How to Measure Surface Gloss Correctly with Linshang LS190 Gloss Meter Malaysia
How to Measure Surface Gloss Correctly with Linshang LS190 Gloss Meter Malaysia
A gloss meter is easy to use, but obtaining consistent results requires more than simply placing the instrument on a surface and reading the number.
The Linshang LS190 Gloss Meter is a portable 60° gloss meter designed for routine industrial quality control. It provides a 0–200 GU measuring range with 1 GU resolution.
For reliable batch comparison and QC inspection, operators should follow a consistent measurement procedure.
Step 1: Confirm That 60° Measurement Is Suitable
Before measuring, confirm that the required test method uses 60° measurement geometry.
The LS190 is a single-angle 60° gloss meter.
60° is widely used for general-purpose gloss measurement, but some very high-gloss or low-gloss applications may require another measurement geometry.
Always check:
Customer specification
Product standard
Internal QC procedure
Approved measurement method
Using the correct geometry is essential when comparing results.
Step 2: Check the Test Surface
The measurement area should represent the actual product surface.
Before testing, inspect the sample for:
Dust
Fingerprints
Oil
Water
Scratches
Surface contamination
Local defects
Contamination can affect reflected light and produce inconsistent readings.
If cleaning is permitted, use an appropriate method that will not alter the surface finish.
Step 3: Prepare the Gloss Meter
Before routine testing, ensure the instrument is ready for measurement according to the manufacturer's operating procedure.
The LS190 uses its standard board/base for calibration and protection.
Keeping this reference surface clean is important.
Avoid touching or scratching the reference surface unnecessarily.
Step 4: Place the Meter Flat on the Surface
Position the LS190 firmly against the test surface.
The measurement area should have good contact with the instrument.
Avoid:
Tilting the meter
Measuring across an edge
Measuring directly over a curved area that prevents proper positioning
Allowing the instrument to move during measurement
Incorrect positioning can reduce measurement consistency.
Step 5: Take the Measurement
Once the instrument is positioned correctly, obtain the GU reading.
For example:
74 GU
This value represents the measured gloss under the LS190's 60° measurement geometry.
Do not judge the product based on a single number unless your QC procedure specifically requires only one measurement.
Step 6: Measure Multiple Positions
Real manufactured surfaces may not be perfectly uniform.
For better representation, take readings at several positions.
For example:
Position 1: 74 GU
Position 2: 75 GU
Position 3: 73 GU
Position 4: 75 GU
Position 5: 74 GU
These results indicate relatively consistent surface gloss.
If the readings are:
74 GU
73 GU
52 GU
75 GU
74 GU
the abnormal position should be investigated.
Step 7: Use the Same Measurement Locations
For batch-to-batch comparison, measurement location matters.
If Batch A is measured at the centre of a component but Batch B is measured near an edge, differences may be caused by surface variation rather than actual process changes.
A better QC procedure defines specific measurement points.
For example:
Point A: centre
Point B: upper left
Point C: lower right
The same positions can then be used for every batch.
Step 8: Keep Measurement Direction Consistent When Necessary
Some surfaces may have directional characteristics caused by:
Brushing
Rolling
Grinding
Printing
Surface texture
Manufacturing processes
For these materials, measurement orientation can influence results.
Use the same orientation when comparing samples.
Consistency is one of the most important principles in gloss measurement.
Step 9: Compare Against an Approved Reference
A GU value by itself does not tell you whether a product is good or bad.
The reading should be compared against:
Approved reference sample
Customer specification
Internal QC standard
Historical production data
Suppose the approved sample is:
78 GU
Current production results:
Batch A: 77 GU
Batch B: 79 GU
Batch C: 78 GU
Batch D: 59 GU
Batch D clearly requires investigation.
Step 10: Establish QC Limits
The LS190 supports QC upper and lower limit settings.
For example, your acceptable range may be:
Lower limit: 70 GU
Upper limit: 85 GU
This makes routine production inspection easier because operators can compare results against predetermined acceptance criteria.
QC limits should be established from the relevant specification or properly approved reference—not chosen arbitrarily.
Step 11: Record Your Measurements
Measurement becomes much more valuable when results are recorded.
Useful information may include:
Date
Product
Batch number
Supplier
Measurement location
GU reading
Operator
PASS / FAIL result
Remarks
The LS190 supports USB communication with a computer and data export to Excel, helping companies maintain organized inspection records.
Step 12: Compare Trends, Not Only Individual Results
Gloss measurement can also help identify gradual process changes.
Consider the average production gloss:
January: 78 GU
February: 77 GU
March: 75 GU
April: 72 GU
May: 69 GU
Even if individual products have not yet caused a major customer complaint, the downward trend may indicate that something in the process is changing.
QC teams can investigate factors such as coating, curing, polishing or raw materials.
Common Gloss Measurement Mistakes
Several simple mistakes can reduce measurement consistency.
Measuring a Dirty Surface
Dust, fingerprints and oil can change the reading.
Measuring Different Locations
Comparing unrelated areas can create misleading results.
Tilting the Instrument
The meter should sit correctly against the measurement surface.
Mixing Measurement Geometries
Do not directly compare measurements obtained using different geometries without considering the applicable method.
Using Only Visual Judgement
The purpose of a gloss meter is to provide objective numerical information.
Ignoring Surface Direction
Directional materials should be measured consistently.
Failing to Keep Records
Without historical data, it becomes harder to identify gradual production changes.
Example: Powder Coating QC Procedure
A simple production procedure could be:
Select three finished samples from each production batch.
Ensure the surfaces are clean.
Measure the same defined positions on every sample.
Record the GU readings.
Compare results against the approved range.
Investigate abnormal results.
Save the measurement record.
This creates a repeatable process that different operators can follow.
Which Industries Can Use This Method?
The same basic measurement principles can apply to suitable:
Paint and coating products
Plastic components
Metal finishes
Powder-coated products
Ceramic surfaces
Tiles
Paper
Printing
Packaging
Furniture
General manufactured products
The exact procedure should always be adapted to the relevant product standard and QC requirement.
Why Measurement Consistency Matters
The purpose of using a gloss meter is not simply to obtain a GU number.
The objective is to produce comparable measurements.
If every operator uses a different method, the data becomes less useful.
A good gloss measurement procedure controls:
Measurement geometry
Sample condition
Measurement position
Measurement direction
Acceptance limits
Data recording
This turns a portable gloss meter into a useful quality-control tool.
Conclusion
Using the Linshang LS190 Gloss Meter correctly is straightforward, but consistent measurement technique is essential.
For reliable results:
Use the correct 60° measurement application, keep the surface and reference standard clean, position the instrument properly, measure consistent locations, establish appropriate QC limits and record the results.
With its 0–200 GU range, 1 GU resolution, QC judgement capability and USB/Excel data management, the LS190 can provide an economical solution for routine surface gloss inspection in Malaysia.
When measurement procedures are standardized, companies can move beyond subjective descriptions such as “too shiny” or “too dull” and build a more objective, traceable quality-control process.
Contact MTM Precision
MTM Precision Sdn. Bhd.
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia
🌐 Website: www.mtmpre.com.my
📧 Email: mtmpre@yahoo.com
📱 WhatsApp: +6016-660 7346
15 Aug 2026