Common Gloss Measurement Errors and How to Avoid Them – Linshang LS190 Malaysia

Common Gloss Measurement Errors and How to Avoid Them – Linshang LS190 Malaysia

 

A gloss meter provides numerical surface-gloss measurements, but incorrect testing methods can still produce inconsistent results.

 

For factories and QA/QC teams using the Linshang LS190 Gloss Meter, understanding common measurement mistakes is important for obtaining repeatable data.

 

The LS190 is a 60° gloss meter with a 0–200 GU measuring range and 1 GU resolution, designed for straightforward routine quality control.

 

Here are some common gloss measurement errors and how to reduce them.

 

Error 1: Measuring a Dirty Surface

 

Dust, fingerprints, oil, moisture and other contamination can affect the way light reflects from a surface.

 

This may change the GU reading.

 

How to Avoid It

 

Before measurement:

 

Inspect the test area

 

Remove loose dust where appropriate

 

Avoid touching the measurement area

 

Ensure the surface condition follows your QC procedure

 

Do not use a cleaning method that changes the actual finish being evaluated.

 

Error 2: Measuring Different Locations Every Time

 

Surface gloss may vary across a product.

 

If one operator measures the centre while another measures near an edge, their results may differ even when both measurements are technically correct.

 

How to Avoid It

 

Define specific measurement locations.

 

For example:

 

Point A – Centre

 

Point B – Upper left

 

Point C – Lower right

 

Use the same positions for every production batch.

 

Error 3: Holding the Meter at an Angle

 

The gloss meter should be positioned correctly against the test surface.

 

Tilting or rocking the instrument can reduce measurement consistency.

 

How to Avoid It

 

Place the LS190 firmly and flat against the surface.

 

Avoid testing locations where the instrument cannot sit properly.

 

Error 4: Measuring a Surface That Is Too Small

 

The measurement area must be suitable for the instrument.

 

Small components, narrow edges or irregular shapes may prevent correct positioning.

 

How to Avoid It

 

Confirm that the available flat surface is large enough for proper measurement before selecting the instrument for the application.

 

Error 5: Ignoring Curved Surfaces

 

A gloss meter designed for routine flat-surface measurement may not sit correctly on strongly curved components.

 

Poor contact can affect the measurement geometry.

 

How to Avoid It

 

Select a suitable flat measurement area whenever possible.

 

For specialized curved components, confirm whether a different measurement method or instrument is required.

 

Error 6: Comparing Different Measurement Angles

 

A GU value should not be considered independently of the measurement geometry.

 

The LS190 uses 60° geometry.

 

If another instrument or supplier report uses 20° or 85°, the results should not simply be compared as though they were obtained using identical conditions.

 

How to Avoid It

 

When specifying gloss requirements, record:

 

Measurement angle

 

GU requirement

 

Applicable standard

 

Measurement procedure

 

This helps avoid misunderstandings between suppliers and customers.

 

Error 7: Ignoring Surface Direction

 

Some materials have directional characteristics.

 

Examples include:

 

Brushed metal

 

Rolled surfaces

 

Printed materials

 

Textured products

 

Certain coated surfaces

 

Measurements taken in different directions may vary.

 

How to Avoid It

 

Use a consistent instrument orientation when measuring directional surfaces.

 

The QC procedure should specify the direction if necessary.

 

Error 8: Using Only One Reading

 

A single measurement may not represent the entire product.

 

Local surface variation can occur.

 

How to Avoid It

 

Where appropriate, measure several defined positions and evaluate the readings according to the relevant QC procedure.

 

For example:

 

Point 1: 76 GU

Point 2: 75 GU

Point 3: 77 GU

Point 4: 76 GU

 

These readings indicate good consistency.

 

If one point measures 51 GU, investigate the local area.

 

Error 9: Using an Incorrect Reference or Calibration Condition

 

The reference standard used with the instrument should be kept in appropriate condition.

 

Dust, scratches or contamination on a reference surface can affect measurement reliability.

 

How to Avoid It

 

Follow the manufacturer's calibration and maintenance instructions.

 

Keep the instrument's standard board/base clean and protected when not in use.

 

Error 10: Touching the Reference Surface

 

Repeatedly touching the reference surface can introduce fingerprints and contamination.

 

How to Avoid It

 

Handle the reference standard carefully and avoid unnecessary contact with the calibrated surface.

 

Error 11: Comparing Measurements from Different Procedures

 

Suppose Factory A measures three positions and reports the average.

 

Factory B measures only one location.

 

Even if both use 60° gloss meters, their reported results may not be directly comparable.

 

How to Avoid It

 

When gloss is an important specification, define a standard operating procedure covering:

 

Instrument

 

Measurement geometry

 

Sample preparation

 

Measurement positions

 

Number of readings

 

Reporting method

 

Error 12: Setting Arbitrary QC Limits

 

An operator may decide that 70–80 GU “looks reasonable” without reference to the actual product requirement.

 

This can create unnecessary rejection or allow unacceptable products to pass.

 

How to Avoid It

 

QC limits should come from appropriate sources such as:

 

Customer requirements

 

Product standards

 

Engineering specifications

 

Approved samples

 

Validated internal QC procedures

 

Error 13: Assuming Higher GU Means Better Quality

 

A higher gloss reading is not automatically better.

 

A matte product may intentionally have a low GU value.

 

How to Avoid It

 

Judge the result against the required specification—not against the assumption that shinier is better.

 

The goal is consistency with the required finish.

 

Error 14: Confusing Gloss with Colour

 

A product can have the correct colour but incorrect gloss.

 

Likewise, it can have the correct gloss but the wrong colour.

 

How to Avoid It

 

Understand the difference:

 

Gloss meter → measures surface gloss

 

Colorimeter → measures colour

 

Appearance-critical products may require both measurements.

 

Error 15: Failing to Record Results

 

If measurements are taken but not recorded, valuable QC information is lost.

 

Historical data can help identify gradual process changes.

 

How to Avoid It

 

Maintain records including:

 

Product

 

Batch

 

Date

 

Measurement location

 

GU value

 

Operator

 

PASS / FAIL result

 

Remarks

 

The LS190 supports USB communication and Excel data export, making systematic record keeping easier.

 

Error 16: Ignoring Trends

 

A product may technically remain within its acceptance limits while gradually changing over time.

 

For example:

 

January: 79 GU

February: 77 GU

March: 75 GU

April: 72 GU

May: 70 GU

 

The readings may still pass, but the trend suggests that the process should be reviewed.

 

How to Avoid It

 

Analyse historical measurement data rather than looking only at individual PASS / FAIL results.

 

Error 17: Buying Based Only on Price

 

The cheapest gloss meter is not automatically the most economical solution.

 

Likewise, the most expensive instrument is not automatically necessary.

 

How to Avoid It

 

Choose according to:

 

Measurement angle

 

GU range

 

Resolution

 

Surface size

 

QC requirements

 

Data requirements

 

Applicable standards

 

For routine 60° measurement within 0–200 GU, the Linshang LS190 can provide a cost-effective solution.

 

Building a Better Gloss Measurement Procedure

 

A good routine can be simple:

 

Confirm the correct measurement geometry.

 

Check the instrument and reference standard.

 

Inspect the test surface.

 

Measure predefined locations.

 

Keep orientation consistent.

 

Take the required number of readings.

 

Compare results with approved QC limits.

 

Record the data.

 

Investigate abnormal readings and trends.

 

Following the same method each time is essential for meaningful comparison.

 

Conclusion

 

Many gloss measurement problems are caused not by the instrument itself, but by inconsistent testing procedures.

 

Dirty surfaces, different measurement locations, incorrect positioning, mixed measurement geometries and poor record keeping can all reduce the usefulness of gloss data.

 

The Linshang LS190 Gloss Meter, with 60° measurement, 0–200 GU range, 1 GU resolution, QC limit functions and USB/Excel data management, provides a practical tool for routine industrial inspection.

 

When combined with a consistent measurement procedure, it can help Malaysian manufacturers improve batch comparison, supplier inspection, production monitoring and overall surface-quality control.

 

Contact MTM Precision

 

MTM Precision Sdn. Bhd.

Showroom & Service Centre

No. 29-1 & 29-2, Jalan Bandar 18,

Pusat Bandar Puchong,

47160 Puchong, Selangor, Malaysia

 

🌐 Website: www.mtmpre.com.my

📧 Email: mtmpre@yahoo.com

📱 WhatsApp: +6016-660 7346


 

15 Aug 2026