acceptance tolerance

 

Certificate format

 

Traceability requirement

 

Accreditation requirement

 

Calibration interval

 

โš ๏ธ Important Note About the Included Standard

 

The WGZ-1B product information states that a 400 NTU Formazine turbidity standard solution is included.

 

However, the stated WGZ-1B measuring range is 0–200 NTU.

 

Users should not assume that the 400 NTU solution can be inserted and measured directly. It may be intended as a stock solution for preparing lower calibration values.

 

Before preparing or diluting any standard:

 

Check the QiWei operating manual

 

Confirm the required calibration points

 

Confirm whether the standard is supplied ready to use

 

Follow the manufacturer’s preparation instructions

 

Use suitable volumetric equipment

 

Do not estimate dilution volumes

 

Do not exceed the instrument’s specified measuring range

 

Incorrectly prepared standards will result in incorrect calibration.

 

๐Ÿงฐ Items Required Before Calibration

 

Depending on the instrument procedure, prepare:

 

QiWei WGZ-1B turbidity meter

 

Correct turbidity standards

 

Clean, matching sample vials

 

Lint-free cleaning cloth

 

Suitable purified or turbidity-free water, if specified

 

Clean transfer equipment

 

Gloves

 

Calibration record sheet

 

QiWei operating manual

 

Do not begin calibration until the required standard concentrations and preparation method are confirmed.

 

โœ… Pre-Calibration Inspection

 

Before using the calibration function:

 

Check that the instrument is clean and undamaged.

 

Inspect the measuring chamber.

 

Check the sample vials for scratches or stains.

 

Confirm the standard concentration.

 

Check the standard’s expiry date.

 

Review its storage condition.

 

Check for contamination or leakage.

 

Allow the instrument and standard to reach suitable measurement conditions.

 

Install fresh batteries or use the correct AC adapter.

 

Read the manufacturer’s calibration instructions.

 

๐Ÿ“ General Calibration Workflow

 

The exact button sequence and calibration points must follow the QiWei manual. A general workflow is:

 

Step 1: Power On the Instrument

 

Place the meter on a stable surface and allow it to initialise.

 

Step 2: Prepare the Required Standard

 

Use the specified standard concentration. If preparation is necessary, follow the manufacturer’s documented procedure.

 

Step 3: Prepare the Vial

 

Fill a clean vial with the standard. Avoid contaminating the inside of the vial or cap.

 

Step 4: Remove Air Bubbles

 

Allow bubbles to disappear. Do not shake the vial aggressively.

 

Step 5: Clean the Vial Exterior

 

Use a clean, lint-free cloth to remove fingerprints, moisture and residue.

 

Step 6: Insert the Vial Consistently

 

Align the vial according to the instrument or vial marking.

 

Step 7: Start the Calibration Procedure

 

Follow the correct WGZ-1B button sequence and calibration-point order.

 

Step 8: Confirm the Calibration

 

Wait for the instrument to complete the procedure before removing the vial.

 

Step 9: Verify the Result

 

Remeasure an appropriate standard to confirm that the instrument response is acceptable.

 

Step 10: Record the Calibration

 

Document the date, standard value, batch number, expiry date, reading, result and operator.

 

๐Ÿซง Why Air Bubbles Affect Calibration

 

The WGZ-1B detects light scattered by material in the vial.

 

Air bubbles can also scatter light and may cause:

 

An incorrectly high reading

 

Poor repeatability

 

Calibration failure

 

Different results after reinserting the same vial

 

Pour the standard carefully and allow bubbles to disappear before measurement.

 

๐Ÿงด How to Handle Turbidity Standards

 

To protect the standard:

 

Check its instructions before use

 

Store it under the specified conditions

 

Keep the container tightly closed

 

Avoid touching the inside of the cap

 

Use clean transfer equipment

 

Do not return used liquid to the original bottle

 

Avoid cross-contamination between concentrations

 

Record the opening date

 

Check the expiry date

 

Dispose of expired material appropriately

 

๐Ÿงผ Why Sample-Vial Condition Matters

 

The sample vial forms part of the instrument’s optical path.

 

Calibration can be affected by:

 

Fingerprints

 

Dust

 

Scratches

 

Water droplets

 

Stains

 

Condensation

 

Mismatched vials

 

Inconsistent vial orientation

 

Use clean, undamaged vials and handle them near the cap or upper section.

 

โŒ Common Calibration Mistakes

 

Avoid the following:

 

Using an expired standard

 

Using an unknown standard concentration

 

Measuring a stock solution outside the instrument range

 

Making an estimated dilution

 

Using dirty or scratched vials

 

Leaving fingerprints on the vial

 

Shaking the standard and creating bubbles

 

Changing the vial orientation between measurements

 

Skipping verification after adjustment

 

Using ordinary tap water as a zero standard without confirmation

 

Returning used standard to the original bottle

 

Failing to record the calibration result

 

๐Ÿ”ง Why Calibration May Fail

 

Calibration failure may be caused by:

 

Incorrect standard concentration

 

Improperly prepared dilution

 

Contaminated standard

 

Expired standard

 

Dirty sample vial

 

Scratched vial

 

Air bubbles

 

Incorrect vial orientation

 

Contaminated measuring chamber

 

Low battery

 

Wrong calibration sequence

 

Instrument drift or optical fault

 

Check the simple causes before assuming that the instrument requires repair.

 

๐Ÿ“… How Often Should WGZ-1B Be Checked?

 

The appropriate interval depends on:

 

Frequency of use

 

Importance of the measurements

 

Site quality procedure

 

Required test method

 

Environmental conditions

 

Instrument transport

 

Previous verification history

 

Customer or consultant requirements

 

Additional verification should be considered:

 

Before important measurements

 

After transporting the instrument

 

After long-term storage

 

After replacing or changing standards

 

When results appear unusual

 

After the instrument is dropped

 

After exposure to moisture or chemicals

 

Follow the organisation’s quality procedure and applicable measurement requirements.

 

๐Ÿ“‹ Suggested Calibration Record

 

Record:

 

Instrument model

 

Serial number

 

Calibration date

 

Calibration-standard value

 

Standard manufacturer

 

Batch or lot number

 

Expiry date

 

Initial reading

 

Final reading

 

Acceptance limit

 

Pass or fail result

 

Operator

 

Corrective action

 

Next review date

 

Good documentation makes it easier to investigate abnormal readings later.

 

โš ๏ธ When Should You Stop Using the Meter?

 

Stop relying on the instrument for critical measurements when:

 

A valid standard repeatedly produces unacceptable results

 

Calibration cannot be completed

 

The measuring chamber is contaminated or damaged

 

Multiple undamaged vials produce inconsistent readings

 

The instrument was dropped or flooded

 

The display or controls behave abnormally

 

The required test method is outside the WGZ-1B specification

 

Arrange inspection or comparison with a suitable reference instrument.

 

๐Ÿ“ QiWei WGZ-1B and Turbidity Standard Supplier Malaysia

 

MTM Precision supplies the QiWei WGZ-1B and water-quality testing instruments for customers in Selangor, Kuala Lumpur, Johor, Penang, Perak, Melaka, Negeri Sembilan, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sabah and Sarawak.

 

When requesting a replacement turbidity standard or technical assistance, provide:

 

Instrument model

 

Required standard concentration

 

Existing standard details

 

Expected calibration points

 

Certificate requirement

 

Application

 

Delivery location

 

๐Ÿ“ž Contact MTM Precision

 

MTM Precision Sdn Bhd

 

Tel / WhatsApp: +6016-660 7346

Office: +603-8080 7172

Email: mtmpre@yahoo.com

Website: www.mtmpre.com.my

 

Showroom & Service Centre:

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Contact MTM Precision for the QiWei WGZ-1B price, availability, turbidity-standard enquiries, sample-vial enquiries and technical support in Malaysia.


 

15 Sep 2026