Why Does My Temperature Humidity Meter Show Different Readings? TASI TA622C Troubleshooting Malaysia
It is common for users to place two temperature and humidity meters next to each other and discover that they do not show exactly the same reading.
One meter may display 25.2°C and 63%RH, while another shows 25.7°C and 67%RH.
Does this mean one of the instruments is faulty?
Not necessarily.
Differences can be caused by measurement accuracy, sensor response time, probe position, airflow, temperature changes, operator influence, calibration condition and sensor ageing.
The TASI TA622C Digital Thermo-Hygrometer is specified for professional temperature and humidity measurement and uses a separate probe, making correct probe positioning particularly important when comparing measurements.
This guide explains what to check when temperature or humidity readings appear different.
1. Check the Accuracy Specification First
Two meters do not need to display exactly the same number to both be operating within their stated specifications.
The TASI TA622C has a specified temperature accuracy of:
±0.5°C
Its relative humidity accuracy is specified at:
±3%RH within 10–90%RH
Suppose the actual humidity is around 60%RH.
Two instruments, each operating within its own permitted tolerance, may still show slightly different values.
This is why comparing only the displayed numbers can be misleading.
2. Accuracy Is Not the Same as Resolution
One instrument may display:
62.8%RH
while another displays:
63%RH
The first instrument has more display digits.
That does not automatically mean it is more accurate.
Resolution describes how finely the instrument displays the measurement.
Accuracy describes how close the measurement is expected to be to the actual value under specified conditions.
Always compare specifications, not decimal places alone.
3. Are the Sensors Actually in the Same Location?
This is one of the easiest mistakes to make.
Two meters may appear to be next to each other, but their sensors could still be several centimetres apart.
In areas with:
Airflow
HVAC outlets
Heat sources
Windows
Equipment
Direct sunlight
even small differences in sensor location can produce different measurements.
When comparing instruments, position the sensing elements as close together as practical without obstructing airflow around them.
4. Allow Both Instruments to Stabilise
Do not compare two meters immediately after moving them into a new environment.
For example:
Outdoor environment → air-conditioned office
or:
Office → warehouse
or:
Normal room → cold room
The sensors need time to respond to the new environmental condition.
One instrument may respond faster than another.
During this period, their readings can be noticeably different even if both eventually stabilise at similar values.
5. Humidity Sensors May Respond at Different Speeds
Temperature and humidity meters do not necessarily have identical response characteristics.
One sensor may react faster to a change in humidity.
Another may take longer to stabilise.
Therefore, when comparing meters, avoid making conclusions based on the first few readings immediately after changing locations.
Wait until both instruments have had sufficient time to stabilise.
6. Your Hand Can Affect the Reading
The human body produces both heat and moisture.
If one sensor is:
Close to your hand
Close to your face
Near your body
while the other sensor is positioned farther away, the two instruments may show different results.
Breathing directly onto a humidity sensor can also cause the RH reading to increase.
The TA622C's separate probe allows the sensing element to be positioned away from the operator more easily.
7. Air-Conditioning Can Create Local Differences
A room does not always have one uniform temperature and humidity.
Conditions near an air-conditioning outlet can be different from conditions at the centre of the room.
Similarly, measurements near:
Return air grille
Door
Window
Ceiling
Floor
may differ.
When comparing meters, make sure both sensors are exposed to approximately the same environment.
8. Direct Sunlight Can Affect Temperature Readings
If one instrument is exposed to sunlight and another is shaded, the temperature readings may differ significantly.
Radiant heat can warm the sensor even if the surrounding air temperature has not changed by the same amount.
For normal ambient measurements, avoid direct sunlight unless solar exposure is intentionally part of the test.
9. Nearby Equipment Can Influence Measurements
Heat-producing equipment can create local temperature differences.
Examples include:
Motors
Computers
Servers
Lighting
Machinery
Power supplies
Production equipment
If one sensor is closer to the equipment, it may show a higher temperature.
This can also affect the relative humidity reading because RH depends on temperature.
10. Why Does Relative Humidity Change When Temperature Changes?
Relative humidity is temperature-dependent.
If air temperature changes while the moisture content remains similar, the RH value can change.
This means a small temperature difference between two sensor locations may also contribute to a difference in relative humidity.
This is one reason temperature and humidity should always be considered together.
11. Why Do Cheap Hygrometers Sometimes Differ More?
Low-cost room hygrometers may be designed primarily for general indication rather than professional inspection.
Different products may have different:
Sensor quality
Accuracy specifications
Response characteristics
Factory calibration
Long-term stability
This does not mean every inexpensive hygrometer is unsuitable.
For basic room monitoring, it may be perfectly adequate.
However, professional users should compare the actual specifications required for their application.
12. Check the Measurement Range
The TASI TA622C has a specified temperature measurement range of:
-40°C to 80°C
and relative humidity measurement range of:
0 to 100%RH
Its specified humidity accuracy of ±3%RH applies within 10–90%RH.
When operating near the limits of an instrument's measurement range, always check the manufacturer's detailed specifications.
13. Sensor Contamination Can Affect Humidity Measurement
Humidity sensors can be affected by their operating environment.
Potential contaminants may include:
Dust
Oil mist
Chemical vapour
Dirt
Condensation
Process contamination
Users should keep the sensing area clean and follow the manufacturer's handling recommendations.
Avoid unnecessarily touching the sensing element.
14. Condensation on the Sensor Can Cause Temporary Problems
Moving a meter rapidly from a warm humid environment into a much colder environment may create condensation under some conditions.
If condensation forms on or around the sensor, readings may temporarily become unreliable.
Allow the instrument to return to suitable operating conditions and stabilise before relying on the measurement.
15. Sensor Ageing Can Affect Performance
Sensors do not necessarily maintain identical performance forever.
Long-term use, environmental exposure and contamination can influence measurement performance.
For professional applications, users should maintain appropriate equipment checking and calibration procedures.
16. Has the Instrument Been Dropped or Damaged?
Physical damage can also affect measurement instruments.
If a meter begins showing unusual readings after:
Being dropped
Receiving an impact
Water exposure
Extreme environmental exposure
Probe damage
Cable damage
inspect the instrument and probe before continuing professional measurements.
If necessary, arrange further checking or calibration.
17. Check the TA622C Separate Probe
Because the TA622C uses a separate probe, inspect the probe and connection if readings appear abnormal.
Check for:
Loose connection
Damaged cable
Physical probe damage
Contamination
Moisture
Improper positioning
Do not immediately assume the sensing element itself has failed.
18. Compare Instruments Properly
If you want to compare two thermo-hygrometers, use a consistent method.
Place both sensing elements:
Close together
Away from your body
Away from direct sunlight
Away from unintended heat sources
Under similar airflow conditions
Then allow sufficient stabilisation time.
Compare the readings only after conditions are reasonably stable.
19. Do Not Use an Unverified Cheap Meter as a Reference Standard
Another common mistake is assuming one meter is correct simply because it is the meter you have used for many years.
For example:
TA622C = 64%RH
Old hygrometer = 59%RH
This does not automatically mean the TA622C is reading 5% too high.
Without an appropriate reference, you do not know which reading is closer to the actual value.
20. When Should Calibration Be Considered?
Calibration may be appropriate when:
Measurements are important to product quality
Customer requirements specify calibration
Internal SOP requires calibration
The instrument is used for audit records
Readings appear suspicious
The instrument has experienced damage
Periodic equipment control is required
The appropriate calibration interval depends on the application and quality requirements.
21. What If TA622C and Another Meter Differ by 1–2%RH?
A small difference does not automatically indicate a problem.
Consider:
Accuracy specifications of both instruments
Probe position
Stabilisation time
Temperature differences
Environmental fluctuations
If both meters are within their specified tolerances, a small difference may be expected.
22. What If the Difference Is Very Large?
If two instruments show a large and persistent difference after proper stabilisation, investigate further.
Check:
Probe positioning
Environmental stability
Sensor contamination
Probe connection
Instrument damage
Battery condition where relevant
Measurement settings
Calibration history
If the discrepancy remains significant, professional inspection or calibration may be appropriate.
23. Why Does TA622C RH Change When I Move the Probe?
This is normally expected if the probe is moved between locations with different environmental conditions.
For example:
Near HVAC outlet
Room centre
Near entrance
Upper warehouse rack
Lower warehouse rack
may all have different temperature and humidity conditions.
This is actually one of the advantages of using a portable separate-probe instrument: it can help reveal local environmental differences.
24. Use MAX / MIN / AVG to Understand Fluctuations
The TA622C provides:
MAX
MIN
AVG
If environmental conditions are changing continuously, these functions can provide a better picture than watching individual readings.
25. Use Data Recording When the Problem Is Intermittent
If users report that:
“Humidity is high sometimes, but normal when we check it.”
a single spot measurement may not be enough.
The TA622C provides data recording capability and supports USB PC communication.
Recorded measurements can help identify when environmental changes occur.
Frequently Asked Questions
Why do two humidity meters show different readings?
Possible reasons include accuracy tolerance, sensor response time, probe position, temperature differences, airflow, calibration and sensor condition.
Does a difference mean one meter is faulty?
Not necessarily. Both instruments may still be operating within their respective specifications.
Should two humidity meters show exactly the same value?
Not necessarily. Measurement instruments have tolerances.
Why does my humidity reading change when I hold the sensor?
Body heat and moisture from your hand or breathing can influence the local environment around the sensor.
Why does the reading change near an air conditioner?
The local temperature and humidity near an HVAC outlet can differ from general room conditions.
What is the humidity accuracy of TASI TA622C?
The specified accuracy is ±3%RH within 10–90%RH.
What is the temperature accuracy of TA622C?
The specified temperature accuracy is ±0.5°C.
Should TA622C be calibrated?
Calibration requirements depend on the application, quality system and measurement requirements.
Looking for Temperature Humidity Meter Calibration or Troubleshooting in Malaysia?
MTM Precision supplies professional temperature and humidity meters, thermo-hygrometers, data loggers and environmental testing instruments in Malaysia.
If you are experiencing inconsistent temperature or humidity readings, or are considering the TASI TA622C Digital Thermo-Hygrometer, contact MTM Precision.
We can help you determine whether the issue may involve measurement technique, application, instrument selection or the need for further checking and calibration.