Common Gloss Measurement Errors and How to Avoid Them – Linshang LS190 Malaysia
Common Gloss Measurement Errors and How to Avoid Them – Linshang LS190 Malaysia
A gloss meter provides numerical surface-gloss measurements, but incorrect testing methods can still produce inconsistent results.
For factories and QA/QC teams using the Linshang LS190 Gloss Meter, understanding common measurement mistakes is important for obtaining repeatable data.
The LS190 is a 60° gloss meter with a 0–200 GU measuring range and 1 GU resolution, designed for straightforward routine quality control.
Here are some common gloss measurement errors and how to reduce them.
Error 1: Measuring a Dirty Surface
Dust, fingerprints, oil, moisture and other contamination can affect the way light reflects from a surface.
This may change the GU reading.
How to Avoid It
Before measurement:
Inspect the test area
Remove loose dust where appropriate
Avoid touching the measurement area
Ensure the surface condition follows your QC procedure
Do not use a cleaning method that changes the actual finish being evaluated.
Error 2: Measuring Different Locations Every Time
Surface gloss may vary across a product.
If one operator measures the centre while another measures near an edge, their results may differ even when both measurements are technically correct.
How to Avoid It
Define specific measurement locations.
For example:
Point A – Centre
Point B – Upper left
Point C – Lower right
Use the same positions for every production batch.
Error 3: Holding the Meter at an Angle
The gloss meter should be positioned correctly against the test surface.
Tilting or rocking the instrument can reduce measurement consistency.
How to Avoid It
Place the LS190 firmly and flat against the surface.
Avoid testing locations where the instrument cannot sit properly.
Error 4: Measuring a Surface That Is Too Small
The measurement area must be suitable for the instrument.
Small components, narrow edges or irregular shapes may prevent correct positioning.
How to Avoid It
Confirm that the available flat surface is large enough for proper measurement before selecting the instrument for the application.
Error 5: Ignoring Curved Surfaces
A gloss meter designed for routine flat-surface measurement may not sit correctly on strongly curved components.
Poor contact can affect the measurement geometry.
How to Avoid It
Select a suitable flat measurement area whenever possible.
For specialized curved components, confirm whether a different measurement method or instrument is required.
Error 6: Comparing Different Measurement Angles
A GU value should not be considered independently of the measurement geometry.
The LS190 uses 60° geometry.
If another instrument or supplier report uses 20° or 85°, the results should not simply be compared as though they were obtained using identical conditions.
How to Avoid It
When specifying gloss requirements, record:
Measurement angle
GU requirement
Applicable standard
Measurement procedure
This helps avoid misunderstandings between suppliers and customers.
Error 7: Ignoring Surface Direction
Some materials have directional characteristics.
Examples include:
Brushed metal
Rolled surfaces
Printed materials
Textured products
Certain coated surfaces
Measurements taken in different directions may vary.
How to Avoid It
Use a consistent instrument orientation when measuring directional surfaces.
The QC procedure should specify the direction if necessary.
Error 8: Using Only One Reading
A single measurement may not represent the entire product.
Local surface variation can occur.
How to Avoid It
Where appropriate, measure several defined positions and evaluate the readings according to the relevant QC procedure.
For example:
Point 1: 76 GU
Point 2: 75 GU
Point 3: 77 GU
Point 4: 76 GU
These readings indicate good consistency.
If one point measures 51 GU, investigate the local area.
Error 9: Using an Incorrect Reference or Calibration Condition
The reference standard used with the instrument should be kept in appropriate condition.
Dust, scratches or contamination on a reference surface can affect measurement reliability.
How to Avoid It
Follow the manufacturer's calibration and maintenance instructions.
Keep the instrument's standard board/base clean and protected when not in use.
Error 10: Touching the Reference Surface
Repeatedly touching the reference surface can introduce fingerprints and contamination.
How to Avoid It
Handle the reference standard carefully and avoid unnecessary contact with the calibrated surface.
Error 11: Comparing Measurements from Different Procedures
Suppose Factory A measures three positions and reports the average.
Factory B measures only one location.
Even if both use 60° gloss meters, their reported results may not be directly comparable.
How to Avoid It
When gloss is an important specification, define a standard operating procedure covering:
Instrument
Measurement geometry
Sample preparation
Measurement positions
Number of readings
Reporting method
Error 12: Setting Arbitrary QC Limits
An operator may decide that 70–80 GU “looks reasonable” without reference to the actual product requirement.
This can create unnecessary rejection or allow unacceptable products to pass.
How to Avoid It
QC limits should come from appropriate sources such as:
Customer requirements
Product standards
Engineering specifications
Approved samples
Validated internal QC procedures
Error 13: Assuming Higher GU Means Better Quality
A higher gloss reading is not automatically better.
A matte product may intentionally have a low GU value.
How to Avoid It
Judge the result against the required specification—not against the assumption that shinier is better.
The goal is consistency with the required finish.
Error 14: Confusing Gloss with Colour
A product can have the correct colour but incorrect gloss.
Likewise, it can have the correct gloss but the wrong colour.
How to Avoid It
Understand the difference:
Gloss meter → measures surface gloss
Colorimeter → measures colour
Appearance-critical products may require both measurements.
Error 15: Failing to Record Results
If measurements are taken but not recorded, valuable QC information is lost.
Historical data can help identify gradual process changes.
How to Avoid It
Maintain records including:
Product
Batch
Date
Measurement location
GU value
Operator
PASS / FAIL result
Remarks
The LS190 supports USB communication and Excel data export, making systematic record keeping easier.
Error 16: Ignoring Trends
A product may technically remain within its acceptance limits while gradually changing over time.
For example:
January: 79 GU
February: 77 GU
March: 75 GU
April: 72 GU
May: 70 GU
The readings may still pass, but the trend suggests that the process should be reviewed.
How to Avoid It
Analyse historical measurement data rather than looking only at individual PASS / FAIL results.
Error 17: Buying Based Only on Price
The cheapest gloss meter is not automatically the most economical solution.
Likewise, the most expensive instrument is not automatically necessary.
How to Avoid It
Choose according to:
Measurement angle
GU range
Resolution
Surface size
QC requirements
Data requirements
Applicable standards
For routine 60° measurement within 0–200 GU, the Linshang LS190 can provide a cost-effective solution.
Building a Better Gloss Measurement Procedure
A good routine can be simple:
Confirm the correct measurement geometry.
Check the instrument and reference standard.
Inspect the test surface.
Measure predefined locations.
Keep orientation consistent.
Take the required number of readings.
Compare results with approved QC limits.
Record the data.
Investigate abnormal readings and trends.
Following the same method each time is essential for meaningful comparison.
Conclusion
Many gloss measurement problems are caused not by the instrument itself, but by inconsistent testing procedures.
Dirty surfaces, different measurement locations, incorrect positioning, mixed measurement geometries and poor record keeping can all reduce the usefulness of gloss data.
The Linshang LS190 Gloss Meter, with 60° measurement, 0–200 GU range, 1 GU resolution, QC limit functions and USB/Excel data management, provides a practical tool for routine industrial inspection.
When combined with a consistent measurement procedure, it can help Malaysian manufacturers improve batch comparison, supplier inspection, production monitoring and overall surface-quality control.
Contact MTM Precision
MTM Precision Sdn. Bhd.
Showroom & Service Centre
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia
🌐 Website: www.mtmpre.com.my
📧 Email: mtmpre@yahoo.com
📱 WhatsApp: +6016-660 7346
15 Aug 2026