How to Use Linshang LS196 Gloss Meter Correctly Malaysia
How to Use Linshang LS196 Gloss Meter Correctly Malaysia
Correct measurement technique is important when using a gloss meter for industrial quality control. Even a good instrument can produce inconsistent results if the surface is dirty, the measurement position changes or the instrument is not placed correctly.
The Linshang LS196 Gloss Meter is a portable 60° gloss meter with a wide 0–1000 GU measurement range, suitable for general and high-gloss surface measurement.
This guide explains the basic steps for obtaining more consistent gloss measurements with the LS196.
Step 1: Identify the Material to Be Measured
Before starting, confirm that gloss measurement is appropriate for the surface and application.
Typical applications can include suitable surfaces such as:
Paint
Coatings
Plastic
Automotive finishes
Metal finishing
Electroplated components
Powder coating
Furniture
Ceramic tiles
Stone
Industrial components
The product specification or applicable testing standard should determine the required measurement procedure.
Step 2: Inspect the Measurement Surface
The condition of the test surface can affect the result.
Before measuring, check for:
Dust
Fingerprints
Oil
Water
Surface contamination
Scratches
Uneven areas
If necessary, clean the surface using an appropriate method that will not alter the finish.
For comparative QC measurements, all samples should be prepared in the same way.
Step 3: Check the Instrument
Before important measurements, check the condition of the LS196.
Follow the manufacturer's recommended calibration procedure when required.
This is particularly important when measurements are being used for:
Production acceptance
Supplier inspection
Customer reports
Batch comparison
Quality investigations
Consistent instrument verification helps improve confidence in the results.
Step 4: Define the Measurement Location
One of the most common causes of different gloss readings is measuring different locations.
A surface that appears uniform may actually have local variations.
For repetitive QC, establish fixed measurement points.
For example:
Point A – Top
Point B – Centre
Point C – Bottom
Use the same locations whenever comparing similar products.
Step 5: Place the LS196 Properly
Position the gloss meter firmly against the test surface.
Avoid:
Tilting the instrument
Moving it during measurement
Measuring across an unstable surface
Allowing gaps between the instrument and sample
Proper positioning is especially important when comparing small differences between samples.
Step 6: Take More Than One Reading
For many industrial applications, one measurement should not automatically represent an entire product.
Consider taking several readings.
For example:
Measurement 1 → Record
Measurement 2 → Record
Measurement 3 → Record
Compare the results to understand the uniformity of the surface.
For larger products, additional measurement locations may be required.
Step 7: Compare With a Reference
A gloss value becomes more useful when it has something to compare against.
Manufacturers can use:
Customer specification
Internal specification
Approved golden sample
Previous production batch
Supplier reference sample
This helps determine whether a measured value is acceptable for the actual application.
Step 8: Use QC Mode for Routine Inspection
The Linshang LS196 provides a QC mode.
Once appropriate acceptance limits are established, operators can compare production measurements against those requirements.
This is useful for repetitive applications such as:
Incoming inspection
Production QC
Final inspection
Supplier inspection
Batch comparison
It can reduce dependence on individual visual judgement.
Step 9: Record the Measurement Results
For professional QA/QC work, do not rely only on the number displayed at the time of measurement.
Record important information such as:
Product model
Batch number
Date
Measurement location
Operator
GU value
Acceptance result
The LS196 supports computer connection and data transmission, allowing results to be exported for further processing and analysis.
Step 10: Analyse Unexpected Results
If a gloss reading is significantly different from normal, repeat the measurement before rejecting the product.
Check:
Surface cleanliness
Measurement location
Instrument positioning
Calibration condition
Surface curvature
Actual production variation
If repeated measurements confirm the difference, investigate the manufacturing process.
Common Mistake: Measuring a Dirty Surface
Fingerprints can be particularly noticeable on glossy surfaces.
If one sample is clean and another contains fingerprints or oil, the readings may not be directly comparable.
Surface preparation should therefore be part of the company's measurement procedure.
Common Mistake: Changing Measurement Positions
Measuring the centre of one sample and the edge of another can introduce unnecessary variation.
Mark or define standard inspection locations whenever possible.
This becomes especially important when several operators use the same instrument.
Common Mistake: Relying on One Reading
A single measurement may miss local variation.
Multiple readings can help identify:
Uneven coating
Polishing differences
Surface defects
Mold variation
Production inconsistency
The required number of readings should be determined according to the product and QC procedure.
Common Mistake: Comparing Different Surface Types
Gloss readings should be interpreted within the context of the material and applicable measurement method.
Do not assume that the same numerical value has the same quality meaning for every product.
A gloss requirement should be established specifically for the surface being inspected.
Why the LS196 Is Useful for High-Gloss Surfaces
One of the main features of the Linshang LS196 is its wide 0–1000 GU measurement range.
This provides additional flexibility when measuring highly glossy and reflective surfaces.
Typical applications can include:
High-gloss coatings
Automotive finishes
Glossy plastics
Polished surfaces
Electroplated components
Decorative finishes
The same instrument can also be used for many conventional gloss measurement applications.
Create a Standard Gloss Measurement SOP
For factories using the LS196 regularly, a written measurement SOP can improve consistency.
The procedure can define:
Surface preparation
Instrument check
Calibration procedure
Measurement locations
Number of readings
Acceptance limits
Data-recording method
Action for out-of-specification results
This helps ensure that different operators follow the same measurement process.
Linshang LS196 Gloss Meter Supplier Malaysia
For Malaysian manufacturers and QA/QC departments, the Linshang LS196 provides a practical solution for objective surface gloss measurement.
Correct measurement technique, consistent inspection locations and proper data recording can help users obtain more meaningful results from the instrument.