Qiwei PHS-25 for Wastewater pH Testing in Factory Laboratories Malaysia

Qiwei PHS-25 for Wastewater pH Testing in Factory Laboratories Malaysia

 

The Qiwei PHS-25 benchtop pH meter can be used for routine wastewater pH checks in factory laboratories, treatment facilities and internal quality-control stations.

 

It is an economical laboratory-style instrument with pH and mV functions, a separate E201 combination electrode and an electrode holder.

 

The PHS-25 is best suited for internal monitoring and trend checking at a fixed testing station. Customers requiring results for regulatory submission or accredited reporting must first confirm the applicable test method, accuracy and calibration-traceability requirements.

 

Why Monitor Wastewater pH?

 

pH is one of the basic parameters used to observe wastewater conditions and treatment performance.

 

Abnormal pH may:

 

Affect biological treatment processes

 

Reduce treatment efficiency

 

Indicate a chemical-dosing problem

 

Damage pumps, pipes or treatment equipment

 

Affect coagulation and precipitation

 

Change the behaviour of dissolved contaminants

 

Affect the toxicity of certain substances

 

Indicate an unexpected process discharge

 

Monitoring pH at several treatment stages helps operators identify changes before the final discharge point.

 

Qiwei PHS-25 Main Specifications

 

pH measuring range: 0.00–14.00 pH

 

pH resolution: 0.01 pH

 

pH accuracy: ±0.05 pH

 

mV measuring range: 0 to ±1999 mV

 

mV resolution: 1 mV

 

mV accuracy: ±10 mV

 

Temperature-compensation range: 0–60°C

 

Standard buffer values: pH 4.00, pH 6.86 and pH 9.18

 

Standard electrode: E201 combination pH electrode

 

Power supply: AC power adaptor

 

Approximate dimensions: 220 × 200 × 85 mm

 

Approximate weight: 1 kg

 

Suitable Wastewater Applications

 

The Qiwei PHS-25 may be considered for routine samples collected from:

 

Equalisation tanks

 

Neutralisation tanks

 

Biological treatment systems

 

Chemical-treatment processes

 

Settling tanks

 

Process-water drains

 

Final treated water

 

Internal wastewater laboratories

 

Incoming and outgoing treatment stages

 

General factory water systems

 

The instrument should be installed in a clean, dry testing area rather than beside an exposed treatment tank.

 

Why Use a Benchtop Meter?

 

A benchtop pH meter is useful when samples from different treatment points are brought to a central laboratory.

 

Advantages include:

 

Stable testing position

 

Separate and replaceable electrode

 

Dedicated electrode holder

 

AC-powered operation

 

Convenient testing of multiple samples

 

pH and mV measurement

 

Easier implementation of a standard laboratory procedure

 

For direct measurement at tanks, drains or outdoor treatment areas, a waterproof portable pH meter may be more suitable.

 

Suggested Wastewater Sampling Points

 

A factory may monitor pH at several locations, depending on its treatment design.

 

Possible sampling points include:

 

Raw wastewater inlet

 

Equalisation tank

 

After acid or alkali dosing

 

Biological treatment stage

 

Clarification or settling stage

 

Final treatment stage

 

Discharge point

 

Using consistent sampling locations allows the operator to compare results and identify where a pH change occurs.

 

Basic Testing Procedure

 

Collect the sample using a clean, suitable container.

 

Label the sampling location, date and time.

 

Test the sample as soon as practical.

 

Calibrate the PHS-25 with fresh buffer solution.

 

Rinse the E201 electrode with distilled or deionised water.

 

Gently blot away excess rinse water.

 

Immerse the electrode bulb and reference junction in the sample.

 

Move the electrode gently to release trapped air.

 

Allow the reading to stabilise.

 

Record the pH value and sample temperature.

 

Rinse the electrode before testing the next sample.

 

Clean and store the electrode correctly after use.

 

Follow the facility’s safety and sampling procedures when handling unknown or hazardous wastewater.

 

Selecting the Correct Calibration Buffers

 

The calibration buffers should cover the expected sample range.

 

For Acidic Wastewater

 

Use:

 

pH 6.86

 

pH 4.00

 

For Alkaline Wastewater

 

Use:

 

pH 6.86

 

pH 9.18

 

For Samples Near Neutral

 

Start with the pH 6.86 buffer and use an appropriate second point when better confidence across the operating range is required.

 

Always use fresh, uncontaminated buffer solution.

 

Challenges When Measuring Wastewater

 

Wastewater is often more difficult to measure than clean water because it may contain suspended solids, oil, chemicals and biological material.

 

Suspended Solids

 

Particles can collect around the electrode bulb or reference junction and produce a slow response.

 

Oil and Grease

 

Oil may coat the glass surface and prevent proper contact with the sample.

 

High or Low Temperature

 

A sample that is much hotter or colder than the calibration buffer may take longer to stabilise.

 

Chemical Residue

 

Strong chemicals may affect electrode performance or shorten its service life.

 

Electrical Interference

 

Motors, pumps and treatment equipment may introduce interference if the meter is installed too close to industrial machinery.

 

Changing Samples

 

Some wastewater continues reacting after collection. Delayed testing may produce a result that does not represent conditions at the sampling point.

 

Cleaning the Electrode

 

After testing wastewater:

 

Rinse the electrode thoroughly.

 

Inspect the glass bulb and junction.

 

Remove visible contamination using a suitable cleaning method.

 

Recondition the electrode when necessary.

 

Recalibrate before the next important measurement.

 

Store it in the correct electrode storage solution.

 

Do not use abrasive materials on the glass bulb.

 

Different contaminants may require different cleaning solutions. The operator should follow the electrode manufacturer’s recommendations and the facility’s chemical-safety procedure.

 

When Should the Electrode Be Replaced?

 

Consider electrode replacement when:

 

Calibration repeatedly fails.

 

Readings remain unstable in fresh buffer.

 

Response is extremely slow.

 

Results cannot be repeated.

 

The glass bulb is cracked.

 

The cable or connector is damaged.

 

Cleaning and rehydration do not restore performance.

 

Testing the PHS-25 with a known working electrode can help determine whether the fault comes from the electrode or the main instrument.

 

PHS-25 vs Online pH Controller

 

The Qiwei PHS-25 is a laboratory instrument. It is not an online pH controller.

 

FunctionQiwei PHS-25Online pH ControllerMeasurement methodCollected sampleContinuous process measurementInstallationLaboratory benchTank, pipe or process linePowerAC adaptorFixed electrical installationContinuous outputNoModel dependentDosing controlNoAvailable on selected systemsBest useRoutine sample testingContinuous monitoring and process control

 

Factories needing continuous pH monitoring, alarms or chemical-dosing control should consider an industrial online pH system.

 

Is the PHS-25 Suitable for Regulatory Reporting?

 

The PHS-25 has a stated accuracy of ±0.05 pH.

 

It can be useful for internal checks and observing treatment trends, but regulatory or accredited testing may require:

 

A specified test method

 

Suitable instrument accuracy

 

Traceable calibration

 

Certified buffer solutions

 

Documented quality-control procedures

 

Maintenance and calibration records

 

Laboratory accreditation

 

Customers should verify these requirements before using any instrument for formal compliance reporting.

 

Other Wastewater Parameters to Consider

 

pH alone does not provide a complete picture of treatment performance.

 

Depending on the application, additional parameters may include:

 

Chemical oxygen demand

 

Dissolved oxygen

 

Ammonia nitrogen

 

Total phosphorus

 

Total nitrogen

 

Turbidity

 

Suspended solids

 

Conductivity

 

Oxidation-reduction potential

 

Residual chlorine

 

Temperature

 

MTM Precision can assist customers in identifying instruments for a broader routine water-quality testing setup.

 

Qiwei PHS-25 Available in Malaysia

 

The Qiwei PHS-25 benchtop pH meter is available from MTM Precision in Malaysia.

 

Factories and wastewater operators may contact us for stock confirmation, package details, replacement-electrode availability and assistance choosing between benchtop, portable and online pH instruments.

 

For a suitable recommendation, please provide:

 

Type of wastewater

 

Expected pH range

 

Sample temperature

 

Testing location

 

Testing frequency

 

Required reporting standard

 

Whether continuous monitoring is needed

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd. (744811-A)

 

Telephone: +603-8080 7172

WhatsApp: +6016-660 7346

Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my

Website: www.mtmpre.com.my

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Business Hours: Monday–Friday, 9:00 AM–6:00 PM


 

19 Sep 2026