Landtek FM-207-500K for Cable, Wire and Rope Pull Testing Malaysia
Landtek FM-207-500K for Cable, Wire and Rope Pull Testing Malaysia
Cable, wire and rope assemblies often need to withstand a specified pulling force without slipping, separating or breaking prematurely.
The Landtek FM-207-500K Digital Force Gauge can measure tensile force up to ±500 kgf, ±4900 N or ±1100 lbf. Its external sensor configuration makes it suitable for integration with a test stand or customised pull-testing fixture.
The instrument can support incoming inspection, production quality control, product development and failure investigation for manufacturers and engineering companies in Malaysia.
What Is a Cable Pull Test?
A cable pull test measures the tensile force applied to a cable, wire, rope or connection.
Depending on the test objective, the operator may need to determine:
The force required to pull a wire from a terminal
The strength of a crimped connection
The maximum load before a cable breaks
Whether a connector remains attached at a specified force
The pull-out strength of an anchor or fitting
Whether a finished assembly meets an internal specification
The difference between production batches
The test may be performed as a proof test or a destructive test.
Proof Test vs Destructive Pull Test
Proof Test
A proof test applies a specified load without intentionally breaking the sample.
For example, a cable assembly may be required to withstand 100 kgf for a defined period. If the assembly remains secure and undamaged, it passes the test.
Real-time measurement is useful because the operator can gradually increase the load and hold it at the required value.
Destructive Test
A destructive test continues until the sample slips, separates or breaks.
The purpose is to identify:
Maximum pull force
Breaking strength
Failure point
Weakest component
Batch-to-batch consistency
The FM-207-500K’s peak-value mode can capture the highest force reached before failure.
Typical Cable, Wire and Rope Applications
The FM-207-500K may be used for:
Electrical cable pull tests
Wire-terminal pull-out tests
Crimp connection testing
Wire-harness inspection
Steel cable testing
Synthetic rope testing
Safety-lanyard component testing
Cable-grip performance testing
Connector retention-force measurement
Hook and end-fitting testing
Anchor and attachment testing
Custom cable-assembly quality control
The suitability of the instrument depends on the expected force, sample dimensions, grip design and applicable test procedure.
Why Grip Selection Is Important
The force gauge measures the force transmitted through the sensor, but the accuracy and repeatability of the complete test also depend heavily on the grip.
An unsuitable grip may allow the sample to:
Slip during the test
Break at the wrong location
Become damaged before the intended test begins
Move out of alignment
Produce inconsistent results
Different samples require different gripping methods.
Possible solutions include:
Cable clamps
Wedge grips
Capstan-style grips
Vise grips
Clevis connections
Hooks and shackles
Custom-machined fixtures
Soft, flexible or coated cables may require a grip that distributes pressure over a larger surface to avoid cutting or crushing the sample.
External Sensor for High-Capacity Testing
The FM-207-500K uses an external sensor rather than relying on a sensor built into a small handheld body.
This provides several advantages for cable and rope testing:
The sensor can be mounted directly to the test structure
The display can remain away from the moving test area
The system can accommodate larger samples and fixtures
The sensor can be connected to a manual, motorised or hydraulic loading system
The operator can monitor readings from a more convenient position
Custom test arrangements can be developed for different products
All sensor connections must be correctly aligned with the pulling direction.
Side loading, twisting and bending may affect the result or damage the sensor.
How to Perform a Cable Pull Test
1. Define the Test Requirement
Determine:
Sample type
Expected force
Pass or fail limit
Proof or destructive test
Required measurement unit
Loading speed
Holding time
Applicable standard or internal procedure
Without a defined method, results from different samples may not be directly comparable.
2. Select the Correct Force-Gauge Range
The FM-207-500K is suitable for forces up to 500 kgf.
For very small terminal or electronic-wire tests, a lower-capacity gauge may provide better resolution. For forces that may exceed 500 kgf, a higher-capacity system is required.
3. Inspect the Fixture
Confirm that the stand, grips, hooks and connectors can safely support the expected load.
The capacity of the overall system is determined by its weakest component.
4. Install the Sample
Secure the sample so it cannot slip unexpectedly.
The sample, grip and sensor should follow the same straight loading axis.
Avoid bending the cable around an unsuitable sharp edge unless the test method specifically requires it.
5. Zero the Instrument
After the sensor and fixtures have been installed, clear any initial force before starting the test.
The display should remain stable at zero.
6. Select the Measurement Mode
Use:
Real-time mode for a proof load or holding-force test
Peak mode for a pull-out, separation or breaking-force test
7. Apply the Load Gradually
Increase the pulling force smoothly.
Avoid sudden jerking because impact loading may temporarily produce a much higher force than intended.
Observe the fixture and stop if it bends, slips or becomes misaligned.
8. Record the Result and Failure Mode
Record more than the force value alone.
A useful pull-test report may include:
Sample identification
Cable or wire size
Terminal or connector type
Grip used
Maximum force
Measurement unit
Loading method
Test speed
Pass or fail result
Location and type of failure
For example, a wire breaking in the middle may indicate a different condition from a terminal pulling out of its crimp.
Real-Time and Peak Measurement
The FM-207-500K supports both real-time and peak-force measurement.
Use Real-Time Mode When:
Applying a specified proof load
Holding a cable at a target force
Monitoring force changes
Adjusting cable tension
Checking gradual movement or slippage
Use Peak Mode When:
Pulling a terminal from a wire
Testing maximum connector retention
Measuring breaking strength
Comparing destructive-test samples
Identifying the highest force before failure
The instrument can store up to 99 measurement records, helping operators retain results from multiple tests.
Common Pull-Testing Problems
The Cable Slips from the Grip
The grip may not match the cable material, diameter or surface. A different grip design or greater contact area may be required.
The Cable Breaks Beside the Clamp
Excessive clamping pressure or a sharp grip edge may be weakening the sample.
Results Vary Between Operators
Differences in sample positioning, loading speed, grip tension or alignment can cause inconsistent results.
The Reading Does Not Return to Zero
Check for residual tension, fixture binding, sensor misalignment or an unstable mounting structure.
The Reading Jumps Suddenly
The sample may be slipping, or the loading mechanism may be applying force unevenly.
The Overload Warning Activates
Stop the test immediately and remove the load safely. The expected force may exceed the selected gauge range.
Cable Pull-Test Safety
Cable and rope samples can release stored energy when they break.
Recommended precautions include:
Keep personnel away from the direct pulling line
Use a suitable protective guard
Inspect cables and fixtures before testing
Confirm the load rating of hooks and connectors
Replace damaged grips
Avoid standing directly behind the sample
Apply force gradually
Wear appropriate personal protective equipment
Stop the test if the fixture becomes unstable
The instrument’s overload protection should not replace proper test planning.
Frequently Asked Questions
Can the FM-207-500K test wire terminals?
Yes, if the expected pull-out force falls within its useful range. Smaller electronic terminals may be better tested with a lower-capacity force gauge.
Can it test steel wire rope?
Potentially, provided the force remains below 500 kgf and a suitable high-capacity grip and stand are used.
Does the gauge include cable grips?
The required grips depend on the sample. Confirm the standard package and select specialised or customised grips separately.
Can it record the breaking force?
Yes. Peak-value mode retains the maximum force reached during the test.
Can it measure cable tension after installation?
The FM-207-500K measures force applied directly through its sensor. Whether it suits an installed cable depends on how the sensor can be incorporated into the loading arrangement.
Can the results be transferred to a computer?
USB data output or optional Bluetooth communication may be available, depending on the supplied accessories and configuration.
Choose a Cable Pull-Testing System in Malaysia
A complete pull-testing solution includes more than the force gauge. The sensor capacity, test stand, grip design and test procedure must work together.
To evaluate your application, send MTM Precision:
Photo of the cable, wire or rope
Sample diameter and length
Connector or terminal details
Expected pulling force
Proof or destructive test requirement
Required loading speed
Number of tests per day
Applicable test method
Calibration or reporting requirements
MTM Precision can help determine whether the Landtek FM-207-500K and a suitable fixture can meet the application.
Contact MTM Precision
MTM Precision Sdn. Bhd. (744811-A)
Telephone: +603-8080 7172
WhatsApp: +6019-665 7346 / +6016-660 7346
Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my
Website: www.mtmpre.com.my
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar Lapan Belas,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.
Supply and technical support are available throughout Malaysia, including Selangor, Kuala Lumpur, Johor, Penang, Perak, Melaka, Negeri Sembilan, Pahang, Kedah, Perlis, Terengganu, Kelantan, Sarawak and Sabah.
16 Sep 2026