How to Measure RF Power Density in uW/m2 and uW/cm2 Malaysia
How to Measure RF Power Density in uW/m2 and uW/cm2 Malaysia
When measuring radio-frequency electromagnetic fields, users may see several different measurement units.
Some RF meters display electric field strength in V/m, while RF power density may be expressed in W/m2, mW/m2 or uW/cm2.
The TENMARS TM-195 3-Axis RF Field Strength Meter is designed for broadband RF electromagnetic field measurements from 50 MHz to 3.5 GHz.
Understanding the different RF measurement units can help technicians record and compare measurements correctly.
What Is RF Power Density?
RF power density describes electromagnetic power over a given area.
Common units include:
W/m2
mW/m2
uW/m2
uW/cm2
mW/cm2
RF field measurements may also be expressed as electric field strength in V/m.
When recording RF measurements, always include the measurement unit together with the number.
For example:
Reading: 500 uW/m2
is much more useful than simply writing:
Reading: 500
Without the unit, the measurement can easily be misunderstood.
What Does uW/m2 Mean?
uW/m2 means microwatts per square metre.
The letter "u" is commonly used as a computer-friendly replacement for the micro symbol.
One microwatt is one-millionth of a watt.
Therefore:
1 uW = 0.000001 W
When an RF meter displays a value in uW/m2, it is showing RF power density over one square metre.
What Does uW/cm2 Mean?
uW/cm2 means microwatts per square centimetre.
This is another unit commonly used when discussing RF power density.
Because one square metre contains 10,000 square centimetres:
1 uW/cm2 = 10,000 uW/m2
This is an important point when comparing RF measurements.
A reading displayed in uW/cm2 cannot be compared directly with another reading displayed in uW/m2 without considering the unit conversion.
Understanding W/m2, mW/m2 and uW/m2
The prefixes change the scale of the measurement.
1 W = 1,000 mW
1 mW = 1,000 uW
Therefore:
1 W/m2 = 1,000 mW/m2
and:
1 W/m2 = 1,000,000 uW/m2
The same RF power density can therefore appear as very different numbers depending on the selected unit.
RF Power Density Conversion Example
Suppose an RF power density measurement is:
0.01 W/m2
The same value can also be written as:
10 mW/m2
or:
10,000 uW/m2
All three values represent the same RF power density.
They are simply expressed using different units.
This is why the unit should always be recorded together with the measurement value.
What Is V/m?
V/m means volts per metre.
It is commonly used to express electric field strength.
For practical RF field surveys, users may choose a suitable unit and then use the same unit consistently throughout the survey.
For example, if Point A is recorded in V/m, Points B, C and D should preferably also be recorded in V/m.
This makes comparison much easier.
How to Measure RF Power Density with TENMARS TM-195
A basic RF field survey can be performed using a structured procedure.
Step 1 - Identify the RF Source
First determine what equipment you want to investigate.
Possible RF sources include:
Wi-Fi routers
Wireless access points
RF transmitters
Telecommunications equipment
Antennas
Wireless communication devices
Before measurement, confirm that the RF source of interest is within the supported frequency range of the instrument.
The TENMARS TM-195 covers 50 MHz to 3.5 GHz.
Step 2 - Select a Measurement Unit
Choose the measurement unit required for your survey.
For example:
V/m
or:
mW/m2
or:
uW/cm2
Avoid changing units unnecessarily while comparing different measurement points.
Step 3 - Select Several Measurement Locations
Do not rely on only one measurement.
For example, you can measure:
Point A - Reference location
Point B - Normal occupied position
Point C - Intermediate position
Point D - Near the RF source
Record each location clearly.
This makes it easier to compare the RF field across the area.
Step 4 - Keep the Measurement Method Consistent
Hold the meter reasonably steady and avoid unnecessarily blocking the sensor.
Try to use a similar measurement method at every location.
For example:
Same unit + similar measurement height + similar observation time
Consistency makes the comparison more meaningful.
Step 5 - Record the Number and Unit
Always record both the measurement value and its unit.
For example:
Point A: 0.35 V/m
Point B: 0.62 V/m
Point C: 1.10 V/m
Do not record only:
Without units, the measurement record may become confusing later.
Why Do RF Measurements Change?
RF readings may change even when the meter remains in the same general area.
Possible reasons include:
Distance from the RF source
Wireless traffic
Transmitter operating condition
Antenna direction
Building structures
Walls
Metal surfaces
RF reflections
Other nearby transmitters
Movement of people
Meter position
This means one instantaneous measurement may not describe the entire RF environment.
Using MAX During RF Measurements
RF signals can fluctuate.
The MAX function can help users capture a higher measured value during the observation period.
For example:
Point A - Current Reading + MAX
Point B - Current Reading + MAX
Point C - Current Reading + MAX
Using the same procedure at every location makes comparison easier.
Using AVG During RF Measurements
Average measurement can also be useful when the RF field changes over time.
Instead of looking only at the highest reading, AVG can provide additional information about the field during the observation period.
A simple survey record may include:
Location / Current / MAX / AVG / Unit
When comparing different locations, try to keep the observation period consistent.
Why Use a 3-Axis RF Meter?
RF fields may arrive at the measurement point from different directions.
Inside buildings, RF propagation can also be affected by walls, equipment, metal structures and multiple wireless transmitters.
The TENMARS TM-195 uses a 3-axis RF measurement system.
This makes practical field measurements easier when the direction of the RF field is unknown or changing.
Can TM-195 Identify the Exact RF Frequency?
No.
The TENMARS TM-195 is a broadband RF field strength meter.
It measures RF electromagnetic field strength within its supported frequency range.
It is not a spectrum analyser.
If several RF sources are operating simultaneously, the TM-195 does not automatically identify which individual frequency is responsible for the reading.
If the application requires detailed frequency identification, specialised RF test equipment may be required.
RF Power Density Meter vs Spectrum Analyser
These instruments answer different questions.
If you want to know:
"What is the RF field level at this location?"
A broadband RF field strength meter such as the TM-195 may be suitable.
If you want to know:
"Which exact RF frequencies are present?"
A spectrum analyser or other frequency-selective instrument may be more appropriate.
Understanding the measurement objective helps avoid choosing the wrong instrument.
Can RF Measurements Be Compared with Exposure Limits?
They can be compared only when the correct methodology and applicable requirements are understood.
Before interpreting an RF measurement against an occupational or public exposure limit, consider:
Applicable standard
Frequency
Exposure category
Averaging period
Measurement method
Instrument requirements
Calibration requirements
A single spot reading should not automatically be treated as a formal RF exposure compliance assessment.
Common RF Measurement Mistakes
Common mistakes include:
Recording a number without its unit
Comparing uW/m2 directly with uW/cm2
Confusing mW with uW
Changing units between measurement locations
Taking only one measurement
Ignoring the frequency range of the meter
Assuming a broadband RF meter identifies individual frequencies
Comparing measurements taken under very different conditions
A consistent measurement method provides more useful results.
Who May Need the TENMARS TM-195?
The TM-195 may be useful for:
Telecommunications technicians
Electrical contractors
Facility management teams
EHS personnel
RF equipment installers
Maintenance departments
Engineering consultants
Commercial building maintenance
Industrial facilities
General RF field surveys
The correct instrument depends on the RF source, frequency range and measurement objective.
TENMARS TM-195 RF Field Strength Meter Malaysia
The TENMARS TM-195 3-Axis RF Field Strength Meter is designed for broadband RF electromagnetic field measurements from 50 MHz to 3.5 GHz.
It can be considered for general RF surveys around compatible wireless, telecommunications and RF transmitting equipment.
If you are unsure which RF meter is suitable, tell us:
What equipment do you want to measure?
What frequency is involved?
Do you need RF field strength measurement or frequency analysis?
MTM Precision can help you select the appropriate testing instrument for your application.
Contact MTM Precision Sdn Bhd
Showroom & Service Centre: Puchong, Selangor, Malaysia
Operating Hours: Monday-Friday, 9:00 AM-6:00 PM
Tel: 03-8080 7172
WhatsApp: +6016-660 7346
Email: mtmpre@yahoo.com
Website: www.mtmpre.com.my
Serving customers throughout Malaysia, including Selangor, Kuala Lumpur, Johor, Penang, Melaka, Negeri Sembilan, Perak, Pahang, Kelantan, Terengganu, Kedah, Perlis, Sabah and Sarawak.
Keywords: RF Power Density Malaysia | TENMARS TM-195 Malaysia | uW/m2 RF Measurement | uW/cm2 RF Measurement | RF Field Strength Meter Malaysia | 3-Axis RF Meter | RF Electromagnetic Field Measurement Malaysia
20 Sep 2026