Qiwei PHS-25 pH Meter Reading Unstable or Slow? Troubleshooting Malaysia

Qiwei PHS-25 pH Meter Reading Unstable or Slow? Troubleshooting Malaysia

 

Is your Qiwei PHS-25 benchtop pH meter showing fluctuating values, taking too long to stabilise or failing to produce repeatable readings?

 

An unstable pH result does not automatically mean that the main instrument is faulty. In many cases, the problem is caused by the E201 electrode, contaminated buffer solution, incorrect calibration, sample conditions or improper electrode storage.

 

The following checks can help identify the likely cause.

 

Common PHS-25 Measurement Problems

 

Users may encounter symptoms such as:

 

pH value continuously moving up and down

 

Reading takes several minutes to stabilise

 

Different result when the same sample is tested again

 

Instrument cannot reach the buffer value

 

Calibration adjustment is insufficient

 

Reading remains around pH 7

 

Display changes when the electrode cable is moved

 

Results become inaccurate after testing dirty samples

 

pH reading changes significantly between hot and cold samples

 

Each symptom may have a different cause, so the electrode and testing procedure should be checked before replacing the instrument.

 

1. The E201 Electrode Is Dry

 

The glass bulb of a pH electrode must remain properly hydrated.

 

If the electrode has been stored dry, its response may become very slow or unstable. A severely dehydrated electrode may fail to calibrate completely.

 

What to Do

 

Inspect the glass bulb for damage.

 

Soak the electrode in the appropriate electrode storage solution.

 

Allow sufficient time for rehydration.

 

Rinse the electrode before calibration.

 

Calibrate again using fresh buffer solution.

 

Do not attempt to solve a dry-electrode problem by storing it permanently in distilled water.

 

2. The Electrode Was Stored in Distilled Water

 

Distilled or deionised water is suitable for rinsing, but it is generally unsuitable for long-term pH-electrode storage.

 

Long-term storage in pure water may affect the reference electrolyte and reduce electrode performance.

 

What to Do

 

Move the electrode to the correct storage solution and allow it to condition before recalibration. If the electrode remains unstable, replacement may be necessary.

 

3. The Glass Bulb Is Dirty

 

Deposits on the electrode surface can prevent proper contact between the glass membrane and the sample.

 

Common contamination includes:

 

Oil and grease

 

Protein

 

Food residue

 

Chemical deposits

 

Scale and minerals

 

Soil and suspended particles

 

Biological growth

 

What to Do

 

Clean the electrode using a solution suitable for the type of contamination. After cleaning, rinse it thoroughly, recondition it in storage solution and recalibrate the meter.

 

Avoid scrubbing the sensitive glass bulb with abrasive material.

 

4. The Buffer Solution Is Old or Contaminated

 

Calibration cannot be more reliable than the buffer solution used.

 

Buffer contamination often occurs when:

 

Used buffer is poured back into the bottle.

 

The electrode is moved between buffers without rinsing.

 

The buffer container is left open.

 

Dirty droppers or tools are inserted into the solution.

 

The buffer has passed its recommended storage period.

 

Different buffer solutions are mixed accidentally.

 

What to Do

 

Use fresh buffer in a clean beaker. Discard the used portion after calibration and do not return it to the original container.

 

The Qiwei PHS-25 uses standard buffer values including:

 

pH 4.00

 

pH 6.86

 

pH 9.18

 

5. The Meter Was Not Calibrated for the Sample Range

 

A meter calibrated only around neutral pH may be less reliable when testing a strongly acidic or alkaline sample.

 

Recommended Buffer Selection

 

For acidic samples: pH 6.86 and pH 4.00

 

For alkaline samples: pH 6.86 and pH 9.18

 

For samples near neutral: Start with pH 6.86

 

Calibration should cover the expected measuring range whenever possible.

 

6. Air Bubbles Are Trapped Around the Electrode

 

An air bubble around the glass bulb or reference junction can cause an unstable or incorrect reading.

 

What to Do

 

Gently move the electrode through the sample or tap it carefully to release trapped air. Do not hit the glass bulb against the beaker.

 

Ensure that the sensitive section of the electrode is adequately immersed.

 

7. The Sample Has Low Conductivity

 

Low-conductivity water can be difficult to measure with a general-purpose pH electrode.

 

Examples may include:

 

Distilled water

 

Deionised water

 

Reverse-osmosis water

 

Very clean process water

 

Low-mineral drinking water

 

Because the sample has very low ionic strength, the displayed pH may drift or take a long time to stabilise.

 

This does not necessarily mean that the meter is defective. Low-conductivity samples may require a suitable electrode and controlled testing procedure.

 

8. The Sample Temperature Is Changing

 

Temperature affects the response of a pH electrode and may also change the actual pH of the sample.

 

If the calibration buffer and sample are at very different temperatures, the result may take longer to stabilise or may not be directly comparable.

 

What to Do

 

Allow buffers and samples to reach a stable temperature.

 

Confirm the temperature-compensation setting.

 

Record the sample temperature together with the pH result.

 

Avoid measuring immediately after moving a sample from refrigeration or heating.

 

The PHS-25 specification indicates a temperature-compensation range of 0–60°C, but the operator should confirm and set the applicable temperature correctly.

 

9. The Electrode Connector Is Loose or Wet

 

Moisture, dirt or a loose connection at the electrode input can cause fluctuating readings.

 

What to Do

 

Switch off the instrument.

 

Inspect the electrode plug and input connector.

 

Make sure the connector is clean and dry.

 

Reconnect the electrode securely.

 

Keep the connection away from spilled buffer and sample liquids.

 

Do not pull the electrode by its cable.

 

10. Strong Electrical Interference Is Present

 

The PHS-25 measures a very small electrical signal from the electrode. Strong electromagnetic interference may affect the stability of the reading.

 

Possible sources include:

 

Motors

 

Pumps

 

High-power electrical equipment

 

Magnetic stirrers

 

Unstable power supplies

 

Damaged extension cables

 

Strong laboratory electrical equipment

 

What to Do

 

Move the pH meter away from the suspected interference source and test it again using a fresh buffer solution.

 

The manufacturer recommends operating the instrument in an area without significant vibration or strong electromagnetic interference.

 

11. The Electrode Has Reached the End of Its Service Life

 

A pH electrode is a consumable component. Its response gradually deteriorates with use, age, storage conditions and sample exposure.

 

Replacement may be required when:

 

Calibration repeatedly fails.

 

Response remains slow after cleaning and rehydration.

 

Readings cannot be repeated.

 

The glass bulb is cracked.

 

The cable or connector is damaged.

 

The electrode remains unstable in fresh buffer.

 

Calibration adjustment has reached its limit.

 

Testing the main unit with a known working electrode can help separate an electrode problem from an instrument problem.

 

Quick Troubleshooting Test

 

Follow this sequence before sending the instrument for checking:

 

Inspect the electrode for physical damage.

 

Confirm that the electrode is hydrated.

 

Prepare fresh pH 6.86 buffer.

 

Rinse the electrode with distilled water.

 

Connect the electrode securely.

 

Place it in the fresh buffer.

 

Allow sufficient stabilisation time.

 

Repeat the test using another fresh buffer.

 

Compare the behaviour with a known working electrode, if available.

 

If the meter works normally with another electrode, the original E201 electrode is probably the source of the problem.

 

When Should the Main Instrument Be Checked?

 

The PHS-25 main unit may require inspection if:

 

The display does not switch on.

 

The reading does not respond to any connected electrode.

 

Values remain abnormal with a known working electrode.

 

The input connector is damaged.

 

The controls cannot complete calibration.

 

The display is incomplete or intermittent.

 

The instrument has suffered liquid entry or electrical damage.

 

Do not open the instrument while it is connected to the power supply.

 

Qiwei PHS-25 Support in Malaysia

 

The Qiwei PHS-25 benchtop pH meter is available from MTM Precision in Malaysia.

 

If your reading is unstable, customers may WhatsApp us a clear photo or short video showing:

 

Instrument model

 

Displayed reading

 

Buffer value being tested

 

Condition of the electrode

 

Electrode connection

 

Calibration procedure being used

 

This information helps determine whether the likely issue involves calibration, buffer solution, electrode condition or the main instrument.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd. (744811-A)

 

Telephone: +603-8080 7172

WhatsApp: +6016-660 7346

Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my

Website: www.mtmpre.com.my

 

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia.

 

Business Hours: Monday–Friday, 9:00 AM–6:00 PM


 

19 Sep 2026