Qiwei PHS-25 E201 pH Electrode Care and Replacement Guide Malaysia

Qiwei PHS-25 E201 pH Electrode Care and Replacement Guide Malaysia

 

The electrode is the most important consumable component of the Qiwei PHS-25 benchtop pH meter.

 

A meter may switch on and display normally, but an aged, dry or contaminated E201 electrode can still produce slow, unstable or inaccurate readings.

 

Correct cleaning and storage can improve measurement reliability and extend electrode service life. However, a pH electrode does not last permanently and will eventually require replacement.

 

What Is the E201 Combination Electrode?

 

The E201 is a combination pH electrode used with the Qiwei PHS-25.

 

It combines the pH-sensitive glass electrode and reference electrode in one assembly. When immersed in a suitable liquid sample, it generates a small electrical signal that the meter converts into a pH value.

 

The sensitive parts normally include:

 

Glass sensing bulb

 

Reference junction

 

Internal reference system

 

Electrode body

 

Cable and connector

 

Damage or contamination affecting any of these components can cause poor measurement performance.

 

Common Signs of an Electrode Problem

 

The electrode may need cleaning, conditioning or replacement if:

 

The pH reading moves continuously.

 

The result takes a long time to stabilise.

 

Repeated tests produce different values.

 

The meter cannot reach the buffer value.

 

Calibration adjustment is insufficient.

 

The reading remains close to neutral in different samples.

 

The display changes when the cable is moved.

 

Results become abnormal after testing dirty samples.

 

Fresh buffer solutions still produce unstable results.

 

The glass bulb is cracked or chipped.

 

The main instrument should not be replaced before the electrode has been properly checked.

 

Rinse the Electrode Between Samples

 

Rinse the E201 electrode with distilled or deionised water before and after each measurement.

 

This helps prevent:

 

Carryover between samples

 

Buffer contamination

 

Chemical mixing

 

Deposits on the glass bulb

 

Incorrect comparison between samples

 

After rinsing, gently blot away excess water using clean laboratory tissue.

 

Do not rub the glass bulb aggressively. Rubbing may create static charge or damage the sensitive surface.

 

Do Not Store the Electrode Dry

 

The glass bulb must remain hydrated for proper response.

 

When an electrode is stored dry, it may:

 

Respond very slowly

 

Drift continuously

 

Fail calibration

 

Require lengthy reconditioning

 

Become permanently damaged

 

Use the storage solution recommended for the electrode and make sure the sensitive bulb remains protected during storage.

 

Do Not Store It Permanently in Distilled Water

 

Distilled or deionised water is appropriate for rinsing but generally unsuitable for long-term pH-electrode storage.

 

Long-term storage in pure water can affect the reference system and reduce electrode performance.

 

If the electrode has been stored incorrectly, place it in the recommended storage solution and allow sufficient conditioning time before recalibration.

 

Cleaning According to the Sample

 

Different samples create different types of contamination.

 

General Water Samples

 

Rinse the electrode thoroughly with distilled or deionised water.

 

Wastewater and Pond Water

 

Inspect the bulb and junction for sludge, algae and suspended solids. Use a suitable cleaning method before rinsing and conditioning the electrode.

 

Oil or Grease

 

Oil can form a film over the glass surface. A cleaning solution compatible with the electrode and contaminant may be required.

 

Food and Protein Samples

 

Protein deposits may block the reference junction or coat the glass membrane. Use the electrode manufacturer’s recommended cleaning procedure.

 

Mineral or Scale Deposits

 

Deposits should be removed using an appropriate electrode-cleaning method. Do not scrape the bulb with hard or abrasive tools.

 

After cleaning, rinse the electrode thoroughly, place it in the correct storage solution and recalibrate the meter.

 

Handling the Glass Bulb

 

The sensing bulb is fragile.

 

To prevent damage:

 

Use the supplied electrode holder.

 

Do not drop the electrode into the beaker.

 

Do not allow the bulb to strike the container.

 

Do not use the electrode as a stirring rod.

 

Do not scrape deposits using sharp tools.

 

Do not force the electrode into solid or semi-solid samples.

 

Keep the protective cap when transporting or storing it.

 

A cracked glass bulb cannot be repaired and the electrode must be replaced.

 

Protect the Cable and Connector

 

An electrode can fail even when the glass bulb appears undamaged.

 

Check for:

 

Cut or crushed cable

 

Loose connector

 

Bent contact

 

Moisture inside the connector

 

Chemical residue

 

Damage caused by pulling the cable

 

Disconnect the electrode by holding the connector, not by pulling on the cable.

 

Keep the meter’s input socket clean and dry.

 

Use Fresh Calibration Buffers

 

The electrode cannot be evaluated properly using contaminated or expired buffer.

 

The Qiwei PHS-25 uses standard buffer values including:

 

pH 4.00

 

pH 6.86

 

pH 9.18

 

Pour a small amount into a clean beaker for calibration. Do not insert a dirty electrode into the original buffer bottle and do not pour used solution back into it.

 

How to Check Whether the Electrode Is Still Working

 

A practical check can be performed using fresh calibration buffers.

 

Inspect the electrode for physical damage.

 

Confirm that the bulb has been properly hydrated.

 

Prepare fresh buffer solutions.

 

Rinse and gently blot the electrode.

 

Place it in pH 6.86 buffer.

 

Allow the reading to stabilise.

 

Rinse the electrode.

 

Repeat the test in pH 4.00 or pH 9.18 buffer.

 

Observe response speed and reading stability.

 

Recalibrate the meter if the response is acceptable.

 

A functioning electrode should show a clear response when moved between buffers with substantially different pH values.

 

Compare With a Known Working Electrode

 

If possible, connect a compatible known working electrode to the PHS-25.

 

If the meter works normally with the replacement electrode, the original electrode is probably defective.

 

If both electrodes produce the same abnormal behaviour, the main instrument, connection, calibration procedure or test environment may require further checking.

 

This comparison helps prevent unnecessary replacement of the meter.

 

When Should the E201 Electrode Be Replaced?

 

Replacement is recommended when:

 

The glass bulb is cracked.

 

The cable or connector is damaged.

 

The electrode repeatedly fails calibration.

 

Response remains extremely slow after conditioning.

 

Readings remain unstable in fresh buffer.

 

Results cannot be repeated.

 

Cleaning does not restore performance.

 

Calibration adjustment has reached its limit.

 

A known working electrode operates normally on the same meter.

 

The exact service life depends on usage, storage and sample conditions. There is no fixed lifetime that applies to every laboratory.

 

Factors That Shorten Electrode Life

 

Electrode life may be reduced by:

 

Continuous use

 

Frequent testing of dirty samples

 

Strong acids or alkalis

 

High-temperature samples

 

Oily or protein-rich liquids

 

Improper storage

 

Long periods of dry storage

 

Physical impact

 

Abrasive cleaning

 

Chemical attack

 

Blocked reference junction

 

Laboratories working with difficult samples should consider keeping a spare electrode available.

 

Daily Electrode-Care Checklist

 

Before use:

 

Inspect the bulb.

 

Confirm that the electrode is hydrated.

 

Check the connector.

 

Prepare fresh buffer.

 

Calibrate the meter.

 

During use:

 

Rinse between samples.

 

Avoid cross-contamination.

 

Keep the bulb adequately immersed.

 

Wait for a stable result.

 

Do not strike the beaker.

 

After use:

 

Rinse the electrode.

 

Clean contamination when required.

 

Inspect the junction.

 

Place it in the correct storage solution.

 

Protect the glass bulb.

 

Keep the connector clean and dry.

 

Should You Buy a Spare Electrode?

 

A spare electrode is useful when:

 

The meter is used every day.

 

Testing cannot stop while a replacement is ordered.

 

The laboratory handles many samples.

 

Wastewater or dirty samples are frequently tested.

 

Several employees use the same instrument.

 

The laboratory is located far from the supplier.

 

The meter is used for important internal QC work.

 

A spare also helps identify whether an unstable result originates from the electrode or the main instrument.

 

Qiwei PHS-25 and E201 Electrode Support in Malaysia

 

The Qiwei PHS-25 benchtop pH meter is available from MTM Precision in Malaysia.

 

Customers may contact us to check the availability of the instrument and compatible replacement electrodes.

 

For troubleshooting assistance, send us:

 

Photo of the instrument model

 

Photo of the electrode

 

Buffer values being used

 

Description of the problem

 

Approximate electrode age

 

Sample type

 

Short video showing the unstable or slow reading

 

This information helps determine whether the electrode requires cleaning, conditioning or replacement.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd. (744811-A)

 

Telephone: +603-8080 7172

WhatsApp: +6016-660 7346

Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my

Website: www.mtmpre.com.my

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Business Hours: Monday–Friday, 9:00 AM–6:00 PM


 

19 Sep 2026