A UV LED lamp may still turn on and look normal even when its actual curing performance has changed.
For industrial applications, checking whether the lamp is illuminated is therefore not enough.
Printing companies, coating manufacturers, electronics factories and UV adhesive users may need to know:
How strong is the UV output?
How much UV energy is delivered during the curing cycle?
Has the UV output changed compared with previous production conditions?
The Linshang LS137 UV Energy Meter is designed for UV LED light sources and provides a practical method for measuring UV irradiance, accumulated UV energy and temperature.
Why Test UV LED Lamp Output?
UV LED curing systems are used in many industrial processes where curing quality directly affects the finished product.
Examples include:
UV printing
UV ink curing
UV coating
UV adhesive curing
Electronics assembly
PCB manufacturing
Optical bonding
Industrial production lines
If UV output changes, the final curing result may also change.
Periodic testing helps production teams understand whether the UV source continues to perform consistently.
A UV LED Lamp Can Be ON but Still Need Checking
One common assumption is:
“The UV lamp is still ON, so it must be okay.”
This is not a reliable measurement method.
A UV LED module can continue producing light while the actual irradiance at the work surface differs from previous conditions.
Possible reasons include:
LED aging
Optical contamination
Cooling problems
Power-supply changes
Increased lamp distance
Mechanical misalignment
Different machine settings
A measurement instrument allows these changes to be investigated quantitatively.
What Does Linshang LS137 Measure?
The LS137 can provide several useful parameters:
UV Energy
UV Irradiance
Peak UV Irradiance
Measurement Duration
Temperature
Maximum Temperature
These measurements allow users to evaluate the curing cycle rather than simply checking whether the UV LED source is operating.
Step 1 – Identify Your UV Light Source
Before performing a measurement, confirm the type of UV source.
Is it:
UV LED
or
Traditional mercury UV lamp?
This distinction matters because different UV meters are designed for different spectral characteristics.
The Linshang LS137 is intended for UV LED sources.
If your system uses a traditional high-pressure mercury UV lamp, another instrument such as the Linshang LS136 may be more appropriate.
Step 2 – Check the UV Wavelength
Identify the wavelength of the UV LED system.
Common industrial UV LED wavelengths include:
365 nm
385 nm
395 nm
405 nm
The Linshang LS137 has a spectral response of approximately:
340–420 nm
and is calibrated using a:
395 nm UV LED source
For this reason, 395 nm UV LED curing systems are particularly relevant applications for the LS137.
Step 3 – Record the Machine Settings
Before taking a measurement, record the current operating conditions.
These may include:
UV power percentage
Conveyor speed
Exposure time
Lamp-to-product distance
UV wavelength
Production mode
Why?
Because measurement results are most useful when they can be reproduced later under comparable conditions.
Step 4 – Position the Meter Correctly
The LS137 should experience UV exposure that represents the actual curing process as closely as practical and within the manufacturer's operating requirements.
For a conveyor system, this may involve measuring along the same curing path used by the product.
For a stationary curing system, the measurement position should represent the relevant product position.
Consistency is important.
If the distance or position changes significantly between tests, the readings may not be directly comparable.
Step 5 – Measure UV Irradiance
UV irradiance indicates the instantaneous strength of the UV radiation reaching the detector.
It is commonly expressed in:
mW/cm²
The LS137 supports UV irradiance measurement up to:
40,000 mW/cm²
This makes it suitable for many high-intensity industrial UV LED curing applications within its specified operating conditions.
Step 6 – Measure Accumulated UV Energy
For curing applications, peak intensity is only part of the information.
Accumulated UV energy indicates how much UV exposure is received during the measurement cycle.
It is commonly expressed in:
mJ/cm²
This parameter becomes especially important in conveyor applications.
A faster conveyor can reduce exposure time and therefore reduce total UV energy even if peak irradiance remains similar.
Step 7 – Check the Peak Value
Peak UV irradiance can be useful when comparing:
Different lamps
Different curing stations
Different power settings
Before and after maintenance
New and aged UV LED modules
However, measurements should be made under comparable conditions.
Changing distance, position or operating mode can also change the reading.
Step 8 – Review Temperature
The LS137 also records temperature and maximum temperature.
This provides additional information about conditions during measurement.
Temperature may be relevant when investigating:
Cooling performance
Heat-sensitive substrates
Electronics components
Process stability
It should be evaluated together with the requirements of the production process.
Don't Just Measure Once
One measurement tells you the current UV condition.
Repeated measurements provide much more value.
For example:
New UV LED module
→ Record baseline.
After several months
→ Measure again.
After maintenance
→ Measure again.
When curing problems occur
→ Compare against baseline.
This creates a history of UV curing performance.
Example: Testing a 395 nm UV LED Printing Machine
Consider a printing company operating a 395 nm UV LED curing system.
When the machine is operating correctly, the production team records:
UV peak irradiance
UV accumulated energy
Exposure duration
Maximum temperature
Conveyor speed
UV power setting
Six months later, operators notice that some prints are not curing as consistently as before.
The LS137 is used again under the same machine settings.
If UV Output Is Lower
Investigate:
UV LED condition
Optical contamination
Cooling
Power supply
Lamp distance
If Peak Intensity Is Similar but Energy Is Lower
Check:
Conveyor speed
Exposure time
Production settings
If Measurements Are Similar to Baseline
Investigate other variables such as:
Ink
Substrate
Ink thickness
Material condition
Measurement helps narrow down the problem.
Example: Comparing Two UV LED Lamps
Suppose a factory has two similar UV curing machines.
Both machines show:
UV Power = 80%
This does not necessarily mean both produce identical UV output.
The LS137 can be used under controlled, comparable conditions to measure each system.
The factory may discover that:
Machine A and Machine B have different actual irradiance or accumulated energy.
This information is much more useful than comparing power percentages alone.
Testing Before and After Cleaning
Optical contamination can gradually reduce UV transmission.
A useful maintenance check can be:
Measure → Clean according to equipment procedure → Measure again
If the measured UV output improves significantly after cleaning, the factory has evidence that contamination was affecting the process.
This can help establish an appropriate preventive-maintenance interval.
Testing After UV LED Replacement
After installing a replacement UV LED module, do not simply assume the machine has returned to the original curing condition.
Measure it.
Record the new:
Peak irradiance
Accumulated energy
Exposure duration
Temperature
This becomes the new reference for future monitoring.
Testing After Increasing Production Speed
Suppose management wants to increase conveyor speed by 20%.
Production capacity improves, but curing exposure changes.
A useful procedure is:
Measure at original speed → Increase speed → Measure again → Check product quality
This provides data showing how the production change affects UV energy.
What Is a Good UV Reading?
There is no universal UV energy or irradiance number that is correct for every application.
The required exposure depends on factors such as:
Ink
Coating
Adhesive
Photoinitiator
Substrate
Material thickness
UV wavelength
Production speed
The correct target should be established from the material requirements and validated production process.
The LS137 measures the UV condition; it does not replace the curing specification of the material manufacturer.
Create a UV LED Inspection Record
For repeatable testing, consider recording:
Date
Machine
UV wavelength
UV LED operating hours
UV power setting
Conveyor speed
Measurement position
Peak UV irradiance
Accumulated UV energy
Exposure duration
Maximum temperature
Product quality result
Over time, this becomes useful preventive-maintenance and QA data.
When Should UV LED Output Be Tested?
Consider measurement:
During machine commissioning
When establishing a new process
After installing a new UV LED module
After maintenance
After cleaning
After changing conveyor speed
After changing curing distance
When product quality changes
During scheduled preventive maintenance
When comparing different machines
The appropriate testing frequency depends on the criticality of the production process.
Who Can Use the LS137?
The Linshang LS137 can be useful for:
Production engineers
Process engineers
Maintenance technicians
QA/QC personnel
Printing companies
Coating manufacturers
Electronics factories
UV adhesive users
UV equipment suppliers
It provides objective UV measurement data that can support process control and troubleshooting.
From “The Lamp Looks Fine” to Measured Data
A better way to manage UV LED curing is:
Do not ask only: “Is the lamp ON?”
Ask:
What is the measured UV intensity?
What is the accumulated UV energy?
How does it compare with our known-good baseline?
This turns UV curing performance into a measurable production parameter.
Looking for a UV LED Lamp Tester in Malaysia?
MTM Precision supplies industrial testing and measurement instruments for Malaysian manufacturing, engineering, maintenance and quality-control applications.
If you need to check the output of your UV LED curing equipment, provide us with:
UV wavelength
UV lamp/source type
Curing machine model
Application
Expected UV intensity
Measurement method
We can assist you in determining whether the Linshang LS137 UV Energy Meter is suitable for your application.
Contact MTM Precision
MTM Precision Sdn. Bhd.
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia
🌐 Website: www.mtmpre.com.my
📧 Email: mtmpre@yahoo.com
📱 WhatsApp: +6016-660 7346
Contact MTM Precision for Linshang LS137 UV Energy Meter enquiries, technical information and quotation in Malaysia.
13 Aug 2026