Force Gauge for Testing Fasteners, Hooks and Lifting Components Malaysia

Force Gauge for Testing Fasteners, Hooks and Lifting Components Malaysia

 

Fasteners, hooks and load-bearing components must be able to withstand the forces expected during assembly or use. Manufacturers and quality-control teams may therefore need to measure pull-out force, retention force, deformation force or breaking force.

 

The Landtek FM-207-500K Digital Force Gauge measures push and pull force up to:

 

±500 kgf

 

±4900 N

 

±1100 lbf

 

Its external force sensor can be incorporated into a suitable test stand or customised fixture for industrial component testing in Malaysia.

 

What Components Can Be Evaluated?

 

Depending on their size and expected load, the FM-207-500K may be used to evaluate:

 

Hooks

 

Bolts and screws

 

Rivets

 

Clips and retaining parts

 

Anchors

 

Buckles

 

Shackles

 

Brackets

 

Pins

 

Handles

 

Latches

 

Wire loops

 

Small load-bearing assemblies

 

Mechanical connection points

 

The instrument’s suitability depends on the required force, fixture design and testing procedure.

 

What Types of Force Tests Can Be Performed?

 

Pull-Out Force Test

 

A pull-out test measures the force required to remove a fastener, anchor, insert or connected component from its installed position.

 

Examples include:

 

Pulling an insert from a plastic component

 

Testing an anchor installed in a sample material

 

Measuring the extraction force of a pin

 

Checking whether a hook remains attached

 

Testing a press-fit component

 

The result can help identify weak connections, poor installation or inconsistent production.

 

Retention-Force Test

 

A retention test confirms whether a component remains secured under a specified pulling load.

 

Unlike a destructive test, the component may only need to withstand a defined force for a certain duration.

 

Real-time mode allows the operator to monitor the applied force and maintain it near the required value.

 

Deformation Test

 

Some components must not bend, open or permanently deform below a specified load.

 

During this test, the operator gradually applies force while observing:

 

Applied force

 

Visible deformation

 

Permanent shape change

 

Movement at the connection point

 

Failure of the supporting material

 

The force value should be recorded together with the type of deformation observed.

 

Breaking-Force Test

 

A breaking-force test continues until the component, joint or supporting material fails.

 

Peak-value mode records the highest force reached before failure.

 

The test may help compare:

 

Different materials

 

Alternative component designs

 

Suppliers

 

Production batches

 

Manufacturing processes

 

Destructive tests require appropriate guarding because broken parts may release suddenly.

 

Why Use an External Force Sensor?

 

For higher-force testing, a separate sensor provides more installation flexibility than a small handheld meter.

 

The FM-207-500K sensor can be:

 

Mounted to a test frame

 

Connected to a pulling mechanism

 

Installed in a compression fixture

 

Positioned close to the component

 

Integrated into a customised testing setup

 

The display can remain in a more convenient viewing position while the sensor measures force at the test point.

 

The applied force must remain aligned with the sensor. Twisting, bending and side loading should be avoided.

 

Example: Hook Pull Test

 

A hook pull test may be used to measure:

 

The force required to straighten or open the hook

 

The pull-out strength of its mounting point

 

The load at which the hook breaks

 

Whether the hook can withstand a specified proof load

 

A typical setup includes:

 

A rigid test frame

 

A suitable external force sensor

 

A connection between the sensor and hook

 

A secure mounting point for the test sample

 

A controlled pulling mechanism

 

A protective guard

 

The exact procedure depends on the hook design and applicable specification.

 

Example: Fastener Pull-Out Test

 

A fastener pull-out test evaluates the force required to remove a screw, bolt, rivet, insert or anchor from a sample.

 

Before testing, define:

 

Fastener type and dimensions

 

Base material

 

Installation depth

 

Installation method

 

Pulling direction

 

Loading speed

 

Maximum required load

 

Pass or fail criterion

 

All samples should be prepared consistently. Otherwise, variations in installation may affect the result more than the fastener itself.

 

How to Set Up the Test

 

1. Define the Test Objective

 

Decide whether the test is intended to measure:

 

Proof load

 

Pull-out force

 

Retention force

 

Deformation force

 

Breaking force

 

Compression force

 

The objective determines the fixture, measurement mode and stopping point.

 

2. Confirm the Expected Force

 

The expected force must fall within the useful range of the FM-207-500K.

 

For much smaller forces, a lower-capacity gauge may provide better measurement resolution. Tests that may exceed 500 kgf require a higher-capacity system.

 

3. Select the Correct Fixture

 

The fixture must hold the component without introducing unwanted bending or slippage.

 

Possible fixtures include:

 

Clevis connections

 

Pulling hooks

 

Clamps

 

Threaded adapters

 

Mounting plates

 

Compression plates

 

Custom-machined holders

 

All fixture components must have an adequate rated capacity.

 

4. Align the Sensor

 

The sensor, fixture and component should follow the same force axis.

 

Incorrect alignment may cause:

 

Unstable readings

 

Uneven loading

 

Premature component failure

 

Damage to the sensor

 

Poor test repeatability

 

5. Zero the Instrument

 

After installing the sensor and fixture, remove any unintended load and set the display to zero.

 

A stable zero reading should be confirmed before every test.

 

6. Select the Measurement Mode

 

Use real-time mode for:

 

Applying a proof load

 

Holding a component at a specified force

 

Observing gradual deformation

 

Adjusting an assembly

 

Use peak mode for:

 

Pull-out testing

 

Breaking-force testing

 

Maximum retention-force testing

 

Destructive testing

 

7. Apply the Force Gradually

 

Apply force smoothly and avoid shock loading.

 

Stop the test if:

 

The required load is reached

 

The component fails

 

The fixture slips

 

The sensor becomes misaligned

 

The overload warning activates

 

The test frame becomes unstable

 

8. Document the Result

 

A useful test record may include:

 

Component description

 

Material

 

Sample identification

 

Fixture used

 

Test direction

 

Maximum force

 

Measurement unit

 

Loading method

 

Failure mode

 

Pass or fail result

 

Operator and test date

 

The FM-207-500K can store up to 99 measurement records for later review.

 

Proof Load Is Not the Same as Breaking Load

 

A proof-load test checks whether a component can withstand a specified load without unacceptable damage.

 

A breaking-load test determines the force at which it fails.

 

These two results should not be treated as interchangeable.

 

A component with a measured breaking force of 400 kgf should not automatically be assigned a safe working load of 400 kgf. Safety factors, material behaviour, repeated loading, environmental conditions and applicable standards must also be considered.

 

Important Limitation for Lifting Components

 

The FM-207-500K can measure force during product development, internal quality checks and comparative testing.

 

However, testing regulated lifting accessories may require:

 

A higher-capacity calibrated test machine

 

Traceable calibration

 

A controlled loading rate

 

A specified proof load

 

Accredited inspection

 

Compliance with a recognised standard

 

Formal test documentation

 

The use of a force gauge alone does not automatically certify a hook, shackle or lifting accessory as safe for lifting.

 

The required standard and certification scope should be confirmed before conducting the test.

 

Common Testing Errors

 

Fixture Is Weaker Than the Test Sample

 

The fixture may bend or fail before the component reaches its expected load.

 

Force Is Applied at an Angle

 

Angular loading can produce bending and inconsistent results.

 

Different Samples Are Installed Differently

 

Variations in installation depth, torque or alignment can affect pull-out results.

 

Load Is Applied Too Quickly

 

Sudden loading may create an impact force and an artificially high peak.

 

Only the Force Value Is Recorded

 

The failure mode is also important. A fastener pulling out is different from the fastener itself breaking.

 

Maximum Range Is Treated as a Working Load

 

The instrument’s maximum measuring range should not be exceeded or routinely treated as the ideal everyday testing point.

 

Frequently Asked Questions

 

Can the FM-207-500K test hooks?

 

Yes, when the expected force is within 500 kgf and a suitable test frame, fixture and safety guard are used.

 

Can it test bolts and anchors?

 

It can measure pull-out or tensile force when the sample and sensor are installed in an appropriate fixture.

 

Can it certify lifting hooks?

 

Not by itself. Formal certification may require a prescribed test machine, accredited procedure and specific documentation.

 

Can it measure compression force?

 

Yes. It supports both tension and compression measurement when correctly installed.

 

Can it capture the maximum force before failure?

 

Yes. Peak-value mode records the highest force reached during the test.

 

Does it include all required fixtures?

 

Specialised fixtures are selected according to the component. Confirm the supplied accessories and arrange custom fixtures where necessary.

 

Select the Correct Component-Testing System in Malaysia

 

Before selecting a force gauge, send MTM Precision:

 

Component photo and dimensions

 

Material and mounting method

 

Expected test force

 

Pull or compression requirement

 

Proof or destructive test requirement

 

Applicable standard

 

Required fixture

 

Number of tests per day

 

Calibration and documentation requirements

 

MTM Precision can help evaluate whether the Landtek FM-207-500K is appropriate for the intended internal testing application.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd. (744811-A)

 

Telephone: +603-8080 7172

WhatsApp: +6019-665 7346 / +6016-660 7346

Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my

Website: www.mtmpre.com.my

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar Lapan Belas,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Supply and technical support are available throughout Malaysia, including Selangor, Kuala Lumpur, Johor, Penang, Perak, Melaka, Negeri Sembilan, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sarawak and Sabah.


 

16 Sep 2026