How to Calibrate QiWei WGZ-1B Turbidity Meter and Use Turbidity Standards Malaysia

How to Calibrate QiWei WGZ-1B Turbidity Meter Malaysia


Why does a turbidity meter need calibration, and which turbidity standard should be used?


The QiWei WGZ-1B Portable Turbidity Meter measures turbidity from 0 to 200 NTU using the 90° scattered-light principle.


Correct calibration and verification help users identify instrument drift, contaminated sample vials, deteriorated standards and measurement errors before testing actual water samples.


⭐ QiWei WGZ-1B Main Specifications


Measurement parameter: Turbidity


Measuring unit: NTU


Measuring range: 0–200 NTU


Minimum display: 0.1 NTU


Measurement principle: 90° scattered light


Repeatability: ≤1%


Zero drift: ±1%


Power supply: 5 × AA batteries or AC adapter


Portable design


LCD display


Automatic power-off


Sample vials and Formazine turbidity standard supplied


🧪 What Is a Turbidity Standard?


A turbidity standard is a liquid with a known reference value used to check or adjust the response of a turbidity meter.


Common purposes include:


Initial instrument setup


Routine performance checks


Calibration


Verification after transport


Troubleshooting unusual readings


Checking performance after long-term storage


Quality-control documentation


A standard is not an ordinary water sample. It must be handled carefully to protect its assigned value.


🔍 Calibration vs Verification


These two terms are often confused.


Verification


Verification checks whether the instrument reads an appropriate reference standard within an acceptable tolerance.


It answers:


Is the instrument still producing an acceptable result?


Calibration


Calibration compares the instrument’s response with one or more known reference values and may involve adjusting the instrument.


It answers:


How does the instrument respond to known turbidity values, and does it require adjustment?


Accredited Calibration


When a customer requires a traceable or accredited certificate, this is different from the user’s routine adjustment procedure.


The customer should confirm:


Required standard


Required calibration points


Acceptance tolerance


Certificate format


Traceability requirement


Accreditation requirement


Calibration interval


⚠️ Important Note About the Included Standard


The WGZ-1B product information states that a 400 NTU Formazine turbidity standard solution is included.


However, the stated WGZ-1B measuring range is 0–200 NTU.


Users should not assume that the 400 NTU solution can be inserted and measured directly. It may be intended as a stock solution for preparing lower calibration values.


Before preparing or diluting any standard:


Check the QiWei operating manual


Confirm the required calibration points


Confirm whether the standard is supplied ready to use


Follow the manufacturer’s preparation instructions


Use suitable volumetric equipment


Do not estimate dilution volumes


Do not exceed the instrument’s specified measuring range


Incorrectly prepared standards will result in incorrect calibration.


🧰 Items Required Before Calibration


Depending on the instrument procedure, prepare:


QiWei WGZ-1B turbidity meter


Correct turbidity standards


Clean, matching sample vials


Lint-free cleaning cloth


Suitable purified or turbidity-free water, if specified


Clean transfer equipment


Gloves


Calibration record sheet


QiWei operating manual


Do not begin calibration until the required standard concentrations and preparation method are confirmed.


✅ Pre-Calibration Inspection


Before using the calibration function:


Check that the instrument is clean and undamaged.


Inspect the measuring chamber.


Check the sample vials for scratches or stains.


Confirm the standard concentration.


Check the standard’s expiry date.


Review its storage condition.


Check for contamination or leakage.


Allow the instrument and standard to reach suitable measurement conditions.


Install fresh batteries or use the correct AC adapter.


Read the manufacturer’s calibration instructions.


📝 General Calibration Workflow


The exact button sequence and calibration points must follow the QiWei manual. A general workflow is:


Step 1: Power On the Instrument


Place the meter on a stable surface and allow it to initialise.


Step 2: Prepare the Required Standard


Use the specified standard concentration. If preparation is necessary, follow the manufacturer’s documented procedure.


Step 3: Prepare the Vial


Fill a clean vial with the standard. Avoid contaminating the inside of the vial or cap.


Step 4: Remove Air Bubbles


Allow bubbles to disappear. Do not shake the vial aggressively.


Step 5: Clean the Vial Exterior


Use a clean, lint-free cloth to remove fingerprints, moisture and residue.


Step 6: Insert the Vial Consistently


Align the vial according to the instrument or vial marking.


Step 7: Start the Calibration Procedure


Follow the correct WGZ-1B button sequence and calibration-point order.


Step 8: Confirm the Calibration


Wait for the instrument to complete the procedure before removing the vial.


Step 9: Verify the Result


Remeasure an appropriate standard to confirm that the instrument response is acceptable.


Step 10: Record the Calibration


Document the date, standard value, batch number, expiry date, reading, result and operator.


🫧 Why Air Bubbles Affect Calibration


The WGZ-1B detects light scattered by material in the vial.


Air bubbles can also scatter light and may cause:


An incorrectly high reading


Poor repeatability


Calibration failure


Different results after reinserting the same vial


Pour the standard carefully and allow bubbles to disappear before measurement.


🧴 How to Handle Turbidity Standards


To protect the standard:


Check its instructions before use


Store it under the specified conditions


Keep the container tightly closed


Avoid touching the inside of the cap


Use clean transfer equipment


Do not return used liquid to the original bottle


Avoid cross-contamination between concentrations


Record the opening date


Check the expiry date


Dispose of expired material appropriately


🧼 Why Sample-Vial Condition Matters


The sample vial forms part of the instrument’s optical path.


Calibration can be affected by:


Fingerprints


Dust


Scratches


Water droplets


Stains


Condensation


Mismatched vials


Inconsistent vial orientation


Use clean, undamaged vials and handle them near the cap or upper section.


❌ Common Calibration Mistakes


Avoid the following:


Using an expired standard


Using an unknown standard concentration


Measuring a stock solution outside the instrument range


Making an estimated dilution


Using dirty or scratched vials


Leaving fingerprints on the vial


Shaking the standard and creating bubbles


Changing the vial orientation between measurements


Skipping verification after adjustment


Using ordinary tap water as a zero standard without confirmation


Returning used standard to the original bottle


Failing to record the calibration result


🔧 Why Calibration May Fail


Calibration failure may be caused by:


Incorrect standard concentration


Improperly prepared dilution


Contaminated standard


Expired standard


Dirty sample vial


Scratched vial


Air bubbles


Incorrect vial orientation


Contaminated measuring chamber


Low battery


Wrong calibration sequence


Instrument drift or optical fault


Check the simple causes before assuming that the instrument requires repair.


📅 How Often Should WGZ-1B Be Checked?


The appropriate interval depends on:


Frequency of use


Importance of the measurements


Site quality procedure


Required test method


Environmental conditions


Instrument transport


Previous verification history


Customer or consultant requirements


Additional verification should be considered:


Before important measurements


After transporting the instrument


After long-term storage


After replacing or changing standards


When results appear unusual


After the instrument is dropped


After exposure to moisture or chemicals


Follow the organisation’s quality procedure and applicable measurement requirements.


📋 Suggested Calibration Record


Record:


Instrument model


Serial number


Calibration date


Calibration-standard value


Standard manufacturer


Batch or lot number


Expiry date


Initial reading


Final reading


Acceptance limit


Pass or fail result


Operator


Corrective action


Next review date


Good documentation makes it easier to investigate abnormal readings later.


⚠️ When Should You Stop Using the Meter?


Stop relying on the instrument for critical measurements when:


A valid standard repeatedly produces unacceptable results


Calibration cannot be completed


The measuring chamber is contaminated or damaged


Multiple undamaged vials produce inconsistent readings


The instrument was dropped or flooded


The display or controls behave abnormally


The required test method is outside the WGZ-1B specification


Arrange inspection or comparison with a suitable reference instrument.


📍 QiWei WGZ-1B and Turbidity Standard Supplier Malaysia


MTM Precision supplies the QiWei WGZ-1B and water-quality testing instruments for customers in Selangor, Kuala Lumpur, Johor, Penang, Perak, Melaka, Negeri Sembilan, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sabah and Sarawak.


When requesting a replacement turbidity standard or technical assistance, provide:


Instrument model


Required standard concentration


Existing standard details


Expected calibration points


Certificate requirement


Application


Delivery location


📞 Contact MTM Precision


MTM Precision Sdn Bhd


Tel / WhatsApp: +6016-660 7346

Office: +603-8080 7172

Email: mtmpre@yahoo.com

Website: www.mtmpre.com.my


Showroom & Service Centre:

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.


Contact MTM Precision for the QiWei WGZ-1B price, availability, turbidity-standard enquiries, sample-vial enquiries and technical support in Malaysia.



15 Sep 2026