How to Perform a Breaking Force Test with the Landtek FM-207-500K Malaysia

How to Perform a Breaking Force Test with the Landtek FM-207-500K Malaysia

 

A breaking-force test determines the maximum force a product, material or connection can withstand before it breaks, separates, slips or permanently fails.

 

The Landtek FM-207-500K Digital Force Gauge measures tension and compression force up to:

 

±500 kgf

 

±4900 N

 

±1100 lbf

 

Its peak-value function captures the highest force reached during the test, while its external sensor can be installed on a suitable high-capacity test stand or customised fixture.

 

What Is Breaking Force?

 

Breaking force is the maximum measured force reached before a sample fails.

 

Failure does not always mean that the sample breaks into two pieces. Depending on the product, failure may occur when:

 

A cable snaps

 

A wire pulls out of a terminal

 

A bonded joint separates

 

A hook opens or deforms

 

A fastener pulls out

 

A clip releases

 

A rope begins to slip

 

A plastic component cracks

 

An assembly separates

 

A compression sample collapses

 

The maximum force and the observed failure mode should both be recorded.

 

Why Perform a Breaking-Force Test?

 

Manufacturers and quality-control teams use breaking-force tests to:

 

Compare different materials

 

Evaluate product designs

 

Check production consistency

 

Compare alternative suppliers

 

Investigate customer complaints

 

Determine the weakest component

 

Verify internal quality requirements

 

Monitor changes after a process adjustment

 

Support research and development

 

The test can reveal whether a product fails at the intended point or at an unexpected connection.

 

Products That May Be Tested

 

Depending on their size and expected breaking force, possible samples include:

 

Cables and wires

 

Rope and straps

 

Hooks and clips

 

Fasteners

 

Terminals and connectors

 

Bonded components

 

Plastic parts

 

Mechanical assemblies

 

Handles and latches

 

Springs

 

Packaging components

 

Construction hardware

 

The expected breaking force must remain within the FM-207-500K measuring range.

 

Breaking Force vs Proof Load

 

These terms describe different tests.

 

Breaking-Force Test

 

The load is increased until the sample fails. The sample is usually damaged and cannot be reused.

 

The objective is to measure the maximum force before failure.

 

Proof-Load Test

 

A defined load is applied without intentionally breaking the sample.

 

The objective is to confirm that the product can withstand the required force without unacceptable damage or permanent deformation.

 

A proof-load value should not be treated as a breaking-force result, and a breaking-force result should not automatically be used as a safe working load.

 

Equipment Required

 

A typical breaking-force test requires:

 

Landtek FM-207-500K display unit

 

External force sensor

 

Suitable high-capacity test stand

 

Pulling or compression mechanism

 

Sample-specific grips

 

Load-rated connectors

 

Secure mounting structure

 

Protective guard

 

Test-recording method

 

The force gauge measures force but does not generate the load. A separate manual, motorised, hydraulic or customised loading mechanism is required.

 

Step 1: Define the Test Method

 

Before testing, determine:

 

Product being tested

 

Expected breaking-force range

 

Pull or compression direction

 

Required measurement unit

 

Sample preparation method

 

Grip or fixture type

 

Loading speed

 

Number of samples

 

Applicable internal or external standard

 

Information to be recorded

 

Tests should use the same procedure if the results will be compared.

 

Changing the fixture, loading speed or sample position may change the measured breaking force.

 

Step 2: Confirm the Correct Gauge Range

 

The FM-207-500K measures up to 500 kgf.

 

It may be suitable when the expected breaking force is within this range and not too small for useful measurement.

 

Choose a lower-capacity force gauge if the expected breaking force is very low. Select a higher-capacity system if the sample may exceed 500 kgf.

 

Avoid selecting a gauge only because its range is larger. The expected test force should match the useful range of the instrument.

 

Step 3: Select the Correct Grip or Fixture

 

The grip must hold the sample securely without creating an unintended weak point.

 

Examples include:

 

Wedge grips

 

Cable clamps

 

Capstan grips

 

Hooks

 

Clevis connections

 

Threaded adapters

 

Compression plates

 

Vise-type grips

 

Custom-machined fixtures

 

If the sample slips from the grip before breaking, the recorded value may represent grip performance rather than the sample’s actual breaking strength.

 

If the grip damages the sample prematurely, the result may also be misleading.

 

Step 4: Inspect the Complete Test System

 

Check the condition and capacity of:

 

Test frame

 

Sensor

 

Hooks

 

Bolts

 

Shackles

 

Clamps

 

Mounting plates

 

Grip surfaces

 

Protective guard

 

Every load-bearing part should safely withstand the expected force.

 

The test system’s overall capacity is limited by its weakest component.

 

Step 5: Align the Sensor and Sample

 

The applied force should pass through the sensor in a straight line.

 

Avoid:

 

Side loading

 

Bending

 

Twisting

 

Off-centre pulling

 

Uneven compression

 

Sudden impact

 

Poor alignment can cause unstable readings, incorrect failure behaviour or sensor damage.

 

Step 6: Zero the Force Gauge

 

Install the sensor, fixture and sample before zeroing the gauge.

 

Remove any unintended load and confirm that the display remains stable at zero.

 

If the reading drifts, inspect the system for:

 

Residual tension

 

Binding fixtures

 

Unstable mounting

 

Environmental vibration

 

Loose sensor connection

 

Do not start the breaking test until the zero reading is stable.

 

Step 7: Select Peak-Value Mode

 

Peak-value mode is normally used for a breaking-force test.

 

As the load increases, the instrument records the highest force reached. When the sample fails and the applied force drops, the maximum value remains available on the display.

 

Real-time mode may also be used while preparing or checking the setup, but peak mode is the key function for capturing the failure force.

 

Step 8: Apply the Force Gradually

 

Increase the load at a controlled rate.

 

Avoid sudden jerking or impact because this may:

 

Produce an artificially high peak

 

Damage the sensor

 

Cause unpredictable sample failure

 

Reduce test repeatability

 

Create a safety hazard

 

Observe the sample and fixture throughout the test, but remain outside the direct failure path.

 

Stop immediately if:

 

The overload warning activates

 

The fixture slips

 

The test frame bends

 

The sensor becomes misaligned

 

A connection point becomes unstable

 

The expected force exceeds the selected range

 

Step 9: Record the Peak Force

 

After the sample fails, record the displayed peak force.

 

A proper test record should include:

 

Sample identification

 

Product description

 

Material

 

Sample dimensions

 

Batch or supplier

 

Measurement unit

 

Peak breaking force

 

Loading method

 

Loading speed

 

Fixture used

 

Failure location

 

Failure description

 

Test date and operator

 

The FM-207-500K can store up to 99 measurement records, but product and failure information should also be documented separately.

 

Step 10: Examine the Failure Mode

 

The force value alone does not explain why the product failed.

 

For example:

 

A cable breaking in its centre may indicate material failure

 

A cable pulling out of a terminal may indicate a crimping problem

 

A hook bending may indicate inadequate stiffness

 

A bolt pulling out with surrounding material may indicate base-material failure

 

A sample slipping from the grip may indicate an invalid test

 

A bonded joint separating cleanly may indicate poor adhesion

 

Photographing the sample after the test can make the report more useful.

 

How Many Samples Should Be Tested?

 

Testing only one sample provides limited information.

 

Where practical, test multiple samples prepared under the same conditions. This helps identify:

 

Average breaking force

 

Lowest result

 

Highest result

 

Variation between samples

 

Unusual failure modes

 

Production inconsistency

 

The number of required samples should follow the applicable standard, customer requirement or internal quality procedure.

 

Common Breaking-Test Mistakes

 

Using the Wrong Measurement Range

 

A gauge with insufficient capacity may be overloaded, while a gauge with excessive capacity may provide less useful resolution for small forces.

 

Allowing the Sample to Slip

 

A slipping sample may not produce its actual breaking-force result.

 

Changing the Loading Speed

 

Different loading speeds may produce different results, especially for plastic, rubber, adhesives and other time-dependent materials.

 

Applying Force at an Angle

 

Misalignment can introduce bending and cause premature failure.

 

Recording Only the Maximum Number

 

The failure mode, sample details and test conditions are also essential.

 

Treating Breaking Force as Safe Working Load

 

A measured breaking force is not automatically the safe load of the product. Appropriate safety factors and applicable standards must be considered.

 

Safety During Destructive Testing

 

When a sample breaks, stored energy may be released suddenly.

 

Recommended precautions include:

 

Install a suitable protective guard

 

Keep personnel away from the loading line

 

Inspect fixtures before each test

 

Use load-rated connections

 

Wear appropriate personal protective equipment

 

Do not touch the sample while it is under load

 

Apply force gradually

 

Stop if any part of the setup becomes unstable

 

Follow the relevant workplace safety procedure

 

High-force destructive testing should only be conducted by trained personnel.

 

Frequently Asked Questions

 

Can the FM-207-500K automatically break the sample?

 

No. It measures force but does not generate it. A test stand or loading mechanism is required.

 

Can it record the highest force before failure?

 

Yes. Peak-value mode captures and retains the maximum measured force.

 

Can it test in both pulling and compression directions?

 

Yes. The external sensor can measure tension and compression when installed correctly.

 

What happens if the sample requires more than 500 kgf to break?

 

The FM-207-500K is not suitable for a test expected to exceed its range. A higher-capacity system is required.

 

Can the test result be used for product certification?

 

That depends on the applicable standard, calibration requirements, equipment specification and accreditation scope. The force gauge alone does not automatically provide certified product approval.

 

Does the test stand come with the gauge?

 

A suitable stand and fixture are normally selected according to the required force and sample design. Confirm the supplied configuration before ordering.

 

Plan Your Breaking-Force Test in Malaysia

 

Before selecting the equipment, send MTM Precision:

 

Photo and dimensions of the sample

 

Expected breaking force

 

Pull or compression test

 

Required test travel

 

Proposed grip or fixture

 

Number of samples

 

Applicable test standard

 

Required loading method

 

Calibration and reporting requirements

 

MTM Precision can help determine whether the Landtek FM-207-500K is suitable for the intended breaking-force application.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd. (744811-A)

 

Telephone: +603-8080 7172

WhatsApp: +6019-665 7346 / +6016-660 7346

Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my

Website: www.mtmpre.com.my

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar Lapan Belas,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Supply and technical support are available throughout Malaysia, including Selangor, Kuala Lumpur, Johor, Penang, Perak, Melaka, Negeri Sembilan, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sarawak and Sabah.


 

16 Sep 2026