How to Calibrate Qiwei PHS-25 Benchtop pH Meter Malaysia
How to Calibrate Qiwei PHS-25 Benchtop pH Meter Malaysia
Regular calibration is essential for obtaining reliable results from the Qiwei PHS-25 benchtop pH meter.
Even when the instrument is operating correctly, the response of its E201 glass electrode can gradually change because of ageing, contamination, dehydration and frequent exposure to different samples.
Calibration adjusts the meter according to known buffer values before actual samples are measured.
Qiwei PHS-25 Calibration Buffers
The Qiwei PHS-25 is supplied for use with the following standard buffer values:
pH 4.00
pH 6.86
pH 9.18
The appropriate calibration points depend on the expected pH range of the sample.
For Acidic Samples
Use pH 6.86 and pH 4.00 buffers.
Suitable examples include acidic process solutions, certain food samples and agricultural solutions.
For Alkaline Samples
Use pH 6.86 and pH 9.18 buffers.
Suitable examples include alkaline cleaning solutions and selected industrial-process samples.
For Samples Around Neutral pH
Start with the pH 6.86 buffer. A second calibration point is recommended when better reliability across a wider measuring range is required.
Items Required for Calibration
Prepare the following before starting:
Qiwei PHS-25 main unit
E201 combination pH electrode
Fresh pH buffer solutions
Clean beakers
Distilled or deionised water
Clean, soft laboratory tissue
Thermometer if the instrument requires manual temperature entry
Electrode storage solution
Do not calibrate the meter using old, contaminated or previously mixed buffer solutions.
Step 1: Inspect the Electrode
Check the E201 electrode before calibration.
Make sure:
The glass bulb is not cracked.
The electrode cable and connector are undamaged.
The sensing bulb is clean and hydrated.
There are no visible deposits on the glass surface.
The electrode has been connected securely to the meter.
A dry electrode may respond very slowly or fail to calibrate. It may need to be soaked in the correct storage solution before use.
Step 2: Prepare Fresh Buffer Solutions
Pour a small amount of each required buffer into separate, clean beakers.
Never insert an unwashed electrode directly into the original buffer bottle. Contamination can change the buffer value and affect future calibrations.
Do not return used buffer solution to its original container.
Step 3: Switch On and Allow the Meter to Stabilise
Connect the Qiwei PHS-25 to its power adaptor and switch it on.
Allow the instrument and electrode to stabilise according to the operating instructions. Calibration is best performed in a stable indoor environment without strong vibration or electromagnetic interference.
The manufacturer specifies an operating environment of approximately 5–40°C with relative humidity not exceeding 80%.
Step 4: Rinse the Electrode
Rinse the electrode with distilled or deionised water.
Gently blot away excess water using clean laboratory tissue. Do not rub the glass bulb because rubbing may create static charge or damage the sensitive glass surface.
Step 5: Calibrate at the Neutral Point
Immerse the electrode bulb and reference junction in the pH 6.86 buffer.
Ensure that:
The sensitive portion of the electrode is adequately immersed.
The electrode does not touch the bottom or side of the beaker.
No air bubble is trapped around the glass bulb.
The reading is allowed sufficient time to stabilise.
Adjust or confirm the instrument according to the PHS-25 operating instructions until it corresponds with the buffer value at the relevant temperature.
Step 6: Rinse Before the Second Buffer
Remove the electrode and rinse it thoroughly with distilled or deionised water.
Blot gently before transferring it into the second buffer. This prevents the neutral buffer from contaminating the acidic or alkaline buffer.
Step 7: Complete the Second Calibration Point
Use:
pH 4.00 for measurements expected below neutral, or
pH 9.18 for measurements expected above neutral.
Immerse the electrode, allow the reading to stabilise and complete the second-point adjustment according to the instrument instructions.
For samples covering both acidic and alkaline ranges, verify the meter across the relevant buffer points before testing.
Step 8: Verify the Calibration
After calibration, rinse the electrode and return it to one of the fresh buffer solutions.
The displayed reading should settle close to the buffer’s specified value, taking temperature and the stated instrument accuracy into account.
The Qiwei PHS-25 provides:
pH resolution: 0.01 pH
Stated pH accuracy: ±0.05 pH
A display resolution of 0.01 pH does not mean the meter has an accuracy of ±0.01 pH.
Step 9: Measure the Sample
Rinse the electrode again before placing it into the actual sample.
Stir the sample gently or move the electrode slightly to remove trapped air, then allow the reading to stabilise. Avoid aggressive stirring because it may introduce bubbles and cause fluctuating readings.
Record relevant information such as:
pH result
Sample temperature
Sample identification
Date and time
Calibration buffers used
Operator name, if required
How Often Should the PHS-25 Be Calibrated?
Calibration frequency depends on the application and testing frequency.
Consider calibrating:
Before each important testing session
Daily when the meter is used regularly
After measuring highly acidic or alkaline samples
After measuring dirty, oily or chemically aggressive samples
After cleaning or replacing the electrode
When the reading becomes slow or unstable
When results do not agree with a reference sample
After the instrument has not been used for a long period
For general demonstration or occasional routine testing, calibration may be performed before each group of measurements.
Why Does the PHS-25 Fail to Calibrate?
Common causes include:
Contaminated Buffer Solution
Used or contaminated buffer may no longer represent its labelled value. Replace it with fresh buffer.
Dry Glass Electrode
A dehydrated electrode can respond slowly or incorrectly. Rehydrate it using the correct electrode storage solution.
Dirty Electrode
Oil, protein, scale or chemical residue may coat the glass surface. Clean the electrode using a method suitable for the contamination.
Aged Electrode
Glass electrodes are consumable items. An old electrode may develop slow response, poor slope or unstable readings even after cleaning.
Incorrect Temperature Setting
If temperature compensation is entered manually, an incorrect setting can affect calibration and measurement.
Unstable Sample or Environment
Strong electrical interference, vibration, temperature changes or an improperly connected electrode can cause fluctuating readings.
Important Electrode Storage Advice
Do not store the E201 electrode permanently in distilled or deionised water. This can damage the reference system and reduce electrode performance.
Use the storage solution recommended for the electrode. Keep the glass bulb hydrated whenever possible.
Proper cleaning and storage can reduce calibration problems and extend electrode service life.
Qiwei PHS-25 Available in Malaysia
The Qiwei PHS-25 benchtop pH meter is available from MTM Precision in Malaysia.
Customers may contact us for current stock, package details, replacement-electrode availability and assistance selecting a suitable pH meter for laboratory, education, agriculture, aquaculture or routine factory testing.
When requesting assistance, please provide:
Instrument model
Type of sample
Expected pH range
Required accuracy
Description of the calibration problem
Photo or short video of the displayed reading
Contact MTM Precision
MTM Precision Sdn. Bhd. (744811-A)
Telephone: +603-8080 7172
WhatsApp: +6016-660 7346
Email: mtmpre@yahoo.com / enquiry@mtmpre.com.my
Website: www.mtmpre.com.my
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.
Business Hours: Monday–Friday, 9:00 AM–6:00 PM
19 Sep 2026