How to Test UV LED Lamp Output with Linshang LS137 UV Energy Meter Malaysia

How to Test UV LED Lamp Output with Linshang LS137 UV Energy Meter Malaysia

 

A UV LED lamp may still turn on and look normal even when its actual curing performance has changed.

 

For industrial applications, checking whether the lamp is illuminated is therefore not enough.

 

Printing companies, coating manufacturers, electronics factories and UV adhesive users may need to know:

 

How strong is the UV output?

 

How much UV energy is delivered during the curing cycle?

 

Has the UV output changed compared with previous production conditions?

 

The Linshang LS137 UV Energy Meter is designed for UV LED light sources and provides a practical method for measuring UV irradiance, accumulated UV energy and temperature.

 

Why Test UV LED Lamp Output?

 

UV LED curing systems are used in many industrial processes where curing quality directly affects the finished product.

 

Examples include:

 

UV printing

 

UV ink curing

 

UV coating

 

UV adhesive curing

 

Electronics assembly

 

PCB manufacturing

 

Optical bonding

 

Industrial production lines

 

If UV output changes, the final curing result may also change.

 

Periodic testing helps production teams understand whether the UV source continues to perform consistently.

 

A UV LED Lamp Can Be ON but Still Need Checking

 

One common assumption is:

 

“The UV lamp is still ON, so it must be okay.”

 

This is not a reliable measurement method.

 

A UV LED module can continue producing light while the actual irradiance at the work surface differs from previous conditions.

 

Possible reasons include:

 

LED aging

 

Optical contamination

 

Cooling problems

 

Power-supply changes

 

Increased lamp distance

 

Mechanical misalignment

 

Different machine settings

 

A measurement instrument allows these changes to be investigated quantitatively.

 

What Does Linshang LS137 Measure?

 

The LS137 can provide several useful parameters:

 

UV Energy

 

UV Irradiance

 

Peak UV Irradiance

 

Measurement Duration

 

Temperature

 

Maximum Temperature

 

These measurements allow users to evaluate the curing cycle rather than simply checking whether the UV LED source is operating.

 

Step 1 – Identify Your UV Light Source

 

Before performing a measurement, confirm the type of UV source.

 

Is it:

 

UV LED

 

or

 

Traditional mercury UV lamp?

 

This distinction matters because different UV meters are designed for different spectral characteristics.

 

The Linshang LS137 is intended for UV LED sources.

 

If your system uses a traditional high-pressure mercury UV lamp, another instrument such as the Linshang LS136 may be more appropriate.

 

Step 2 – Check the UV Wavelength

 

Identify the wavelength of the UV LED system.

 

Common industrial UV LED wavelengths include:

 

365 nm

 

385 nm

 

395 nm

 

405 nm

 

The Linshang LS137 has a spectral response of approximately:

 

340–420 nm

 

and is calibrated using a:

 

395 nm UV LED source

 

For this reason, 395 nm UV LED curing systems are particularly relevant applications for the LS137.

 

Step 3 – Record the Machine Settings

 

Before taking a measurement, record the current operating conditions.

 

These may include:

 

UV power percentage

 

Conveyor speed

 

Exposure time

 

Lamp-to-product distance

 

UV wavelength

 

Production mode

 

Why?

 

Because measurement results are most useful when they can be reproduced later under comparable conditions.

 

Step 4 – Position the Meter Correctly

 

The LS137 should experience UV exposure that represents the actual curing process as closely as practical and within the manufacturer's operating requirements.

 

For a conveyor system, this may involve measuring along the same curing path used by the product.

 

For a stationary curing system, the measurement position should represent the relevant product position.

 

Consistency is important.

 

If the distance or position changes significantly between tests, the readings may not be directly comparable.

 

Step 5 – Measure UV Irradiance

 

UV irradiance indicates the instantaneous strength of the UV radiation reaching the detector.

 

It is commonly expressed in:

 

mW/cm²

 

The LS137 supports UV irradiance measurement up to:

 

40,000 mW/cm²

 

This makes it suitable for many high-intensity industrial UV LED curing applications within its specified operating conditions.

 

Step 6 – Measure Accumulated UV Energy

 

For curing applications, peak intensity is only part of the information.

 

Accumulated UV energy indicates how much UV exposure is received during the measurement cycle.

 

It is commonly expressed in:

 

mJ/cm²

 

This parameter becomes especially important in conveyor applications.

 

A faster conveyor can reduce exposure time and therefore reduce total UV energy even if peak irradiance remains similar.

 

Step 7 – Check the Peak Value

 

Peak UV irradiance can be useful when comparing:

 

Different lamps

 

Different curing stations

 

Different power settings

 

Before and after maintenance

 

New and aged UV LED modules

 

However, measurements should be made under comparable conditions.

 

Changing distance, position or operating mode can also change the reading.

 

Step 8 – Review Temperature

 

The LS137 also records temperature and maximum temperature.

 

This provides additional information about conditions during measurement.

 

Temperature may be relevant when investigating:

 

Cooling performance

 

Heat-sensitive substrates

 

Electronics components

 

Process stability

 

It should be evaluated together with the requirements of the production process.

 

Don't Just Measure Once

 

One measurement tells you the current UV condition.

 

Repeated measurements provide much more value.

 

For example:

 

New UV LED module

 

→ Record baseline.

 

After several months

 

→ Measure again.

 

After maintenance

 

→ Measure again.

 

When curing problems occur

 

→ Compare against baseline.

 

This creates a history of UV curing performance.

 

Example: Testing a 395 nm UV LED Printing Machine

 

Consider a printing company operating a 395 nm UV LED curing system.

 

When the machine is operating correctly, the production team records:

 

UV peak irradiance

 

UV accumulated energy

 

Exposure duration

 

Maximum temperature

 

Conveyor speed

 

UV power setting

 

Six months later, operators notice that some prints are not curing as consistently as before.

 

The LS137 is used again under the same machine settings.

 

If UV Output Is Lower

 

Investigate:

 

UV LED condition

 

Optical contamination

 

Cooling

 

Power supply

 

Lamp distance

 

If Peak Intensity Is Similar but Energy Is Lower

 

Check:

 

Conveyor speed

 

Exposure time

 

Production settings

 

If Measurements Are Similar to Baseline

 

Investigate other variables such as:

 

Ink

 

Substrate

 

Ink thickness

 

Material condition

 

Measurement helps narrow down the problem.

 

Example: Comparing Two UV LED Lamps

 

Suppose a factory has two similar UV curing machines.

 

Both machines show:

 

UV Power = 80%

 

This does not necessarily mean both produce identical UV output.

 

The LS137 can be used under controlled, comparable conditions to measure each system.

 

The factory may discover that:

 

Machine A and Machine B have different actual irradiance or accumulated energy.

 

This information is much more useful than comparing power percentages alone.

 

Testing Before and After Cleaning

 

Optical contamination can gradually reduce UV transmission.

 

A useful maintenance check can be:

 

Measure → Clean according to equipment procedure → Measure again

 

If the measured UV output improves significantly after cleaning, the factory has evidence that contamination was affecting the process.

 

This can help establish an appropriate preventive-maintenance interval.

 

Testing After UV LED Replacement

 

After installing a replacement UV LED module, do not simply assume the machine has returned to the original curing condition.

 

Measure it.

 

Record the new:

 

Peak irradiance

 

Accumulated energy

 

Exposure duration

 

Temperature

 

This becomes the new reference for future monitoring.

 

Testing After Increasing Production Speed

 

Suppose management wants to increase conveyor speed by 20%.

 

Production capacity improves, but curing exposure changes.

 

A useful procedure is:

 

Measure at original speed → Increase speed → Measure again → Check product quality

 

This provides data showing how the production change affects UV energy.

 

What Is a Good UV Reading?

 

There is no universal UV energy or irradiance number that is correct for every application.

 

The required exposure depends on factors such as:

 

Ink

 

Coating

 

Adhesive

 

Photoinitiator

 

Substrate

 

Material thickness

 

UV wavelength

 

Production speed

 

The correct target should be established from the material requirements and validated production process.

 

The LS137 measures the UV condition; it does not replace the curing specification of the material manufacturer.

 

Create a UV LED Inspection Record

 

For repeatable testing, consider recording:

 

Date

 

Machine

 

UV wavelength

 

UV LED operating hours

 

UV power setting

 

Conveyor speed

 

Measurement position

 

Peak UV irradiance

 

Accumulated UV energy

 

Exposure duration

 

Maximum temperature

 

Product quality result

 

Over time, this becomes useful preventive-maintenance and QA data.

 

When Should UV LED Output Be Tested?

 

Consider measurement:

 

During machine commissioning

 

When establishing a new process

 

After installing a new UV LED module

 

After maintenance

 

After cleaning

 

After changing conveyor speed

 

After changing curing distance

 

When product quality changes

 

During scheduled preventive maintenance

 

When comparing different machines

 

The appropriate testing frequency depends on the criticality of the production process.

 

Who Can Use the LS137?

 

The Linshang LS137 can be useful for:

 

Production engineers

 

Process engineers

 

Maintenance technicians

 

QA/QC personnel

 

Printing companies

 

Coating manufacturers

 

Electronics factories

 

UV adhesive users

 

UV equipment suppliers

 

It provides objective UV measurement data that can support process control and troubleshooting.

 

From “The Lamp Looks Fine” to Measured Data

 

A better way to manage UV LED curing is:

 

Do not ask only: “Is the lamp ON?”

 

Ask:

 

What is the measured UV intensity?

 

What is the accumulated UV energy?

 

How does it compare with our known-good baseline?

 

This turns UV curing performance into a measurable production parameter.

 

Looking for a UV LED Lamp Tester in Malaysia?

 

MTM Precision supplies industrial testing and measurement instruments for Malaysian manufacturing, engineering, maintenance and quality-control applications.

 

If you need to check the output of your UV LED curing equipment, provide us with:

 

UV wavelength

 

UV lamp/source type

 

Curing machine model

 

Application

 

Expected UV intensity

 

Measurement method

 

We can assist you in determining whether the Linshang LS137 UV Energy Meter is suitable for your application.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd.

 

Showroom & Service Centre

 

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia

 

🌐 Website: www.mtmpre.com.my

📧 Email: mtmpre@yahoo.com

📱 WhatsApp: +6016-660 7346

 

Contact MTM Precision for Linshang LS137 UV Energy Meter enquiries, technical information and quotation in Malaysia.


 

13 Aug 2026